Performance of an Investigational Blood Glucose Monitoring System in a Clinical Setting

The purpose of this study is to evaluate the performance and acceptability of an investigational blood glucose monitoring system, which includes meter and sensor strip.

Completed

Looking for future studies?

Notify Me

Key information

Age range

18 year and older

Sex eligibility

All sexes

Study type

Interventional

Phase

Not applicable

Primary location

AMCR Institute

Escondido, California, 92026, United States

Who can participate

Healthy volunteers accepted: No

Only the study team can determine whether someone qualifies for participation.

Inclusion criteria

  • Males and females, 18 years of age and above
  • Type 1 or type 2 diabetes
  • Currently self-testing blood glucose at home at least twice daily
  • Able to speak, read, and understand English
  • Willing to complete all study procedures

Exclusion criteria

  • Currently Pregnant
  • Hemophilia or any other bleeding disorder
  • Employee of competitive medical device company
  • Cognitive disorder or other condition which, in the opinion of the investigator, would put the person at risk or seriously compromise the integrity of the study

Treatment and study plan

Ninja 2 Investigational Blood Glucose Monitoring System

Device

Untrained subjects with diabetes performed Blood Glucose(BG) tests from the subject's capillary blood obtained from fingerstick, palm, and forearm using the Ninja 2 investigational meter and an investigational sensor. All BG results were compared to a reference laboratory glucose method. Subjects' success at performing basic tasks using only the User Guide were rated by study staff.

Primary outcomes

  1. Numbers of Fingerstick Blood Glucose (BG) Results Within +/- 5 to 20mg/dL (<75 mg/dL) or Within +/- 5% to 20% (>=75 mg/dL) of Laboratory Glucose Method

    Time frame: 2 hours

    Subjects with diabetes tested subject fingerstick blood using an investigational blood glucose monitoring system (BGMS), which included an investigational meter and sensor. BGM results were compared to a reference lab glucose method - Yellow Springs Instrument (YSI) Analyzer. BG results were used to calculate the number of BG results within +/- 5 to 20 mg/dL (for reference BG results <75mg/dL) or +/- 5 to 20% (for reference BG results >=75mg/dL) of the reference method results.

Secondary outcomes

  1. Numbers of Palm Blood Glucose (BG) Results Within +/- 5 to 20 mg/dL (<75 mg/dL) or Within +/- 5% to 20% (>=75 mg/dL) of Laboratory Glucose Method

    Time frame: 2 hours

    Subjects with diabetes used an investigational blood glucose monitoring system (BGMS) with subject capillary blood obtained from the palm. BGM results were compared to a reference lab glucose method - Yellow Springs Instrument (YSI) Analyzer. BG results were used to calculate the number of BG results within +/- 5 to 20 mg/dL (for reference BG results <75mg/dL) or +/- 5% to 20% (for reference BG results >=75mg/dL) of the reference method results.

  2. Number of Subjects Rated as<=3 Performing Basic Meter Tasks (Labeling Comprehension)

    Time frame: 2 hours

    Subjects reviewed the instructions for use (User Guide and Quick Reference Guide) to learn to use the system. Study staff then observed and rated the subjects (1 to 4) on their success at performing the tasks. Scale: 1.Success in performing tasks correctly without assistance. 2.Successful with additional review of User Guide. 3.Successful with additional review and study staff assist similar to review of a specific function during a Customer Service call. 4.Subject did not perform task correctly and study staff intervention was required.

  3. Numbers of Forearm Blood Glucose (BG) Results Within +/- 5 to 20 mg/dL (<75 mg/dL) or Within +/- 5% to 20% (>=75 mg/dL) of Laboratory Glucose Method

    Time frame: 2 hours

    Subjects with diabetes used an investigational blood glucose monitoring system (BGMS) with subject capillary blood obtained from the forearm. BGM results were compared to a reference lab glucose method - Yellow Springs Instrument (YSI) Analyzer. BG results were used to calculate the number of BG results within +/- 5 to 20 mg/dL (for reference BG results <75mg/dL) or +/- 5% to 20% (for reference BG results >=75mg/dL) of the reference method results.

Sponsors and collaborators

Lead sponsor

Ascensia Diabetes Care

Industry

Registry information

Official study title

Clinical Trial Protocol for Ninja 2 System With TATSU Sensor

Important dates

Study start
2010
Primary completion
2010
Study completion
2010
First posted
Dec 30, 2010
Registry last updated
Feb 29, 2016

OpenTrials presents study information sourced from ClinicalTrials.gov. The official registry record should be consulted for the latest information.