NCT Number: NCT07145463
This is a prospective, multicenter, randomized controlled clinical study designed to evaluate the real-world performance of the VitaFlow Liberty® Flex Transcatheter Aortic Valve Retrievable and Steerable Delivery System (Investigational Device) for treating severe aortic stenosis (AS) in challenging anatomies. The study will compare it against the VitaFlow Liberty® Retrievable Delivery System (Control Device), which lacks steerability.
Key Study Elements:
Objective: The primary objective is to assess the device's performance using the incidence of a composite endpoint before hospital discharge. This endpoint includes permanent pacemaker implantation (PPI), valve-in-valve (ViV) implantation, or moderate-to-severe paravalvular leakage (PVL).
Design: A prospective, multicenter RCT with 1:1 randomization (Experimental: VitaFlow Liberty® Flex vs. Control: VitaFlow Liberty®). Approximately 15 sites in China will participate. Subjects undergo follow-up at discharge, 30 days, and 1 year post-procedure. An independent data center handles management and analysis.
Population: Patients with severe AS (echo confirmed: peak velocity ≥4.0 m/s, mean gradient ≥40 mmHg, or AVA ≤1.0 cm²/AVAi ≤0.6 cm²/m²) AND preoperative CTA showing a challenging aortic-left ventricular angle >60°. NYHA class ≥II is required.
Devices:
Investigational: VitaFlow Liberty® Flex (Retrievable & Steerable Delivery System - Models DSRS21/24/27/30/A series; Loading Tools LT-S series).
Control: VitaFlow Liberty® (Retrievable Delivery System - Models DSR21/24/27/30; Loading Tools LT series).
Endpoints:
Primary: Composite of PPI, ViV, or moderate-to-severe PVL before discharge.
Secondary: Include individual components of the primary endpoint (ViV, PVL) immediately post-procedure, procedural success (VARC-3), technical assistance rates, valve retrievals, implantation depth, arch/valve crossing performance, valve hemodynamics (gradient, area, leak, LVEF), NYHA class, Major Adverse Cardiac and Cerebrovascular Events (MACCE - all-cause death, MI, stroke, reoperation), major vascular complications (at discharge/30 days).
Key Inclusion: Severe AS, challenging anatomy (aortic-LV angle >60°), NYHA ≥II, informed consent.
Key Exclusions: Device/contrast allergies, anticoagulant intolerance, active infection, severe vascular disease prohibiting access, ascending aorta ≥55mm, unsuitable aortic root anatomy, intracardiac mass/thrombus, recent MI (<30 days), severe concomitant mitral/tricuspid regurgitation, cardiogenic shock, severe LV dysfunction (LVEF<20%), hematologic abnormalities, pregnancy/breastfeeding, participation in other device trials.
Visits: Screening (≤30d pre-op), Procedure (intra-op to 24h post-op), Discharge (≤7d post-op), 30d Follow-up (±7d), 12m Telephone FU (±1m).
Sample Size: Planned enrollment of 232 subjects (116 per group), calculated for superiority testing. Based on an expected composite endpoint rate of 21% for the Flex system vs. a historical rate of 38% for non-steerable systems (Superiority margin Δ1=0%, one-sided α=2.5%, Power=80%, accounting for 5% dropout).
Purpose: This study aims to demonstrate the superiority of the retrievable and steerable VitaFlow Liberty® Flex delivery system in reducing the composite rate of key adverse events (PPI, ViV, significant PVL) at discharge compared to the non-steerable system, specifically in patients with severe AS and anatomically challenging aortic-left ventricular angles.
Interested in participating?
