Post-Market Clinical Investigation of the Clareon® IOL

The primary purpose of this study is to evaluate the long-term (3 years) visual acuity and adverse event outcomes for the Clareon® Intraocular Lens (IOL). A comparison to historical safety and performance endpoint (SPE) rates as reported in the European Standard International Organization for Standardization (EN ISO) 11979-7:2014 will be conducted at one year. The secondary purpose of this study is to evaluate the visual acuity outcomes with the Clareon IOL at Years 2 and 3.

Completed

Looking for future studies?

Notify Me

Key information

Age range

22 year and older

Sex eligibility

All sexes

Study type

Interventional

Phase

Not applicable

Primary location

Alcon Investigative Site, Footscray, Victoria, Australia

Loading trial locations.

About this study

Subjects will attend a total of 12 study visits over a period of approximately 36 months: One preoperative screening visit, two operative visits, and nine post-implantation visits. Both eyes will be implanted, with the second eye implantation to occur 2-21 days after the first. Primary endpoint data will be collected at the Year 1 visit (Day 330-420 post-implantation from second eye surgery).

Who can participate

Healthy volunteers accepted: No

Only the study team can determine whether someone qualifies for participation.

Key Inclusion Criteria:

  • Diagnosed with bilateral cataracts
  • Planned small incision cataract removal surgery
  • Able to comprehend and willing to sign a statement of informed consent and complete all required post-implantation visits

Key Exclusion Criteria:

  • Subjects who may reasonably be expected to require ocular surgical treatment or refractive surgical procedures at any time during the study
  • Certain eye conditions, as specified in the protocol, including but not limited to glaucoma, diabetic retinopathy, diabetic macular edema, and macular degeneration
  • Pregnant or lactating, current or planned during the course of the study

Other protocol-defined inclusion/exclusion criteria may apply.

Treatment and study plan

Clareon® IOL

Device

Foldable, single-piece posterior chamber IOL intended for long-term use over the lifetime of the cataract patient. This device is CE-marked (European Conformity) in the countries with participating Investigators.

Other names: Model SY60WF

cataract surgery

Procedure

Routine small incision cataract surgery with unilateral IOL implantation

Primary outcomes

  1. Percentage of First Implanted Eyes Achieving Best-Corrected Distance Visual Acuity (BCDVA) of 0.3 Logarithm Minimum Angle of Resolution (logMAR) or Better at 1 Year Post-Implantation

    Time frame: Year 1 post-implantation from second eye surgery

    Visual acuity (VA) was tested for the first implanted eye using the correction obtained from the manifest refraction and 100% contrast electronic Early Treatment Diabetic Retinopathy Study (ETDRS) charts at a distance of 4 meters (m) from the eye. VA was measured in logarithm of the minimum angle of resolution (logMAR), with 0.02 logMAR increment corresponding to a single letter on an ETDRS chart. A BCDVA of 0.3 logMAR is equivalent to 20/40 Snellen visual acuity. VA was compared to the historical safety and performance endpoint (SPE) of 92.5% [as reported in the European Standard International Organization for Standardization (EN ISO) 11979-7:2014].

  2. Percentage of Second Implanted Eyes Achieving BCDVA of 0.3 logMAR or Better at 1 Year Post-Implantation

    Time frame: Year 1 post-implantation from second eye surgery

    VA was tested for the second implanted eye using the correction obtained from the manifest refraction and 100% contrast electronic ETDRS charts at a distance of 4 m from the eye. VA was measured in logMAR, with 0.02 logMAR increment corresponding to a single letter on an ETDRS chart. A BCDVA of 0.3 logMAR is equivalent to 20/40 Snellen visual acuity. VA was compared to the historical SPE of 92.5% (EN ISO 11979-7:2014).

Secondary outcomes

  1. Percentage of First Implanted Eyes Achieving BCDVA of 0.3 logMAR or Better at 2 Years Post-Implantation

    Time frame: Year 2 post-implantation from second eye surgery

    VA was tested for the first implanted eye using the correction obtained from the manifest refraction and 100% contrast electronic ETDRS charts at a distance of 4 m from the eye. VA was measured in logMAR, with 0.02 logMAR increment corresponding to a single letter on an ETDRS chart. A BCDVA of 0.3 logMAR is equivalent to 20/40 Snellen visual acuity. No hypothesis testing was prespecified for this endpoint.

  2. Percentage of Second Implanted Eyes Achieving BCDVA of 0.3 logMAR or Better at 2 Years Post-Implantation

    Time frame: Year 2 post-implantation from second eye surgery

    VA was tested for the second implanted eye using the correction obtained from the manifest refraction and 100% contrast electronic ETDRS charts at a distance of 4 m from the eye. VA was measured in logMAR, with 0.02 logMAR increment corresponding to a single letter on an ETDRS chart. A BCDVA of 0.3 logMAR is equivalent to 20/40 Snellen visual acuity. No hypothesis testing was prespecified for this endpoint.

  3. Percentage of First Implanted Eyes Achieving BCDVA of 0.3 logMAR or Better at 3 Years Post-Implantation

    Time frame: Year 3 post-implantation from second eye surgery

    VA was tested for the first implanted eye using the correction obtained from the manifest refraction and 100% contrast electronic ETDRS charts at a distance of 4 m from the eye. VA was measured in logMAR, with 0.02 logMAR increment corresponding to a single letter on an ETDRS chart. A BCDVA of 0.3 logMAR is equivalent to 20/40 Snellen visual acuity. No hypothesis testing was prespecified for this endpoint.

