NCT Number: NCT05357716

Preliminary Research Evaluation for Ambulatory Leadless Electrocardiogram Recorder Trial

The clinical study is intended to implement and test the collection and analysis of paired ECG data from the HeartWatch and two comparator devices, an Event Recorder and Holter. The HeartWatch is indicated for extended diagnostic evaluation of patients with transient symptoms of possible cardiac origin such as syncope and palpitations, as well as patients at risk for arrhythmias, but without significant symptoms. While the use of the device itself can be managed by both healthcare professionals and patients, interpretation of the collected data for diagnosis is restricted to healthcare professionals, as the data must be loaded and analyzed separately after recording. Patients indicated for extended ambulatory ECG testing will be approached for participation. Subjects will wear the HeartWatch and an Event Recorder for up to 72 hours or the HeartWatch and Holter monitor for up to 48 hours. Subjects will be asked to document their activities (standing, sitting, walking, exercise, or laying down). Event recorder subjects will collect user-triggered and auto-triggered data, while Holter subjects will record diary information on their activities and any relevant symptoms. Adverse events and user preference for one device versus the other being used in that arm will also be collected from all subjects. Paired ECG data from the HeartWatch and comparator devices is expected to be similar in terms of the events documented and the proportion of data that is clinically interpretable.

Completed

Looking for future studies?

Notify Me

Key information

Who can participate

Healthy volunteers accepted: No

Only the study team can determine whether someone qualifies for participation.

Inclusion criteria

  • At least 22 years of age at time of consent
  • Clinically-indicated for an ambulatory Event Recorder or Holter monitor
  • Able to follow the protocol
  • Provision of written-informed consent

Exclusion criteria

  • Known allergy to any component of the Event Recorder
  • Known allergy to any component of the Holter monitor
  • Known allergy to any component of the HeartWatch
  • Dextrocardia

Treatment and study plan

HeartWatch

Device

The HeartWatch arm band provides continuous, high-quality single lead electrocardiograph (ECG) data in a variety of environments. The ECG data is transmitted via Bluetooth and stored on the user's smartphone. It is then uploaded to the cloud for subsequent analysis using either a commercial ECG software package or a customized software package from HelpWear.

Event Recorder

Device

A standard 1 or 2-lead Sirona Event Recorder will be used to collect high-resolution data. Subjects will be asked to wear the clinical monitor for at least 48 hours and no longer than 84 hours. ECG signals will be continuously collected as long as the skin patches are attached to the skin and the recorder is properly connected to the wires.

Holter monitor

Device

A standard 5-lead SEER Holter will be used to collect 3-channels of high-resolution data. Subjects will be asked to wear the clinical, conventional monitor for at least 36 hours and no longer than 60 hours. During this time ECG signals will be continuously collected.

Primary outcomes

  1. Pairing HeartWatch and Event Recorder

    Time frame: Up to 6 months, with 3 months of analysis time (9 months total)

    Can paired user-triggered (tagged) and auto-triggered ECG data from the HeartWatch and Event Recorder be successfully collected?

  2. Pairing HeartWatch and Holter

    Time frame: Up to 6 months, with 3 months of analysis time (9 months total)

    Can paired continuous ECG data from the HeartWatch and Holter monitor be successfully collected?

Secondary outcomes

  1. Body Position and Activity

    Time frame: Up to 6 months, with 3 months of analysis time (9 months total)

    Does body position and activity affect the quality of the HeartWatch, Event Recorder, and Holter data?

Other outcomes

  1. Adverse Events

    Time frame: Up to 6 months, with 3 months of analysis time (9 months total)

    Incidence of adverse events among the HeartWatch and comparator devices and the types of adverse events

  2. User Preferences

    Time frame: Up to 6 months, with 3 months of analysis time (9 months total)

    User preferences of subjects with the HeartWatch versus the comparator devices

Sponsors and collaborators

Lead sponsor

HelpWear Inc.

Industry

Registry information

Acronym: PRE-ALERT

Important dates

Study start
2020
Primary completion
2023
Study completion
2023
First posted
May 3, 2022
Registry last updated
Oct 4, 2023

OpenTrials presents study information sourced from ClinicalTrials.gov. The official registry record should be consulted for the latest information.

View the official ClinicalTrials.gov record (opens in a new tab)

This listing is for discovery and informational purposes only. It is not medical advice, does not guarantee that a study is recruiting, and does not determine eligibility. Contact the study team and a qualified healthcare professional when considering participation.

Published trials that share one or more normalized conditions with this study.