TREO Abdominal Stent-Graft System
DeviceThe TREO Abdominal Stent-Graft System is intended for the treatment of infrarenal abdominal aortic aneurysms with or without iliac involvement.
NCT Number: NCT04697784
The purpose of this study is to determine the long-term performance of the TREO Abdominal Stent-Graft as a treatment for patients with Infrarenal Abdominal Aortic Aneurysms or Aorto-iliac Aneurysms.
This study is active but is not currently recruiting participants.
18 year and older
All sexes
Interventional
Not applicable
Vascular Surgery Specialists, Phoenix, Arizona, United States
This is a prospective, multi-center, non-randomized, single-arm, post-market, non-interventional study of treatment with the TREO Abdominal Stent-Graft in subjects with Infrarenal Abdominal Aortic Aneurysms or Aorto-iliac Aneurysms. The study will include a minimum of 300 subjects treated with the TREO Abdominal Stent-Graft at up to 55 investigational sites in the US.
The primary objective is the collection of real world safety and effectiveness outcomes of the TREO Abdominal Stent-Graft System in an all-comers population eligible for the endovascular treatment of AAA in routine clinical practice, with emphasis on subjects that experience a device stent-strut or barb fracture.
Healthy volunteers accepted: No
Only the study team can determine whether someone qualifies for participation.
Inclusion criteria
Exclusion criteria
The TREO Abdominal Stent-Graft System is intended for the treatment of infrarenal abdominal aortic aneurysms with or without iliac involvement.
Time frame: Through 5 Years post-procedure
Incidence of fracture in a stent-strut or separation of one or more proximal fixation barbs as confirmed by the Imaging Core Laboratory upon review of post-implant imaging studies.
Time frame: Through 5 Years post-procedure
Incidence of secondary intervention for adverse events related to or caused as a result of a fracture in a stent-strut or separation of one or more proximal fixation barbs. Relatedness to the device will be confirmed by the Clinical Events Committee (CEC) through review of available treatment and procedure records and reports of imaging studies.
Time frame: Through 5 Years post-procedure
Defined as the following:
Time frame: Through 5 Years post-procedure
Incidence of the following Major Adverse Events
Time frame: Through 5 Years post-procedure
Reporting of procedural data collected during implant of the TREO Abdominal Stent-Graft System including: Procedure time (minutes), Fluoroscopy time (minutes), Contrast volume (mL), Access method (i.e. percutaneous, surgical cut down), Length of ICU stay (hours), Length of hospital stay post-procedure (days), Anesthesia Type.
Time frame: Through 5 Years post-procedure
Incidence of adverse events occurring during or following the implant determined by the CEC to be procedure-related.
Time frame: 12 months post-implant
Successful aneurysm treatment is defined as:
Time frame: Through 5 Years post-procedure
Rate of mortality attributed to any causality as confirmed by the CEC.
Time frame: Through 5 Years post-procedure
Rate of mortality attributed to the following causes as confirmed by the CEC: death due to a rupture, death within 30 days or prior to hospital discharge from primary procedure, or death within 30 days or prior to hospital discharge for a secondary procedure designed to treat the original aneurysm.
Time frame: Through 5 Years post-procedure
Incidence of rupture of the native aneurysm sac post-implantation of the endograft as confirmed by the CEC.
Time frame: Through 5 Years post-procedure
Incidence of secondary procedures designed to treat or repair the original aneurysm treated with the TREO Abdominal Stent-Graft System
Time frame: Through 5 Years post-procedure
Incidence of conversion to open surgical AAA repair during the initial EVAR procedure secondary to any procedure-related complications as determined by the CEC.
Time frame: Through 5 Years post-procedure
Incidence of stent-graft occlusion defined as the unintentional obstruction of the vascular/endograft lumen with minor obstruction (0-25%), minimal obstruction (26-74%), moderate obstruction (75-99%) or occlusion (100%) as confirmed by the Imaging Core Lab due to causes such as twisting or kinking of the prosthesis, oversizing and fabric pleating, or failure of the implant to fully open, or to mural thrombus deposition.
Time frame: Through 5 Years post-procedure
Incidence of stenosis or kinking of the TREO Stent-Graft as confirmed by the Imaging Core Lab.
Time frame: Through 5 Years post-procedure
Incidence of changes in the structural integrity in a material component of the stent-graft such as a stent-strut fracture or separation of the proximal fixation barbs as confirmed by the Imaging Core Lab.
Time frame: Through 5 Years post-procedure
Incidence of the increase in the aneurysm sac diameter > 5 mm at post-implant follow-up visits relative to the diameter determined at the first post-procedural imaging study, as confirmed by the Imaging Core Lab.
Time frame: Through 5 Years post-procedure
Incidence of the longitudinal movement of all or part of a stent or attachment system for a distance of >10 mm relative to anatomical landmarks that were determined at the first post-procedural imaging study as confirmed by the Imaging Core Lab.
Time frame: Through 5 Years post-procedure
Incidence of Type I, II, III, IV or V Endoleaks defined as the persistence of blood flow outside the lumen of the stent-graft but within the native aorta or adjacent vascular segment being treated by the stent-graft as determined by the Imaging Core Lab.
Bolton Medical
Industry
Post-Approval Study of the TREO Abdominal Stent-Graft System (P190015) in Patients With Infrarenal Abdominal Aortic and Aorto-iliac Aneurysms
Acronym: TREO PAS
OpenTrials presents study information sourced from ClinicalTrials.gov. The official registry record should be consulted for the latest information.
View the official ClinicalTrials.gov record (opens in a new tab)This listing is for discovery and informational purposes only. It is not medical advice, does not guarantee that a study is recruiting, and does not determine eligibility. Contact the study team and a qualified healthcare professional when considering participation.
Published trials that share one or more normalized conditions with this study.
NCT06578741
Abdominal Aortic Aneurysm, Aneurysm
Stanford, California, United States
View Trial DetailsNCT05864560
Abdominal Aortic Aneurysm, Aneurysm
Pleven, Bulgaria
View Trial DetailsNCT06990048
Abdominal Aortic Aneurysm, Aneurysm
Lebanon, New Hampshire, United States
View Trial DetailsNCT04751578
Abdominal Aortic Aneurysm, Aneurysm
Arnhem, Gelderland, Netherlands
View Trial Details