Shanghai Fourth People's Hospital Tongji University
Shanghai, Shanghai Municipality, China
Location status: Recruiting
NCT Number: NCT07257367
This observational study aims to evaluate the accuracy and safety of pressure sensing sheath blood pressure monitoring compared to traditional invasive and non-invasive blood pressure monitoring methods during neuroendovascular interventional procedures.
The study will enroll 50 adult patients undergoing elective neuroendovascular procedures requiring general anesthesia and continuous invasive blood pressure monitoring. Blood pressure will be simultaneously measured using three methods: (1) pressure sensing sheath, (2) radial arterial line, and (3) non-invasive cuff monitoring.
The primary outcome is the accuracy of blood pressure measurements from the pressure sensing sheath compared to radial arterial line measurements. Secondary outcomes include the incidence of access site complications, procedure duration, patient comfort scores, and cost-effectiveness analysis.
This prospective, single-center study will be conducted at Shanghai Fourth People's Hospital Affiliated to Tongji University from August 2025 to May 2027.
Interested in participating?
Request Info18 year and older
All sexes
Observational
Shanghai, Shanghai Municipality, China
Location status: Recruiting
Real-time, accurate hemodynamic monitoring is crucial during various interventional procedures. Currently, the clinical "gold standard" for continuous invasive arterial blood pressure monitoring is achieved through peripheral arterial catheterization (typically radial artery) connected to a pressure transducer, known as Radial Artery Catheterization (RAC). Although RAC provides beat-to-beat blood pressure data, it has several inherent limitations.
First, RAC insertion is an additional invasive procedure requiring extra time and technical skill, potentially delaying the start of the primary procedure. Studies have shown that RAC insertion requires an average of 10.7 minutes of additional time, with delays exceeding 80 minutes possible in complex cases. Second, the catheterization process may cause patient discomfort; research indicates that approximately 31.6% of patients experience pain after RAC insertion, with about 30% finding the pain bothersome. Additionally, RAC is associated with various potential complications, including radial artery occlusion (incidence approximately 5.5%), hand ischemia, infection, and thrombosis.
As an alternative, non-invasive blood pressure (NIBP) monitoring is widely used due to its convenience and safety. However, NIBP provides only intermittent readings and may fail to capture critical blood pressure fluctuations in rapidly changing hemodynamic scenarios. This is particularly important in neurointerventional procedures, where real-time blood pressure monitoring is essential for preventing and managing complications such as vasospasm and thrombosis.
### Pressure Sensing Sheath Technology
Given the limitations of traditional monitoring methods, pressure sensing sheath technology has emerged as an innovative blood pressure monitoring approach. This technology integrates a miniature pressure sensor within the vascular access sheath, enabling continuous invasive blood pressure monitoring while establishing vascular access. Theoretically, this approach can simultaneously address the time consumption, patient discomfort, and monitoring discontinuity associated with traditional methods.
Internationally, pressure sensing sheath technology, represented by EndoPhys Corporation, has received U.S. FDA 510(k) clearance and entered clinical use. Purdy et al. first published accuracy validation research on pressure sensing sheath technology in 2017. Froehler et al. completed the first prospective controlled trial (Clinical Trial Registration Number: NCT03239847) in 2018, initially confirming the clinical value of this technology in neurointerventional procedures. However, existing studies are primarily single-center, small-sample investigations focused mainly on European and American populations, with a lack of randomized controlled trial evidence.
### Study Rationale and Innovation
#### Filling Evidence Gaps
Currently, there is a lack of high-quality prospective randomized controlled trial evidence to systematically verify whether pressure sensing sheath monitoring is non-inferior to the gold standard RAC in accuracy, and to comprehensively compare safety, procedural efficiency, and patient-physician satisfaction. Existing international studies have relatively small sample sizes (20-40 cases), and their external validity and generalizability require further verification.
Vascular anatomical structures and hemodynamic characteristics show certain racial differences. Vascular diameter, elasticity, and blood pressure variability patterns in the Chinese population may differ from those in European and American populations. This study will be the first to systematically evaluate the accuracy and safety of pressure sensing sheath technology in a Chinese population, providing scientific evidence for establishing application standards and operational specifications suitable for Chinese clinical practice.
This is a prospective, single-center, self-controlled randomized, non-inferiority clinical study.
