Reference Dose Levels During Fluoroscopically Guided Procedures in Operating Rooms

This study is conducted by a working group of the French Society of Medical Physics (SFPM). Its main aim is to establish reference dose levels for the most common procedures performed in operating rooms using mobile X-ray systems, hence helping medical physicists and surgeons to evaluate their practice and optimize patient radiation protection.

This is a multi-centric prospective study involving 73 medical institutions of different categories (public university hospitals, clinics, centers dedicated to cancer treatment, etc.). It consists on progressively collecting anonymous data for 15 to 30 procedures from a list of 62 types of procedures, belonging to 7 surgery specialties (neurosurgery, orthopedic surgery, digestive surgery, urology, cardiology, vascular surgery and multi-specialty). Collected data include patient BMI and information about the X-ray equipment, the medical procedure and the dosimetric parameters. Data collection doesn't require the access to the patient medical record and doesn't impact his medical care.

Proposed dose reference levels will be expressed in terms of KAP (Kerma-Area-Product), fluoroscopy time and air Kerma. Moreover, multiple statistical analyses will be done to investigate the impact of different variables on the procedure X-ray doses.

Completed

Looking for future studies?

Notify Me

Key information

Conditions

Age range

16 year–100 year

Sex eligibility

All sexes

Study type

Observational

Primary location

Hopital Lariboisiere

Paris, 75010, France

About this study

Context:

Reference dose levels, introduced by the International Commission on Radiological Protection (ICRP) in 1996, are intended to help professionals using medical imaging to evaluate their practice and optimize patient's doses. European directive 2013/59/Euratom emphasizes the need to establish, use and regularly review reference levels for diagnostic and interventional procedures and to publish guidelines on this topic. Furthermore, national reference levels were established for conventional radiology and CT exams. They are used as a tool for the optimization of medical practices. Besides, a working group of the French Society of Medical Physics (SFPM) has published in 2017 reference levels for interventional radiology procedures performed in a dedicated room and using fixed angiographic equipment. Although radiation protection is a big challenge encountered in operating rooms using mobile X-ray equipment, few data are available in the literature regarding the use of this type of equipment.

In this context, the SFPM suggested the creation of a working group in 2017 to establish reference levels for the main procedures realized in the operating rooms using mobile X-ray systems. The aim of this working group is to help medical physicists and surgeons in evaluating their practice and optimizing patient radiation protection.

Scope of the study: The main purpose of the study is to establish dosimetric reference levels, in terms of KAP (Kerma-Area-Product). Additional objectives consist on proposing reference levels in terms of fluoroscopy time and air kerma and on studying the impact of different parameters (surgical specialty, body mass index, use of zoom, fluoroscopy pulse rate) on the delivered doses.

Study type and cohort: This is a multi-centric prospective study which consists on progressively collecting anonymous data for 15 to 30 procedures from a list of 62 surgical procedures (n=62) belonging to 7 different surgery specialties. Selected specialties are neurosurgery (n=12), orthopedic surgery (n=15), digestive surgery (n=6), urology (n=10), cardiology (n=6), vascular surgery (n=11), multi-specialty (n=2). 73 health institutions of different categories (University hospitals, clinics, centers dedicated to cancer treatment, etc..) confirmed their participations to the study.

Data collection: Data will be collected using a password-protected Excel spreadsheet. This excel file will be sent to all the participating institutions at the beginning of data collection. Every institution will send the excel file completed with anonymized data to the scientific responsible of the study through the secure email server of the APHP (Assistance Publique des Hopitaux de Paris)) institution.

Analysis: The main goal of this study is to publish descriptive statistics parameters (mean, standard deviation, median, 1st quartile, 3rd quartile) for each dosimetric data (KAP, fluoroscopy time, kerma, etc..) and for each category of surgical procedures. Results will be presented with a 95 percentage confidence interval.

Moreover, correlation analysis between several variables will be done. The differences between detector and equipment technologies will be also studied using appropriate statistical tests.

Who can participate

Healthy volunteers accepted: No

Only the study team can determine whether someone qualifies for participation.

Inclusion criteria

  • Fluoroscopy guided procedures performed on adult patients (≥16 years old)
  • Fluoroscopy guided procedures performed with mobile X-ray systems. Procedures performed outside of the operating rooms but using mobile X-ray systems are included in this study.
  • Fluoroscopy guided procedures performed using mini C-arms or O-arms

Exclusion criteria

  • Fluoroscopy guided procedures performed on pediatric patients (<16 years old)
  • Fluoroscopy guided procedures performed with a fixed interventional radiology imaging system even if it was located in the operating rooms .
  • Medical procedures performed using both a mobile X-ray system and a fixed CT scanner.
  • Medical procedures performed using a mobile intra-operative CT scanner

Treatment and study plan

Primary outcomes

  1. Total KAP (Kerma-Air-Product)

    Time frame: KAP is provided by the X-ray equipment at the end of the procedure.

    KAP or Kerma-Area-Product is the main dosimetric parameter representing the patient exposure. Participants should choose the appropriate unit of the KAP among a list of five units (microGray.m², milliGray.centimeter², centiGray.centimeter², deciGray.centimeter², Gray.centimeter²). Note that Gray is the international unit of the radiation quantity.

