Skip to main content
OpenTrials
Completed

NCT Number: NCT04417270

Reliability of the Freestyle Libre CGM in the Inpatient Setting During the COVID-19 Surge

The FDA has provided an emergency waiver for the use of non-invasive continuous glucose monitors (CGM) in hospitals, so frontline health care providers (in hospitals) can remotely monitor patients with diabetes thus reducing patient's discomfort, limiting exposure to COVID-19 and preserving critical personal protective equipment (PPE). The FreeStyle Libre 14-day system is a continuous glucose monitoring system consisting of a handheld reader (smart phone may be used) and a sensor applied with adhesive to back of the upper arm. In order to evaluate the reliability of the Freestyle Libre CGM for in-patient use, we propose a study which will examine the correlation between the libre CGM data and capillary blood glucose test - the current standard of care taken by the Accuchek Inform II platform.

Completed

Looking for future studies?

Notify Me

Key information

Age range

18 year–99 year

Sex eligibility

All sexes

Study type

Interventional

Phase

Not applicable

Primary location

Lenox Hill Hospital

New York, 10022, United States

About this study

In hospitalized patients diagnosed with COVID-19, diabetes mellitus is associated with a higher mortality. Patients with controlled blood glucose have a lower mortality rate than those who are uncontrolled. However, a rigid blood glucose control - which may lead to hypoglycemia- is associated with higher mortality rates. Thus, adequate glucose monitoring is important for hospitalized diabetic patients.

Previous studies have shown that CGMS, in the outpatient setting, engender better glycemic control thereby reducing risk of hypoglycemia. CGMS offer a replacement for finger stick blood glucose monitoring which may not identify an incidence of hypoglycemia until symptoms drive the patient to test. Furthermore, finger stick blood glucose monitoring is painful and time consuming thus causing many patients to avoid this aspect of diabetes self-care management.

Freestyle Libre CGMs sample interstitial fluid glucose subcutaneously. Measurement frequencies typically range from 1 to 15 minutes and most commonly are every 5 minutes.

Previous studies in hospitalized patients living with diabetes mellitus on basal bolus regimen have shown:

  • Use of the CGM increased detection of both hypoglycemic and hyperglycemic events versus those sensed by POCT.

One investigation identified 88 postprandial hyperglycemic excursions in patients with CGM use as opposed to 61 in those with POCT. Another noted that when comparing the CGM use with that of POCT, the former identified 55 hypoglycemic events whereas the latter only found 12.

  • No difference in mean daily glucose CGM and POCT readings. In a study of 38 patients, the investigators noted no difference in mean daily glucose between the CGM and POCT.

Who can participate

Healthy volunteers accepted: No

Only the study team can determine whether someone qualifies for participation.

Inclusion criteria

  • Participants acquired from among all non-critical care in patients who are living with Diabetes Mellitus. Because the patients will have to be instructed on the use of the CGM, this study will be limited to English speaking/reading adults (18 years of age or older), who possess the cognitive and physical ability to participate.

Exclusion criteria

  • Minors, pregnant women, and critically ill.

Treatment and study plan

Freestyle Libre 14 day CGM system

Device

The FreeStyle Libre 14 day system is a continuous glucose monitoring system consisting of a handheld reader and a sensor worn on the back of the upper arm.

Accuchek Inform II platform

Device

The Accu-Chek Inform II system is designed to deliver data with speed and efficiency, helping to provide optimal care. Wireless meter quickly transfers data. The Accu-Chek Inform II system is the first point-of-care blood glucose device to offer truly wireless technology at the meter level.

Primary outcomes

  1. Comparison of values between two devices

    Time frame: Through end of patient hospitalization, an average of 14 days

    Correlation between blood glucose results from the Freestlye Libre 14-day CGM and Accuchek Inform II platform using the Bland-Altman method to determine whether the two glucose methods agree

Sponsors and collaborators

Lead sponsor

Northwell Health

Other

Registry information

Important dates

Study start
2020
Primary completion
2020
Study completion
2020
First posted
Jun 4, 2020
Registry last updated
Mar 2, 2021

OpenTrials presents study information sourced from ClinicalTrials.gov. The official registry record should be consulted for the latest information.

View the official ClinicalTrials.gov record (opens in a new tab)

This listing is for discovery and informational purposes only. It is not medical advice, does not guarantee that a study is recruiting, and does not determine eligibility. Contact the study team and a qualified healthcare professional when considering participation.

Published trials that share one or more normalized conditions with this study.