The study will enroll patients undergoing elective transradial interventional procedures requiring continuous invasive blood pressure monitoring (meeting inclusion criteria without exclusion criteria). Using a self-controlled randomized design, each patient will simultaneously receive both pressure sensing sheath blood pressure monitoring (experimental group) and traditional radial arterial line invasive blood pressure monitoring combined with standard blood pressure cuff monitoring (control group). The primary study endpoint is at 7 days. By synchronously comparing the performance of both monitoring methods in the same patient, individual differences are eliminated, demonstrating that pressure sensing sheath monitoring is non-inferior to traditional radial arterial line invasive blood pressure monitoring systems, thereby providing a superior monitoring option for clinical practice.
### Randomization Scheme
Laterality Randomization: A random sequence will be generated to randomly determine whether the pressure sensing sheath monitoring system is inserted into the left or right radial artery, with the control group monitoring system inserted into the contralateral radial artery.
Monitoring Time Point Randomization: Block randomization design will be used to randomly determine specific blood pressure measurement time points within preset monitoring time windows, ensuring time synchronization and randomness of monitoring for both groups.
Experimental Group: Pressure sensing sheath monitoring system (inserted via radial artery)
Control Group: Traditional radial arterial line invasive blood pressure monitoring system combined with standard blood pressure cuff monitoring
---
## STUDY POPULATION
This study primarily targets patients requiring transradial interventional procedures with continuous invasive blood pressure monitoring. Disease diagnostic criteria include:
Indications for Interventional Procedures: According to relevant clinical guidelines and expert consensus, diseases requiring transradial interventional treatment primarily include acute cerebral infarction, aneurysms, arteriovenous malformations, carotid artery stenosis, and other cerebrovascular diseases. Specific diagnostic criteria reference the latest cerebrovascular disease diagnosis and treatment guidelines, including comprehensive evaluation of clinical symptoms, imaging examinations (CT/CTA/MRI/MRA/DSA), and laboratory test results.
Indications for Continuous Invasive Blood Pressure Monitoring: According to clinical needs and standard medical operational procedures, patients requiring radial arterial catheterization for invasive blood pressure monitoring include: interventional procedure patients requiring real-time, accurate blood pressure monitoring to guide treatment; patients with potentially unstable hemodynamics requiring close monitoring; patients requiring precise blood pressure control during procedures to prevent complications.
Mean Arterial Pressure (MAP) Agreement Analysis:
Secondary Efficacy Endpoints:
Secondary Safety Endpoints:
---
## STUDY PROCEDURES
### Visit Schedule
The study consists of 4 visits: Visit 1 (enrollment visit), Visit 2 (intraoperative visit), Visit 3 (24 hours post-procedure), and Visit 4 (7 days post-procedure). All visits are clinical follow-ups.
The study employs prospective data collection at 4 predetermined visit time points: Visit 1 (enrollment), Visit 2 (intraoperative), Visit 3 (24 hours post-procedure), and Visit 4 (7 days post-procedure). Standardized Case Report Form (CRF) templates ensure data collection consistency and completeness. Intraoperative blood pressure monitoring data will be synchronously collected to ensure temporal matching between pressure sensing sheath monitoring and traditional radial arterial line invasive blood pressure monitoring.
### Data Management Plan
Detailed standard operating procedures for data collection will be established, clearly defining data collection content and requirements for each visit time point. For primary endpoint data, double data entry will be employed to ensure data accuracy. Data collection checklists will be established to ensure important data items are not missed. For blood pressure monitoring data, strict synchronous measurement standards will be established to ensure both monitoring methods are measured at the same time points under identical conditions.
Multiple levels of data quality control measures will be implemented. Real-time data monitoring systems will validate and quality-check intraoperative blood pressure monitoring data. Data quality checkpoints will regularly verify completeness, accuracy, and consistency, with timely identification and correction of missing or abnormal data. Data anomaly identification and processing workflows will verify and correct abnormal data promptly. Source data verification mechanisms will ensure consistency between CRF data and source documents, guaranteeing data authenticity and traceability.
---
## BIAS CONSIDERATIONS
This study fully considers various potential biases and control measures during design and implementation. To control information bias, the study employs standardized equipment calibration to reduce measurement errors, with all pressure monitoring equipment calibrated before use to ensure measurement accuracy, and standardized training to ensure data collection consistency. For selection bias, subjects are screened strictly according to inclusion and exclusion criteria to ensure study population homogeneity, with comprehensive follow-up plans to minimize loss to follow-up and detailed recording of reasons for incomplete visits.