Request Info18 year and older
All sexes
Interventional
Not applicable
Shanghai, Shanghai Municipality, 200032, China
Location status: Recruiting
Junbo Ge, Dr
CONTACT
Wenzhi Pan, Dr
CONTACT
A Prospective, Multicenter, Randomized Controlled Clinical Trial Evaluating the VitaFlow Liberty® Flex Transcatheter Aortic Valve Retrievable and Steerable Delivery System for Severe Aortic Stenosis in Challenging Anatomies
Type: Prospective, multicenter, open-label, 1:1 randomized controlled superiority trial
Randomization: Centralized electronic system
Blinding: Unblinded (procedural nature precludes blinding)
Study Arms:
Experimental: VitaFlow Liberty® Flex TAVR System (Retrievable & Steerable)
Control: VitaFlow Liberty® TAVR System (Retrievable, Non-Steerable)
Follow-up Schedule:
Pre-discharge (≤7 days post-op)
30 days (±7 days)
12 months (±1 month, telephone)
Oversight: Independent Clinical Events Committee (CEC), Data Monitoring Committee (DMC), and Core Laboratories for imaging/ECG analysis
Centers: ~15 experienced TAVR centers across China
Device Component:VitaFlow Liberty® FLEX (Investigational) VitaFlow Liberty® (Control) Delivery System Type:Retrievable & Steerable(Investigational) Retrievable (Non-Steerable)(Control)
Diagnosis: Severe symptomatic AS confirmed by echocardiography:
Peak aortic velocity ≥4.0 m/s OR Mean gradient ≥40 mmHg OR Aortic valve area (AVA) ≤1.0 cm² (or AVAi ≤0.6 cm²/m²) Anatomical Inclusion: Pre-op CT angiography (CTA) with 3D reconstruction demonstrating aortic-LV angle >60° (validating challenging anatomy).
Symptoms: NYHA Functional Class ≥II.
Key Exclusion Criteria:
Allergy to device materials (nitinol), contrast, or antiplatelets/anticoagulants Active infection/endocarditis Severe vascular disease prohibiting access Ascending aorta diameter ≥55mm Unsuitable aortic root anatomy (e.g., heavy calcification affecting expansion) Intracardiac thrombus/mass Recent MI (<30 days) Severe mitral/tricuspid regurgitation LVEF <20% Hematologic disorders (leukopenia, thrombocytopenia, coagulopathy) Pregnancy or concurrent device trials
Primary Endpoint:
Composite of PPI, ViV implantation, or moderate-to-severe PVL before discharge.
Secondary Endpoints:
Procedural:
ViV rate (immediate post-op) Moderate-to-severe PVL (immediate post-op) Technical assistance crossing rate Valve retrieval attempts Valve implantation depth Arch/valve crossing performance Procedural success (VARC-3 criteria*)
Clinical:
PPI rate (discharge, 30 days) MACCE** (discharge, 30 days): All-cause death, MI, stroke, reoperation Major vascular complications
Hemodynamic/Echocardiographic:
Valve orifice area, mean gradient, peak velocity Degree of stenosis/regurgitation LVEF (immediate, discharge, 30 days)
Functional: NYHA Class improvement (discharge, 30 days)
*VARC-3: Valve Academic Research Consortium-3 criteria for technical success.
**MACCE: Major Adverse Cardiac and Cerebrovascular Events.
Assumptions:
Historical composite endpoint rate (non-steerable systems): 38% Expected composite rate (steerable system): 21% Superiority margin (Δ1): 0% Power: 80% | One-sided α: 2.5% | Dropout: 5% Sample Size: 232 subjects (116 per arm)
Analysis:
Primary endpoint: Chi-square/Fisher's exact test Secondary endpoints: Mixed-effects models accounting for site variability Subgroup analyses: Valve size, center volume, baseline aortic-LV angle
Screening (≤30 days pre-op): Echocardiography, CTA, NYHA assessment, informed consent.
Procedure: TAVR under standard anesthesia. Randomization pre-insertion. Discharge (≤7 days): Echo, ECG, NYHA, MACCE/vascular complication assessment. 30-Day Visit: Clinical exam, echo, ECG, lab tests, NYHA, MACCE. 12-Month Follow-up: Telephone assessment for mortality/MACCE.
Approved by institutional review boards (IRBs) at all sites. Conducted per ICH-GCP, Declaration of Helsinki, and Chinese regulatory requirements.
Independent DMC reviews safety data quarterly.
This trial addresses a critical unmet need in TAVR: improving outcomes in patients with challenging aortic anatomies. If superiority is demonstrated, the steerable delivery system may become the preferred option for:
Reducing PPI rates (linked to anatomical precision) Mitigating PVL (via optimal deployment) Avoiding ViV (through recapture/repositioning) Enhancing first-attempt success in tortuous anatomies
Conclusion:
This rigorously designed RCT will provide Level A evidence on the clinical utility of steerable TAVR technology in high-risk anatomies, potentially establishing a new standard for managing complex AS patients.