Secondary outcomes

  1. Year of construction of the equipment and date of the last external quality control (QC) (yyyy)

    Time frame: baseline

    Year of construction of the equipment and date of the last external quality control

  2. Image receptor type of the equipment

    Time frame: baseline

    Participants should specify whether the images are obtained using a flat panel detector or an X-ray image intensifier.

  3. DACS Connection

    Time frame: baseline

    Participants should specify If the X-ray equipment is connected to a Dose Archiving and a Communication system (DACS) (Yes/no)

  4. Patient weight (kg) and height (cm)

    Time frame: Patient weight and height are provided from the equipment or given by the medical team during the procedure

    Weight and height will be combined to report BMI in kg/m^2.

  5. Size of Fields of View (FOV) (cm)

    Time frame: Baseline, The most used FOV during the procedure is provided by the medical staff

    Size of the smallest and largest Field of View (FOV); size of the largest FOV adapted to the size of the phantom; size of the most used FOV for the concerned procedure. (cm)

  6. Differences between the displayed and measured X-ray doses (percentage)

    Time frame: baseline

    Differences between the displayed and measured KAP for the smallest and biggest FOV; Differences between the displayed and measured air kerma available in the last quality control (QC) report. (Percentages).

  7. Total air kerma of the procedure including the one performed of an O-arm equipment

    Time frame: Data are provided by the X-ray equipment at the end of the procedure

    Participants should specify the unit by choosing the appropriate one among a list of 5 units (micro Gray, milli Gray, centi Gray, deci Gray, Gray).

  8. Total number of image frames (number)

    Time frame: Data are provided by the X-ray equipment at the end of the procedure

    Total number of image frames (number)

  9. Fluoroscopy mode used during the procedure

    Time frame: Data are provided by the X-ray equipment at the end of the procedure

    Participants should specify whether continuous, pulsed or mixed fluoroscopy modes are used during the procedure

  10. Fluoroscopy pulse rate ( pulses/second)

    Time frame: Data are provided by the X-ray equipment at the end of the procedure

  11. Fluoroscopy time (hh:mm:ss) of the procedure including the one performed of an O-arm equipment

    Time frame: Data are provided by the X-ray equipment at the end of the procedure

  12. KAP (Kerma-Area-Product) attributable to fluoroscopy

    Time frame: Data are provided by the X-ray equipment at the end of the procedure

    Participants should choose the appropriate unit of the KAP among a list of five units (µGy.m², mGy.cm², cGy.cm², dGy.cm², Gy.cm²).

  13. Medical specialty of the procedure

    Time frame: Data are provided by the medical staff during the procedure

    Participants should specify the medical specialty of the procedure by selecting one from a predefined list of seven categories: Orthopedic surgery, neurosurgery, cardiology, vascular surgery, urology, digestive surgery and multi-specialty.

  14. Name/type of the procedure

    Time frame: Data are provided by the medical staff during the procedure

    Participants should specify the medical description of the procedure.

  15. Additional information about the procedure giving further indications about the used materials or the anatomical region of interest.

    Time frame: Data are provided by the medical staff during the procedure

  16. Total KAP for procedures performed on an O-arm equipment

    Time frame: KAP is provided by the X-ray equipment at the end of the procedure

    Participants should choose the appropriate unit of the KAP among a list of five units (µGy.m², mGy.cm², cGy.cm², dGy.cm², Gy.cm²).

  17. Total DLP (Dose-Length-Product) of the procedures performed on an O-arm equipment.DLP is always expressed in mGy.cm.

    Time frame: DLP is provided by the X-ray equipment at the end of the procedure

    The DLP (Dose Length Product) is expressed in milliGray *centimeters (mGy.cm)

  18. CTDIvol (Volume Computed Tomography Dose Index) of the procedures performed on an O-arm equipment. CTDI is always expressed in mGy.

    Time frame: CTDIvol is provided by the X-ray equipment at the end of the procedure

    The CTDIvol is a standardized measure of radiation dose output of a CT scanner. This dose is expressed in milliGrays (mGy).

  19. Type of the dosimetric phantom used to calculate the CTDI of an acquisition for the procedures performed on an O-arm equipment.

    Time frame: Data are provided by the X-ray equipment at the end of the procedure

    Participants should select between a "head phantom (16 cm)" or a "body phantom (32 cm)"

Sponsors and collaborators

Lead sponsor

Societe Francaise de Physique Medicale

Other

Collaborators

  • C2i Santé, FRANCE
  • Fréjus-Saint-Raphaël hospital, FRANCE
  • Hopital Lariboisière
  • University Hospital, Angers
  • University hospital of Montpellier, FRANCE

Registry information

Official study title

Reference Dose Levels During Fluoroscopically Guided Procedures Performed Using Mobile X-ray Systems in Operating Rooms

Important dates

Study start
2018
Primary completion
2019
Study completion
2019
First posted
Aug 28, 2018
Registry last updated
May 13, 2020

OpenTrials presents study information sourced from ClinicalTrials.gov. The official registry record should be consulted for the latest information.

View the official ClinicalTrials.gov record (opens in a new tab)

This listing is for discovery and informational purposes only. It is not medical advice, does not guarantee that a study is recruiting, and does not determine eligibility. Contact the study team and a qualified healthcare professional when considering participation.

Published trials that share one or more normalized conditions with this study.