Confounding bias control is a key feature of this study, employing a self-controlled randomized design where each patient simultaneously receives both blood pressure monitoring strategies. By synchronously comparing the performance of both monitoring methods in the same patient, individual differences are effectively eliminated. Additionally, synchronous measurement of blood pressure data from both monitoring methods during the procedure eliminates the impact of temporal factors on results. To avoid outcome-driven bias, the study protocol predefines the primary endpoint analysis method as Bland-Altman agreement analysis, with non-inferiority criterion set as 95% limits of agreement within ±10 mmHg, avoiding post-hoc selection of the most favorable analysis method.
---
## STATISTICAL ANALYSIS PLAN
### Sample Size Estimation
The target sample size for this study is 50 cases. This sample size was not determined through statistical calculation but was comprehensively considered based on relevant literature and study design characteristics. According to FDA guidance on medical device clinical trials, non-inferiority studies evaluating medical device accuracy can have relatively small sample sizes when using self-controlled designs.
Referring to previous research on pressure sensing sheath technology, Purdy et al.'s 2017 accuracy validation study enrolled 20 patients, and Froehler et al.'s 2018 prospective controlled trial (Clinical Trial Registration Number: NCT03239847) enrolled 40 patients, both confirming the clinical value of this technology. Based on sample size requirements for Bland-Altman agreement analysis, relevant statistical literature recommends a minimum of 30-50 samples for methodological comparison studies to obtain reliable agreement assessment results.
Since this study employs a self-controlled randomized design where each patient simultaneously receives both blood pressure monitoring strategies as their own control, individual variation is effectively eliminated, providing higher statistical power compared to traditional independent sample designs. Considering single-center study feasibility, expected study duration, and enrollment rate of interventional procedure patients, 50 cases can both meet statistical requirements and have good operability. This sample size references conventions from international similar medical device accuracy assessment studies and can provide sufficient evidence support for the primary study endpoint.
### Data Set Definitions
According to study design and analysis objectives, the following data sets are defined:
Efficacy Analysis Set: All subjects meeting inclusion criteria, without exclusion criteria, who complete synchronous intraoperative blood pressure monitoring. This data set is used for primary endpoint and secondary efficacy endpoint analysis.
Safety Analysis Set: All subjects receiving at least one blood pressure monitoring method, used for safety endpoint analysis, including monitoring-related complication incidence and patient-reported puncture site pain and discomfort scores.
If the target population for analysis is a subset of the data set, subsets will be marked as corresponding target populations for subgroup analysis based on different blood pressure ranges or procedure types.
### Missing Data Handling
This study employs prospective design with strict visit arrangements and data quality control measures to minimize data missingness. Primary analysis uses complete case analysis, analyzing only subjects with complete paired blood pressure monitoring data to ensure synchronous measurement data completeness and comparability.
For subjects with missed visits, investigators will record in detail reasons for incomplete visits and make efforts to obtain other relevant information from subjects. For lost-to-follow-up subjects, available clinical data or vital signs will continue to be collected from interventional centers, referring hospitals, general practitioners, etc., according to protocol. Lost-to-follow-up subjects will not be replaced. Sensitivity analysis will employ different missing data handling methods for comparison, including last observation carried forward, to evaluate the impact of missing data on study conclusions.
### Descriptive Analysis
Comprehensive descriptive analysis will be performed on all collected variables to characterize main variable features. Continuous variables will be described using mean ± standard deviation or median (interquartile range) according to data distribution characteristics. Categorical variables will be described using frequencies and percentages.
Baseline variable descriptive analysis includes subject demographic characteristics (name, gender, age, height, weight, body mass index), medical history (smoking history, drinking history, family history, hypertension, diabetes history, peripheral artery disease history, cerebrovascular disease history, coronary heart disease history, PCI history, dyslipidemia history, etc.), clinical indicators (blood pressure, heart rate, complete blood count, coagulation function, liver function, kidney function, electrolytes, blood glucose, electrocardiogram), imaging examination results, and concomitant treatment status. Endpoint variables including blood pressure monitoring data, procedural efficiency indicators, and safety events will also undergo corresponding descriptive analysis.