Healthy volunteers accepted: No
Only the study team can determine whether someone qualifies for participation.
Inclusion criteria
Exclusion criteria
Transcatheter Aortic Valve Implantation/Replacement (TAVI/TAVR) has gradually emerged as a new treatment option for patients with severe aortic stenosis who are ineligible for open-heart valve replacement or at high surgical risk.
Time frame: Perioperative/Periprocedural"
Incidence of composite endpoint events before discharge (permanent pacemaker implantation, valve-in-valve, or moderate-to-severe paravalvular leak) calculated as number of occurrences/total number of occurrences
Time frame: Perioperative/Periprocedural
Incidence of valve-in-valve calculated as: Occurred/Total
Time frame: Perioperative/Periprocedural
Incidence of Moderate-to-severe paravalvular leak calculated as: Occurred/Total
Time frame: Perioperative/Periprocedural
Incidence of Other technical assistance crossing rate (immediately post-procedure) calculated as: Occurred/Total
Time frame: Perioperative/Periprocedural
Incidence of valve retrievals calculated as: Occurred/Total
Time frame: Perioperative/Periprocedural
Valve implantation depth calculated as Intraoperative images
Time frame: Perioperative/Periprocedural, 30 days,
Incidence of Permanent Pacemaker Implantation calculated as: Occurred/Total
Time frame: Perioperative/Periprocedural
Arch Crossing and Valve Crossing Performance calculated as assessment scale
Time frame: Perioperative/Periprocedural, 30 days,
Valve Performance Evaluation, Including: Valve Orifice Area, Mean Transvalvular Pressure Gradient, Peak Transvalvular Velocity, Degree of Valve Stenosis, Degree of Valve Regurgitation, Degree of Paravalvular Leak, Left Ventricular Ejection Fraction (LVEF) calculated as cardiac uhrasonography
Time frame: Perioperative/Periprocedural, 30 days
NYHA Functional Class Assessment NYHA classification: Divided into four levels, Grade I: No discomfort in daily activities, and no restriction in physical activities. Grade II: Mild activity restriction, no symptoms at rest, but daily activities (such as climbing stairs) can cause fatigue and palpitation. Grade III: obvious activity limitation, no symptoms at rest, but slight activity (such as walking) will cause symptoms. Grade Ⅳ: There are symptoms at rest, and any activity aggravates discomfort.
Time frame: Perioperative/Periprocedural, 30 days,
Incidence of Major Adverse Cardiac and Cerebrovascular Events (MACCE),MACCE Includes: All-Cause Mortality, Myocardial Infarction, All Stroke, Reoperation Due to Valve Dysfunction calculated as: Occurred/Total
Time frame: Perioperative/Periprocedural, 30 days
Incidence of Major Vascular Complications calculated as: Occurred/Total
Time frame: Perioperative/Periprocedural
Procedural Success Rate (VARC-3 Definition) calculated as: Occurred/Total
Contact information is provided by the study sponsor or research team.
Shanghai Zhongshan Hospital
Other
A Prospective, Multicenter, Randomized Controlled Clinical Study Evaluating the VitaFlow Liberty® Flex Transcatheter Aortic Valve Retrievable and Steerable Delivery System for the Treatment of Severe Aortic Stenosis
Acronym: Flex-TAVR
OpenTrials presents study information sourced from ClinicalTrials.gov. The official registry record should be consulted for the latest information.
View the official ClinicalTrials.gov record (opens in a new tab)This listing is for discovery and informational purposes only. It is not medical advice, does not guarantee that a study is recruiting, and does not determine eligibility. Contact the study team and a qualified healthcare professional when considering participation.
Published trials that share one or more normalized conditions with this study.
NCT05283356
Aortic Valve Disease, Aortic Valve Stenosis
Vigo, Pontevedra, Spain
View Trial DetailsNCT05172973
Aortic Valve Disease, Aortic Valve Insufficiency
Newport Beach, California, United States
View Trial DetailsNCT06773780
Aortic Valve Disease, Aortic Valve Stenosis
Istanbul, Turkey (Türkiye)
View Trial DetailsNCT07709858
Aortic Valve Disease, Aortic Valve Stenosis
Seoul, South Korea
View Trial Details