Element Materials Technology
Louisville, Colorado, 80027, United States
NCT Number: NCT06980766
The purpose of this study is to provide supporting documentation of the blood pressure performance for the Philips EarlyVue VS30 monitor on the intended adult and pediatric population. The same arm sequential method with dual observer auscultation will be used to collect data.
Looking for future studies?
Notify Me3 year–85 year
All sexes
Observational
Louisville, Colorado, 80027, United States
The Philips EarlyVue VS30 monitor, and approved Philips Gentle Care cuffs will be evaluated in comparison to dual reference auscultation. The study will consist of a minimum of 85 participants, specifically, 35 children aged 3-12 years and 50 participants older than 12 years. The participants enrolled will approximate the distribution of blood pressures and biceps circumferences as outlined in ISO 81060-2:2018/AMD 2:2024. The maximum number of participants enrolled will not exceed 170 participants.
Blood pressure measurements will follow ISO 81060-2:2018/AMD 2:2024, Section 5.2.4.1 Same arm sequential method.
Two trained observers, typically a Registered Nurse (RN), Licensed Practical Nurse (LPN), or Respiratory Therapist (RT) will observe (listen to) the Korotkoff sounds at the artery at the site of cuff placement. The reference blood pressure measurements observed by the auscultators will be performed sequentially with the device under test. The auscultators will complete 1 or 2 initial baseline auscultatory reference blood pressure measurements and device under test measurements. This is followed by 3 to 8 paired blood pressure readings of auscultatory reference blood pressure measurements alternated with device under test measurements. All participants must contribute 3 valid paired data points requiring three DUT readings paired with three reference blood pressure readings. The reference blood pressure readings for analysis will be the average of the reference dual auscultation values before and after DUT reading.
The data included for the final analysis will contain no fewer than 35 participants aged 3-12 years, and no fewer than 50 participants over the age of 12 and a minimum of 255 valid paired observations.
Healthy volunteers accepted: Yes
Only the study team can determine whether someone qualifies for participation.
Inclusion criteria
Exclusion criteria
Manual blood pressure via sphygmomanometer
Or
Non invasive blood pressure via earlyvue VS30 Monitor
Time frame: 2-4 weeks
Noninvasive blood pressure measurement accuracy for the Philips EarlyVue VS30 on the intended population in comparison to the reference sphygmomanometer via same arm sequential method with dual auscultators
Philips Clinical & Medical Affairs Global
Industry
Element Boulder Clinical Investigation Plan Philips Non-Invasive Blood Pressure Validation Study PR 2024-610
OpenTrials presents study information sourced from ClinicalTrials.gov. The official registry record should be consulted for the latest information.
View the official ClinicalTrials.gov record (opens in a new tab)This listing is for discovery and informational purposes only. It is not medical advice, does not guarantee that a study is recruiting, and does not determine eligibility. Contact the study team and a qualified healthcare professional when considering participation.