Clinique Beau Soleil
Montpellier, 34070, France
NCT Number: NCT05055076
This is a monocentric retrospective, open-label clinical study in patients after unilateral or bilateral implantation of hydrophobic acrylic monofocal toric intraocular lens PODEYE TORIC and executed only with standard of care examinations.
The investigational device is CE approved. The investigational device and all study products, including the devices used for the study examinations, have been used within the intended use specifications from the manufacturer. In addition, there were no invasive or other burdening examinations.
The investigator is an accredited and experienced cataract surgeon and researcher.
The primary endpoint is to determine rotational stability in a minimum of 120 eyes of patients with PODEYE TORIC IOL implantation after cataract surgery. Patients will be enrolled for the documentation and evaluation of retrospective data up to 6 months postoperatively.
Looking for future studies?
Notify Me35 year and older
All sexes
Observational
Montpellier, 34070, France
Age-related changes of the proteins in the crystalline lens lead to cataract formation. Cataracts are a common condition in adults over 40 years of age, and surgical replacement of the cataractous lens with an intraocular lens (IOL) remains an effective way to restore vision to cataract patients.
On patients with a low amount of pre-existing corneal astigmatism (defect in the sphericity of the cornea with a focal point depending on the meridian), the implantation of a spherical monofocal IOL is the state- of- the-art when targeting a good correction of distance visual function. However, in case of pre-existing corneal astigmatism, the implantation of spherical IOLs might not be the best solution, as such lenses only correct for spherical errors and leave the astigmatism uncorrected.
Toric IOLs are obtained by adding a cylinder power to the basis optic of a given spherical power, thus correcting for both spherical and cylindrical refractive errors. Toric IOLs provide the opportunity to reduce or eliminate astigmatism, thus offering patients with pre-existing astigmatism the possibility of restoring their distance vision with reduced or no need for the use of spectacles or contact lenses with a cylindrical correction. Not correcting the astigmatism component at the time of cataract surgery is an important cause for not obtaining planned emmetropia following IOL implantation. A previous study found that more than 22% of patients undergoing cataract surgery have substantial corneal astigmatism superior than 1.25 D and would benefit from toric IOL implantation 1).
This study will add to the knowledge on toric IOLs and help improve the care of patients and the outcomes of cataract surgery after toric IOL implantation.
The study is a monocentric, retrospective study to investigate rotational stability and visual performance after mono- or bilateral implantation of PODEYE TORIC intraocular lens following cataract surgery, executed only with standard of care examinations. The study outcomes will be used to confirm the safety performance for a stable IOL positioning of the investigation al device.
Preoperative, surgical and postoperative data (up to 6 months after surgery) will be collected retrospectively by reviewing the patient charts.
The investigational device PODEYE TORIC IOL is CE approved and commercially available in the country where this clinical investigation is being carried out. The investigational device and all study products, including the devices used for the study examinations, have been used within the intended use specifications from the manufacturer. In addition, no invasive or other burdening examinations will occur for the patient.
Healthy volunteers accepted: No
Only the study team can determine whether someone qualifies for participation.
Inclusion criteria
Exclusion criteria
Patients will be contacted consecutively and will be enrolled after the patient informed leaflet has been given to them and after obtaining their consent to the scientific use of their data collected retrospectively (preoperative, surgery, postoperative routine visits). Patients will only be enrolled after the routine 4-6 months visit has been performed.
Time frame: 6 months
Rotational stability is an important parameter on toric IOLs, as postoperative IO L rotation can lead to inappropriate correction of corneal astig matism and decrease in uncorrected distance visual acuity.
The primary study endpoint is the absolute value of rotation between the day of surgery and the postoperative visit 6 months after surgery. The absolute value of rotation should be less than 10° in 90% of the eyes implanted with POD EYE TORIC lens according to safety declarations of EN ISO 11979-7:2018.
Time frame: 6 months
Manifest refraction will be assessed using the spherical equivalent: Manifest refraction Spherical
Equivalent (MRSE). Negative cylinder notation will be used. MRSE is calculated by the following formula:
MRSE = sphere + 1/2 cylinder. Descriptive analysis of sphere, cylinder and MRS E will be reported. Stability of the MRSE will be assessed by summarizing the change in MRSE between adjacent visits ( mean, standard deviation, minimum and maximum, and 95% confidence interval).
Time frame: 6 months
Corrected and Uncorrected Distance Visual Acuity data will be measured and summarized (mean, standard deviation, minimum and maximum) at each visit, as far as available, and the change between visits with a 95% confidence interval will be provided.
Time frame: 6 months
The refractive astig matism predictability achieved with the PhyslO L toric calculator will be assessed and compared to other toric calculators. The prediction error will be calculated as the difference between the measured postoperative manifest refraction and the predicted residual astigmatism.
Beaver-Visitec International, Inc.
Industry
Retrospective Study to Investigate Rotational Stability and Clinical Outcomes After Implantation of a Monofocal Toric Intraocular Lens (IOL)
OpenTrials presents study information sourced from ClinicalTrials.gov. The official registry record should be consulted for the latest information.
View the official ClinicalTrials.gov record (opens in a new tab)This listing is for discovery and informational purposes only. It is not medical advice, does not guarantee that a study is recruiting, and does not determine eligibility. Contact the study team and a qualified healthcare professional when considering participation.
Published trials that share one or more normalized conditions with this study.
NCT06399211
Aphakia, Aphakia, Postcataract
View Trial DetailsNCT04987216
Astigmatism, Cataract
Berlin, Germany
View Trial DetailsNCT04778501
Astigmatism, Cataract
Makati City, Philippines
View Trial DetailsNCT04866719
Astigmatism, Cataract
Alicante, Spain
View Trial Details