Skip to main content
OpenTrials
Completed

NCT Number: NCT05034185

Real-World Validation of an Artificial Intelligence Characterization Support (CADx) System

Colorectal cancer (CRC) is a leading cause of cancer-related morbidity and mortality worldwide, with rates of CRC predicted to increase. Colonoscopy is currently the gold standard of screening for CRC. Artificial intelligence (AI) is seen as a solution to bridge this gap in adenoma detection, which is a quality indicator in colonoscopy. AI systems utilize deep neural networks to enable computer-aided detection (CADe) and computer-aided classification (CADx). CADe is concerned with the detection of polyps during colonoscopy, which in turn is postulated to help decrease the adenoma miss-rate.

In contrast, CADx deals with the interpretation of polyp appearance during colonoscopy to determine the predicted histology. Prediction of polyp histology is crucial in helping Clinicians decide on a "resect and discard" or "diagnose and leave strategy". It is also useful for the Clinician to be aware of the predicted histology of a colorectal polyp in determining the appropriate method of resection in terms of safety and efficacy. While CADe has been studied extensively in randomized controlled trials, there is a lack of prospective data validating the use of CADx in a clinical setting to predict polyp histology.

The investigators plan to conduct a prospective, multi-centre clinical trial to validate the accuracy of CADx support for prediction of polyp histology in real-time colonoscopy.

Completed

Looking for future studies?

Notify Me

Key information

Age range

40 year and older

Sex eligibility

All sexes

Study type

Observational

Primary location

Changi General Hospital, National University Hospital, Singapore General Hospital and Tan Tock Seng Hospital

Singapore, 529889

About this study

Colonoscopy is currently the gold standard of screening for CRC. A 1% increase in adenoma detection rate (ADR) estimated to be associated with a 3% decreased risk of interval CRC. AI systems can be broadly divided into CADe (for detection) and CADx (for diagnosis, or prediction of polyp histology in the context of colonoscopy). CADe has been extensively studied, with several randomized controlled trials and meta-analysis showing a higher ADR when CADe is used compared to the control groups without CADe.

Besides the ADR, predicted polyp histology is a key component in the performance of colonoscopy as this enables the Clinician to make a decision regarding its management, as described above. In this regards, image-enhanced endoscopy (IEE) is often used to help Clinicians determine if colorectal polyps found on colonoscopy are neoplastic or hyperplastic. The most commonly used non-magnification classification is the NBI International Colorectal Endoscopic (NICE), while the Japan NBI Expert Team (JNET) classification is used where endoscopy systems with optical magnification and the proper training is available. However, these classification systems have varying diagnostic accuracy and interobserver agreement. Previous prospective studies looking at CADx have utilized endocytoscopy and autofluorescence imaging (CAD-AFI) with positive results. However, the major limitation in these CADx studies is that these imaging systems are costly and are not readily available in most centres worldwide. Furthermore, most Clinicians performing colonoscopies have not been trained in these modalities of imaging and will have to rely completely on the CADx function to detect polyps if these imaging modalities are used, without being able to fall back on their experience and training should there be doubts about the accuracy of a CADx diagnosis in a real-world setting.

The Fujifilm 7000 System (Fujifilm Corp., Tokyo) has been in routine clinical use in all tertiary institutions in Singapore. The CAD EYE system was developed by Fujifilm Corp to aid Clinicians in colonoscopy with CADe and CADx functions. The basic functions and handling of the colonoscope, as well as the endoscopy processing unit, are similar to what is currently available in clinical practice, with the added CAD EYE software. The controller has been configured to allow the operator to activate and deactivate the CAD function depending on the need for it. These functions can be turned on and off using a button on the controller by the Clinician. The CADe and CADx functions operate when white light and blue laser imaging (BLI) are used, respectively. This provides a unique opportunity to externally validate the use of the CADx support tool by evaluating its diagnostic accuracy with final polyp histology as the gold standard, while also comparing its performance in a clinical setting against a Clinician using IEE (which is the conventional method of predicting polyp histology in colonoscopy).

The investigators plan to conduct a prospective, multi-centre clinical trial to validate the accuracy of CADx support for prediction of polyp histology in real-time colonoscopy.

Who can participate

Only the study team can determine whether someone qualifies for participation.

Inclusion criteria

  • Patients who have an indication for colonoscopy and who have at least one polyp detected during colonoscopy
  • 40 years of age and above
  • Consent obtained for the study

Exclusion criteria

  • Less than 39 years of age
  • Declined participation in study
  • Patients with no polyps detected during colonoscopy
  • Patients with inflammatory bowel disease
  • Patients with known unresected colorectal cancer

Treatment and study plan

Computer-aided diagnosis (CADx) support tool

Device

The CADx support tool operates when the Clinician switches the preconfigured CAD EYE function on using a button on the controller while the scope system is in BLI mode. This is performed after the Clinician first makes an optical prediction of polyp histology using IEE as described. The CADx support tool will make a prediction of polyp histology as "hyperplastic" or "neoplastic".

Primary outcomes

  1. To evaluate the diagnostic performance of the CADx support tool compared to optical prediction of polyp histology by the Clinician in real-time colonoscopy in a clinical setting

    Time frame: 1 year

    Polyp histology used as gold standard

Secondary outcomes

  1. To determine the diagnostic performance of CADx versus optical prediction of polyp histology by endoscopist in the subgroup analysis

    Time frame: 1 year

    Subgroups include bowel preparation, size of polyp and location

Sponsors and collaborators

Lead sponsor

Changi General Hospital

Other

Collaborators

  • National University Hospital, Singapore
  • Singapore General Hospital
  • Tan Tock Seng Hospital

Registry information

Official study title

Real-World Validation of an Artificial Intelligence Characterization Support (CADx) System for Prediction of Polyp Histology in Colonoscopy: A Prospective Multicentre Study

Important dates

Study start
2021
Primary completion
2022
Study completion
2022
First posted
Sep 5, 2021
Registry last updated
Feb 22, 2023

OpenTrials presents study information sourced from ClinicalTrials.gov. The official registry record should be consulted for the latest information.

View the official ClinicalTrials.gov record (opens in a new tab)

This listing is for discovery and informational purposes only. It is not medical advice, does not guarantee that a study is recruiting, and does not determine eligibility. Contact the study team and a qualified healthcare professional when considering participation.

Published trials that share one or more normalized conditions with this study.