  4. Percentage of Second Implanted Eyes Achieving BCDVA of 0.3 logMAR or Better at 3 Years Post-Implantation

    Time frame: Year 3 post-implantation from second eye surgery

    VA was tested for the second implanted eye using the correction obtained from the manifest refraction and 100% contrast electronic ETDRS charts at a distance of 4 m from the eye. VA was measured in logMAR, with 0.02 logMAR increment corresponding to a single letter on an ETDRS chart. A BCDVA of 0.3 logMAR is equivalent to 20/40 Snellen visual acuity. No hypothesis testing was prespecified for this endpoint.

  5. Uncorrected Distance Visual Acuity (UCDVA) at 1 Year Post-Implantation - First Implanted Eye

    Time frame: Year 1 post-implantation from second eye surgery

    VA was tested for the first implanted eye without visual correction using 100% contrast electronic ETDRS charts at a distance of 4 m from the eye. +0.25 diopter (D) spherical power was applied to correct for optical infinity. VA was measured in logMAR, with 0.02 logMAR increment corresponding to a single letter on an ETDRS chart. A UCDVA of 0.0 logMAR is equivalent to 20/20 Snellen visual acuity, with a lower numeric value representing better visual acuity. No hypothesis testing was prespecified for this endpoint.

  6. UCDVA at 1 Year Post-Implantation - Second Implanted Eye

    Time frame: Year 1 post-implantation from second eye surgery

    VA was tested for the second implanted eye without visual correction using 100% contrast electronic ETDRS charts at a distance of 4 m from the eye. +0.25 D spherical power was applied to correct for optical infinity. VA was measured in logMAR, with 0.02 logMAR increment corresponding to a single letter on an ETDRS chart. A UCDVA of 0.0 logMAR is equivalent to 20/20 Snellen visual acuity, with a lower numeric value representing better visual acuity. No hypothesis testing was prespecified for this endpoint.

  7. UCDVA at 2 Years Post-Implantation - First Implanted Eye

    Time frame: Year 2 post-implantation from second eye surgery

    VA was tested for the first implanted eye without visual correction using 100% contrast electronic ETDRS charts at a distance of 4 m from the eye. +0.25 D spherical power was applied to correct for optical infinity. VA was measured in logMAR, with 0.02 logMAR increment corresponding to a single letter on an ETDRS chart. A UCDVA of 0.0 logMAR is equivalent to 20/20 Snellen visual acuity, with a lower numeric value representing better visual acuity. No hypothesis testing was prespecified for this endpoint.

  8. UCDVA at 2 Years Post-Implantation - Second Implanted Eye

    Time frame: Year 2 post-implantation from second eye surgery

    VA was tested for the second implanted eye without visual correction using 100% contrast electronic ETDRS charts at a distance of 4 m from the eye. +0.25 D spherical power was applied to correct for optical infinity. VA was measured in logMAR, with 0.02 logMAR increment corresponding to a single letter on an ETDRS chart. A UCDVA of 0.0 logMAR is equivalent to 20/20 Snellen visual acuity, with a lower numeric value representing better visual acuity. No hypothesis testing was prespecified for this endpoint.

  9. UCDVA at 3 Years Post-Implantation - First Implanted Eye

    Time frame: Year 3 post-implantation from second eye surgery

    VA was tested for the first implanted eye without visual correction using 100% contrast electronic ETDRS charts at a distance of 4 m from the eye. +0.25 D spherical power was applied to correct for optical infinity. VA was measured in logMAR, with 0.02 logMAR increment corresponding to a single letter on an ETDRS chart. A UCDVA of 0.0 logMAR is equivalent to 20/20 Snellen visual acuity, with a lower numeric value representing better visual acuity. No hypothesis testing was prespecified for this endpoint.

  10. UCDVA at 3 Years Post-Implantation - Second Implanted Eye

    Time frame: Year 3 post-implantation from second eye surgery

    VA was tested for the second implanted eye without visual correction using 100% contrast electronic ETDRS charts at a distance of 4 m from the eye. +0.25 D spherical power was applied to correct for optical infinity. VA was measured in logMAR, with 0.02 logMAR increment corresponding to a single letter on an ETDRS chart. A UCDVA of 0.0 logMAR is equivalent to 20/20 Snellen visual acuity, with a lower numeric value representing better visual acuity. No hypothesis testing was prespecified for this endpoint.

Sponsors and collaborators

Lead sponsor

Alcon Research

Industry

Registry information

Important dates

Study start
2018
Primary completion
2019
Study completion
2021
First posted
Oct 20, 2017
Registry last updated
Aug 21, 2023

OpenTrials presents study information sourced from ClinicalTrials.gov. The official registry record should be consulted for the latest information.

View the official ClinicalTrials.gov record (opens in a new tab)

This listing is for discovery and informational purposes only. It is not medical advice, does not guarantee that a study is recruiting, and does not determine eligibility. Contact the study team and a qualified healthcare professional when considering participation.

Published trials that share one or more normalized conditions with this study.