DIDGET (Investigational Blood Glucose Monitoring System)
DeviceSubjects with diabetes (with parent/guardian assistance, if applicable)and health care professionals (HCPs) each performed duplicate Blood Glucose (BG) tests from the subject's capillary blood on the DIDGET. To test the full range of glucose concentrations, a venipuncture was performed on some subjects to obtain enough blood for modifying (spiking or glycolyzing) to the extreme glucose concentrations. Results were compared to a laboratory glucose method - the Yellow Springs Instrument (YSI) Analyzer.