This is a prospective observational diagnostic accuracy and feasibility study conducted in the Central Denmark Region.
Adult patients assessed by participating emergency medical services (EMS) units with symptoms suggestive of acute coronary syndrome (ACS), without ST-segment elevation on the prehospital electrocardiogram, and managed with planned transport to hospital are eligible for inclusion.
During routine EMS assessment, a venous blood sample (<0.5 mL) is obtained and analyzed using a portable point-of-care device for measurement of high-sensitivity cardiac troponin I (hs-TnI). The measurement is performed according to manufacturer instructions and local procedures. Troponin values are recorded for research purposes only and are not used to guide clinical decision-making. EMS personnel are not provided with interpretation guidance or clinical thresholds for hs-TnI.
Clinical assessment, treatment, triage, and transport decisions follow standard regional protocols and are independent of study participation.
Clinical data are obtained from prehospital records and hospital electronic health records. Final diagnoses, including acute coronary syndrome, are determined based on standard clinical evaluation, including serial in-hospital troponin measurements and other diagnostic procedures as clinically indicated.
The primary analysis evaluates the diagnostic performance of prehospital hs-TnI for identifying acute coronary syndrome and predicting major adverse cardiac events within 30 days. Diagnostic performance will be reported using sensitivity, specificity, positive and negative predictive values, likelihood ratios, and area under the receiver operating characteristic curve.
Secondary analyses assess associations between hs-TnI concentrations and clinical outcomes, including hospital admission, length of stay, coronary angiography, coronary revascularization, and 30-day mortality.
Feasibility outcomes include the proportion of successful blood sampling, the proportion of valid measurements, and the time from patient contact to availability of the biomarker result.
The study is designed to generate feasibility and diagnostic accuracy data to inform the design and statistical planning of future interventional studies in the prehospital setting.
The study has been assessed by The Central Denmark Region Committees on Health Research Ethics, which determined that the project does not require notification or approval under Danish legislation.