### Primary Analysis
Hypotheses:
Primary analysis method employs Bland-Altman agreement analysis, calculating mean and standard deviation of differences between the two methods, creating Bland-Altman scatter plots, calculating 95% limits of agreement, and evaluating fixed and proportional bias. Non-inferiority judgment criterion is 95% limits of agreement within ±10 mmHg.
Simultaneously, agreement evaluation will calculate Lin's concordance correlation coefficient and Pearson correlation coefficient, and create equivalence plots. As a reference comparison, agreement between pressure sensing sheath monitoring and standard blood pressure cuff monitoring will be synchronously evaluated. Since a self-controlled randomized design is employed where each patient serves as their own control, effectively eliminating inter-individual differences, primary analysis does not require adjustment for confounding factors.
### Sensitivity Analysis
To evaluate study conclusion robustness, multiple sensitivity analyses will be performed, including:
Safety analysis will employ descriptive statistical methods, analyzed according to adverse event definitions and grading (NCI-CTCAE 4.0). Primary analysis focuses on monitoring-related complication incidence, including vasospasm, hematoma, infection, thrombosis, pseudoaneurysm, etc., as well as patient-reported puncture site pain and discomfort scores (VAS score).
Complication incidence and severity will be analyzed by visit time point, including intraoperative complications (puncture-related immediate complications, puncture failure rate, significant hemodynamic changes), postoperative complications within 24 hours (delayed bleeding, hematoma, vasospasm or dissection, vascular occlusion, abnormal puncture site healing), and short-term follow-up complications (7 days post-procedure: infection, vascular occlusion, pseudoaneurysm, persistent neurological impairment).
Safety event analysis will use frequencies and percentages for description. When necessary, Fisher's exact test or McNemar's test (for paired data) will be used to compare differences in monitoring method-related complication incidence.
---
## QUALITY CONTROL
Quality control objectives for this study align with ICH guidelines, ensuring scientific validity, completeness, accuracy, and traceability of study data, with particular attention to quality control of blood pressure monitoring data precision and synchronization.
### Personnel Training
All medical staff participating in the study will receive standardized training in pressure sensing sheath monitoring equipment operation to ensure operational consistency and accuracy. Training content includes: pressure sensing sheath monitoring system operational procedures, standard operations for traditional radial arterial line blood pressure monitoring, standardized data collection processes, adverse event identification and reporting, and study protocol requirements and procedures.
Training will employ a combination of theoretical instruction and practical operation to ensure all research personnel proficiently master equipment operation skills, with effectiveness confirmed through assessment.
### Data Quality Control
A comprehensive data quality control system will be established with regular data quality checks to ensure data completeness and accuracy. Real-time data monitoring will record and validate intraoperative blood pressure monitoring data in real-time, ensuring data collection accuracy and completeness. Data completeness checks will regularly verify completeness and logical consistency, promptly identifying and correcting missing or abnormal data. Data consistency verification will compare data consistency across different visit time points, ensuring data continuity and reliability. Key data validation will perform double verification of primary endpoint data, ensuring core data accuracy. Source data verification will ensure consistency between data and source documents, guaranteeing data authenticity and traceability.
### Equipment Calibration
All pressure monitoring equipment must be calibrated before use to ensure measurement accuracy meets study requirements. Equipment standardization employs uniformly specified monitoring equipment to ensure consistent equipment performance. Regular calibration according to equipment manual requirements calibrates pressure transducers regularly, establishing equipment calibration archives. Calibration records detail each calibration time, results, and operator, ensuring calibration process traceability. Equipment maintenance ensures all monitoring equipment is in good working condition, with regular equipment maintenance and upkeep, and timely handling of equipment failures.
---
## ETHICS AND REGULATORY CONSIDERATIONS
This protocol, written informed consent form, and materials directly related to subjects must be submitted to the Ethics Committee and receive written Ethics Committee approval before formally commencing the study. Investigators must submit continuing review reports one month before ethics approval letter expiration to apply for approval extension.
Upon study suspension and/or completion, investigators must notify the Ethics Committee in writing. Investigators must promptly report all changes occurring in study work to the Ethics Committee (such as protocol and/or informed consent form amendments), and must not implement these changes without Ethics Committee approval, unless the changes are made to eliminate obvious and immediate risks to subjects. In such cases, the Ethics Committee will be notified.
