Clinimark Laboratory
Louisville, Colorado, 80027, United States
NCT Number: NCT02048436
The purpose of this study is to validate the accuracy of the Perminova N4 Monitoring system Respiration Rate technology.
Looking for future studies?
Notify Me18 year–45 year
All sexes
Observational
Louisville, Colorado, 80027, United States
After IRB Approval, eighteen volunteer test subjects will be entered into an accuracy study that is designed to validate accuracy of the respiratory rate technology of the Perminova N4 Monitoring system (Device Under Test). The subjects will be selected to represent a range of body types including small, average, muscular, and large with a range of BMI's.
A range of stable respiratory rates will be elicited from each volunteer test subject. The rates will be approximately 5, 10, 15, 20, 25, 30, 35 and 40 breaths per minute; with some natural variation from these exact numbers.
Each subject will be instrumented with an open system mask that allows for measurement of the end tidal carbon dioxide (EtCO2) respiratory rate and tidal volume. The EtCO2 monitor will be used as the Accuracy Reference Device (Reference) in this study. Performance of the currently released ECG Impedance Respiratory Rate monitor (Acceptance Standard) will be used to establish the acceptance criteria in this study by comparing to Reference.
Healthy volunteers accepted: Yes
Only the study team can determine whether someone qualifies for participation.
Inclusion criteria
Exclusion criteria
Time frame: 1 week
Clinimark, LLC
Other
OpenTrials presents study information sourced from ClinicalTrials.gov. The official registry record should be consulted for the latest information.
View the official ClinicalTrials.gov record (opens in a new tab)This listing is for discovery and informational purposes only. It is not medical advice, does not guarantee that a study is recruiting, and does not determine eligibility. Contact the study team and a qualified healthcare professional when considering participation.
Published trials that share one or more normalized conditions with this study.
NCT06911541
Capnography, Respiratory Rate
Bergen, Norway
View Trial Details