Quantitative Characterization of Safe Irrigation for Ventricular Shunt Catheters

The objective of this pilot study is to perform a set of pressure and flow rate measurements in order to quantitatively characterize safe irrigation or "flushing" of a ventricular shunt catheter, a routine maneuver performed by a neurosurgeon during a shunt revision surgery. This maneuver involves injecting saline with a syringe to clear plugged shunt catheters. Based on data obtained in Stage I of the study, a device has been fabricated which provides the same small-volume pulse of fluid used by surgeons. In Stage II of the study, pressure and flow rate measurements generated by this specialized flusher device will be collected and measured.

Completed

Looking for future studies?

Notify Me

Key information

Age range

30 day–35 year

Sex eligibility

All sexes

Study type

Interventional

Phase

Not applicable

Primary location

Boston Children's Hospital

Boston, Massachusetts, 02115, United States

About this study

Primary Objective: To collect pressure and flow rate measurements during shunt revision surgeries in human patients.

Pressure recordings of the injected fluid will be obtained using disposable off-the-shelf sterile pressure transducers (which are typically used for intracranial and arterial pressures monitoring) readily available in the operating room. First, the surgeon will attempt to flush the catheter using the Alcyone in-line Flusher device. If this fails, the surgeon may elect to attempt to flush using a standard syringe attached to the catheter. The flow rates will be measured by video, recording the syringe during the flushing maneuver, and then extracting plunger motion information by applying image processing on the acquired videos. In the event the recording equipment is not available (in that these operations are usually unscheduled and sometimes urgent), the same irrigation techniques may be used without the pressure recordings.

Secondary Objective: To analyze the collected data and establish maximal allowable injection pressure and flow rate. The collected data segments for each patient will be analyzed and maximal values for pressure and the saline injection rate will be recorded in a spreadsheet. Inter-patient variation in the results, presumably due to different degrees of occlusion, will be statistically evaluated. We will also determine if any of the catheters which appear occluded in the operating room resume flow with the flushing maneuvers.

Who can participate

Healthy volunteers accepted: No

Only the study team can determine whether someone qualifies for participation.

Inclusion criteria

  • Subjects must be 30 days to 35 years of age, of any race or gender.
  • Subjects must be scheduled to undergo a neurosurgical procedure potentially requiring a revision of neurosurgical catheters.
  • Subjects aged >18 years of age must be willing and able to provide written informed consent. Parental consent will be required for minors.

Exclusion criteria

  • Inability or unwillingness of subject or parent/guardian to give informed consent/assent.
  • Judgment of the investigator, surgeon, or any member of the care team (including anesthesiology or operating room nursing staff) that participation in the study will interfere with, or be detrimental to, administration of optimal healthcare of the subject.
  • Inability or unwillingness of subject to endure the test or any other condition that would preclude or bias the results of the study according to the judgment of the investigator.

Treatment and study plan

Cerebrospinal Fluid (CSF) Drainage

Device

Placement of Alivio in line flusher during CSF drainage

Other names: Alivio in line flusher

Primary outcomes

  1. To Demonstrate That the Flusher Can be Primed, Connected, and Activated in a Sterile Clinical Setting Per the Instructions for Use by the Treating Investigator .

    Time frame: Intra-procedural

    Criteria for success:

    Investigator Determination if the Flusher can be properly primed. Investigator Determination if the Flusher can be easily connected to current shunt components.

    Investigator Determination if the Flusher can be activated.

    The following scenarios were also assessed:

    A noticeable resistance to compression/collapse of the dome to indicate that the Ventricular Catheter is too tightly blocked to be flushed.

    If the dome does not refill the Ventricular Catheter may still be non-flowing. If the dome refills, the flushing procedure may be repeated as needed and/or as determined reasonable by the surgeon.

Sponsors and collaborators

Lead sponsor

Alcyone Therapeutics, Inc

Industry

Registry information

Important dates

Study start
2016
Primary completion
2016
Study completion
2016
First posted
Jan 11, 2016
Registry last updated
Sep 21, 2020

OpenTrials presents study information sourced from ClinicalTrials.gov. The official registry record should be consulted for the latest information.

View the official ClinicalTrials.gov record (opens in a new tab)

This listing is for discovery and informational purposes only. It is not medical advice, does not guarantee that a study is recruiting, and does not determine eligibility. Contact the study team and a qualified healthcare professional when considering participation.

Published trials that share one or more normalized conditions with this study.