### Informed Consent
Investigators must provide subjects or their legal representatives with an easily understandable Ethics Committee-approved informed consent form and allow subjects or their legal representatives sufficient time to consider the study. Subjects may not be enrolled before obtaining signed written informed consent from subjects. During subject participation, subjects will be provided with all updated versions of informed consent forms and written information. Informed consent forms should be retained as important clinical trial documents for inspection.
### Confidentiality Measures
Results from this project research may be published in medical journals, but personal information will be kept confidential according to legal and regulatory requirements. Unless required by relevant laws, patient personal information will not be disclosed. When necessary, government regulatory authorities, hospital ethics committees, and related personnel may inspect patient data according to regulations.
---
Study Period: September 15, 2025 to May 31, 2027
Estimated Timeline:
This will be the first prospective, self-controlled randomized trial in a Chinese population to systematically evaluate pressure sensing sheath blood pressure monitoring technology. Results will provide high-quality evidence for clinical application of this innovative monitoring technology, potentially improving blood pressure monitoring in interventional procedures, reducing patient discomfort and complications, and improving procedural efficiency. If non-inferiority is confirmed, this technology could become an important alternative for blood pressure monitoring in interventional procedures.
Healthy volunteers accepted: No
Only the study team can determine whether someone qualifies for participation.
Inclusion criteria
Exclusion criteria
Continuous invasive blood pressure monitoring via pressure sensing sheath inserted through radial artery during interventional procedure. The pressure sensing sheath integrates a miniature pressure sensor within the vascular access sheath, enabling simultaneous vascular access and continuous blood pressure monitoring.
Traditional invasive blood pressure monitoring via radial arterial catheterization (RAC) on the contralateral radial artery, combined with standard non-invasive blood pressure cuff monitoring. This serves as the gold standard comparator for blood pressure measurement accuracy.
Time frame: Intraoperative (duration of procedure, approximately 1-4 hours)
Agreement analysis of mean arterial pressure (MAP) measurements between pressure sensing sheath monitoring and radial arterial line monitoring using Bland-Altman method. Non-inferiority criterion: 95% limits of agreement within ±10 mmHg. Synchronous comparison of both methods in the same patient to eliminate individual differences.
Time frame: Intraoperative (duration of procedure, approximately 1-4 hours)
Bland-Altman analysis of systolic (SBP) and diastolic (DBP) blood pressure agreement between the two monitoring methods. Calculate Pearson correlation coefficient and Lin's concordance correlation coefficient.
Time frame: From patient entry to operating room until completion of procedure setup (approximately 10-30 minutes)
Time required to establish stable blood pressure monitoring from puncture initiation, procedure preparation time, and impact on total procedure duration.
Time frame: From procedure start through 7 days post-procedure
Composite of puncture-related complications including: vasospasm, hematoma, infection, thrombosis, pseudoaneurysm formation, vessel occlusion, and neurological impairment. Assessed at multiple time points: intraoperative, 24 hours post-procedure, and 7 days post-procedure.
Time frame: 24 hours and 7 days post-procedure
Visual Analog Scale (VAS) pain score (0-10 points) for puncture site discomfort and impact on daily activities.
Contact information is provided by the study sponsor or research team.
Shanghai Fourth People's Hospital Tongji University
Other
Accuracy and Safety Assessment of Pressure Sensing Sheath Blood Pressure Monitoring Compared to Traditional Invasive and Non-invasive Blood Pressure Monitoring in Interventional Procedures: A Prospective, Single-Center, Self-Controlled Randomized Study
OpenTrials presents study information sourced from ClinicalTrials.gov. The official registry record should be consulted for the latest information.
View the official ClinicalTrials.gov record (opens in a new tab)This listing is for discovery and informational purposes only. It is not medical advice, does not guarantee that a study is recruiting, and does not determine eligibility. Contact the study team and a qualified healthcare professional when considering participation.
Published trials that share one or more normalized conditions with this study.
NCT07240831
Blood Pressure, Blood Pressure Monitoring
Birmingham, Alabama, United States
View Trial DetailsNCT07556354
BMI, Blood Cholesterol
Birmingham, Alabama, United States
View Trial DetailsNCT07457970
Blood Pressure Monitoring, Cardiovascular Diseases
Columbia, South Carolina, United States
View Trial DetailsNCT07482930
Autonomic Nervous System (ANS) Functioning and Mood State, Blood Pressure Monitoring
Sinop, Turkey (Türkiye)
View Trial Details