NCT Number: NCT01460992

ProMRI AFFIRM Study of the EVIA/ENTOVIS Pacemaker With Safio S Pacemaker Leads

This investigation is designed to demonstrate the clinical safety of the EVIA/ ENTOVIS pacemaker system when used under specific MRI conditions.

Completed

Looking for future studies?

Notify Me

Key information

Age range

18 year and older

Sex eligibility

All sexes

Study type

Observational

Primary location

Allgemeines Krankenhaus Linz, Linz, Austria

Loading trial locations.

About this study

In the past, MR scans were always contraindicated for pacemaker patients. If particular prerequisites and conditions are fulfilled, MR scans can now be conducted on patients with BIOTRONIK's EVIA/ENTOVIS pacemakers in combination with the Safio S pacemaker leads, scheduled to be part of this clinical study.

The specific MRI conditions for this investigation with the EVIA/ENTOVIS pacemaker family and Safio S pacemaker leads are defined in the study protocol

Who can participate

Healthy volunteers accepted: No

Only the study team can determine whether someone qualifies for participation.

Inclusion criteria

  • Pacemaker system consisting solely of the EVIA/ENTOVIS pacemaker and Safio S pacemaker leads
  • The pacemaker system has to be implanted, repositioned or exchanged at least 5 weeks prior to enrollment
  • Age more than 18 years
  • Understand the nature of the procedure
  • Able and willing to complete MRI testing
  • Able and willing to activate and use the Cardio Messenger
  • Give written informed consent
  • Able to complete all testing required by the clinical protocol
  • Ability to measure atrial and/or ventricular pacing threshold(s)* (at 0.4 ms)
  • All pacing thresholds do not exceed 2.0V @0.4ms.
  • Available for follow-up visit at the investigational site
  • Patient body height greater or equal to 140 cm
  • Pectoral implantation
  • The ascertained lead impedance is between 200 and 1500 Ohms.

Exclusion criteria

  • No EVIA/ENTOVIS /Safio S pacemaker system implanted
  • Pacing threshold(s) (at 0.4 ms) and sensing amplitudes are not measurable
  • Meet one or more of the contraindications
  • Being pregnant
  • Have a life expectancy of less than three months
  • Cardiac surgery already scheduled in the next three months
  • Enrolled in another cardiac clinical investigation
  • Have other medical implants that may interact with MRI, e.g. abandoned pacemaker/ICD leads, lead extensions, other active medical devices, non-MRI compatible devices (e.g. mechanical tricuspid valve)
  • Have other metallic artifacts/components in body that may interact with MRI

Treatment and study plan

Evia/Entovis pacemaker system under Magnetic Resonance Imaging (MRI) conditions

Procedure

patients will undergo an MRI scan for 30 minutes.

Other names: Evia/Entovis SR-T (single chamber with telemetry function), DR-T (dual chamber with telemetry function), Atrial/ventricular lead(s): Safio S 53 and S 60., Safio S 45

Primary outcomes

  1. Primary hypothesis 1: MRI and pacing system related Serious Adverse Device Effect (SADE) free rate

    Time frame: 3 months

    MRI and pacing system related Serious Adverse Device Effect (SADE) Free-Rate is greater than 90%.

  2. Primary hypotheses 2 and 3: pacing threshold rise (atrial and ventricular)

    Time frame: 1 month

    The atrial or ventricular MRI induced pacing threshold rise is not larger than or equal to 0.5 V between Pre-MRI and 1-month follow-up

  3. Primary hypothesis 4: P-wave sensing attenuation

    Time frame: 1 month

    P-wave attenuation: P-wave amplitude decrease (between pre-MRI follow-up and 1-month follow-up) exceeding 50% or a P-wave amplitude at 1-month follow-up smaller or equal to 1.5 mV

  4. Primary hypothesis 5: R-wave sensing attenuation

    Time frame: 1 month

    R-wave attenuation: R-wave amplitude decrease (between pre-MRI follow-up and 1-month follow-up) exceeding 50% or a R-wave amplitude at 1-month follow-up smaller or equal to 5.0 mV.

Sponsors and collaborators

Lead sponsor

Biotronik SE & Co. KG

Industry

Registry information

Official study title

Master Study for the MRI Compatibility of the EVIA/ENTOVIS Pacemaker in Combination With Safio S Pacemaker Lead

Acronym: ProMRIAFFIRM

Important dates

Study start
2012
Primary completion
2013
Study completion
2013
First posted
Oct 27, 2011
Registry last updated
Nov 28, 2013

OpenTrials presents study information sourced from ClinicalTrials.gov. The official registry record should be consulted for the latest information.

View the official ClinicalTrials.gov record (opens in a new tab)

This listing is for discovery and informational purposes only. It is not medical advice, does not guarantee that a study is recruiting, and does not determine eligibility. Contact the study team and a qualified healthcare professional when considering participation.

Published trials that share one or more normalized conditions with this study.