King Chulalongkorn Memorial Hospital
Patumwan, Bangkok, 10330, Thailand
NCT Number: NCT01599351
The objective of this study is to evaluate the radiation dose to patient at two positions (prone and left lateral) and to medical staff during Endoscopic retrograde cholangiopancreatography (ERCP) procedure using dose area product (DAP) meter and thermoluminescent dosimeter (TLD). Data were recorded on 20 patients at prone position and 20 patients at left lateral position.
Looking for future studies?
Notify Me18 year–80 year
All sexes
Interventional
Phase 4
Patumwan, Bangkok, 10330, Thailand
The radiographic-fluoroscopic system manufactured by Siemens Model POLYSTAR was used for ERCP procedure at King Chulalongkorn Memorial Hospital. Dose Area Product (DAP) manufactured by PTW Model DIAMENTOR E, a transmission ionization chamber was attached to X-ray collimator to record the dose-area (cGy.cm2) in order to determine the entrance surface air kerma (ESAK, mGy) of the patients at two positions (prone and left lateral) while the Kodak Portal Pack for localization imaging was placed on the patient's couch to determine the exposed area (cm2) on the surface of the patient. TLD was attached at five positions, left eye, thyroid, left forearm, lower abdomen and left leg of the medical staff during ERCP procedure.
Healthy volunteers accepted: No
Only the study team can determine whether someone qualifies for participation.
Inclusion criteria
Exclusion criteria
Fluoroscopy was used during ERCP procedure. With automatic brightness adjustment, patients with thicker body will get greater radiation dose.
Other names: Siemens Model Polystar
Time frame: within 12 hours after procedure
Transparent ionization chamber was placed at the light beam diaphragm of the x-ray tube to measure the dose in the air and the verification film was placed on the couch at fluoroscopic area to determine exposed area. Patient's dose-area product will express in cGy.cm2.
Time frame: Within 12 hours after procedure
Thermoluminescent dosimeters (TLD) were placed at left eye, thyroid, left forearm, lower abdomen, and left leg of endoscopists. The TLDs used in this study was TLD-100 which was lithium fluoride crystal doped with magnesium and titanium (LiF:Mg, Ti). After procedures all labeled TLDs were read by automatic TLD reader (Harshaw 5500, Thermo Scientific Corp., MA, USA).
Chulalongkorn University
Other
Different Radiation Exposures to Patient and Endoscopists Between Performing ERCP in Patients Lying Prone and Left Lateral Decubitus
OpenTrials presents study information sourced from ClinicalTrials.gov. The official registry record should be consulted for the latest information.
View the official ClinicalTrials.gov record (opens in a new tab)This listing is for discovery and informational purposes only. It is not medical advice, does not guarantee that a study is recruiting, and does not determine eligibility. Contact the study team and a qualified healthcare professional when considering participation.
Published trials that share one or more normalized conditions with this study.
NCT01677481
Exposure to Ionizing Radiation
Ahmedabad, Gujarat, India
View Trial DetailsNCT01794325
Cardiovascular Diseases, Coronary Disease
Porto Alegre, Rio Grande do Sul, Brazil
View Trial DetailsNCT01334931
Exposure to Ionizing Radiation, Occupational Exposure to Radiation
Porto Alegre, Rio Grande do Sul, Brazil
View Trial Details