IKFE - Institute for Clinical Research and Development
Mainz, 55116, Germany
NCT Number: NCT01504620
The goal of this single-center, comparative, open label, in-vitro diagnostic device performance evaluation was to investigate the performance of two sanofi-aventis BGMSs (BGStar® and iBGStar®) with regard to system accuracy (Polaris), precision and Hct influence (Helios).
Looking for future studies?
Notify Me18 year–75 year
All sexes
Interventional
Phase 4
Mainz, 55116, Germany
The main study (Polaris) investigated blood glucose measuring accuracy and intra-assay precision while the substudy (Helios) investigated potential haematocrit interference of sanofi-aventis BGMSs BGStar® and iBGStar® and partly of competitor BGMSs according to ISO 15197 [1] and TNO 2001 Quality Guideline [2], respectively.
The primary objective of the system accuracy evaluation is to identify the number of blood glucose readings of both sanofi-aventis devices within the ISO 15197 area of acceptable accuracy of ± 20% [DEV%] for blood samples with glucose concentration > 75 mg/dL (4.2 mmol/L) and within the ISO 15197 area of acceptable accuracy of ± 15 mg/dL (± 0.83 mmol/L) for blood samples with glucose concentration < 75 mg/dL (4.2 mmol/L) as compared to the StatStrip® Connectivity reference method. The primary objective of the haematocrit interference evaluation is to demonstrate that the BGStar® and iBGStar® devices meet the Hct interference requirements according to TNO 2001 Quality Guideline when used to measure BG of venous whole blood as compared to a venous plasma reference method (YSI 2300 STAT Plus™) at Hct levels between 35 - 55% in terms of:(1) a mean deviation not exceeding ± 10% [DEV%] for hyperglycaemic glucose concentrations (150 - 180, 250 - 280, 320 - 350 mg/dL) and (2)a mean deviation not exceeding ± 18 mg/dL (1.0 mmol/L) [DEVtotal] for hypo- and normoglycaemic glucose values (50 - 60, 100 - 120 mg/dL) compared to the reference method at Hct levels between 35 - 55%.
Healthy volunteers accepted: Yes
Only the study team can determine whether someone qualifies for participation.
Inclusion criteria
Exclusion criteria
Infusion of glucose to achieve high blood glucose levels
Other names: 5 % glucose solution
i.v. infusion of insulin
Other names: insuman rapid
Time frame: within 10 min up to 300 min after start of experiment
After start of the experiment, blood glucose will be measured by means of the test devices and blood will be drawn for assessment of venous blood glucose concentration with a standard laboratory reference method
Time frame: within 10 min up to 300 min after start of experiment
Blood samples from 5 individuals with different levels of blood glucose were each measured with 10 meters of both investigational devices, the BGStar® and iBGStar®, ten times resulting in a total number of 500 measurements.
Time frame: within 10 min up to 300 min after start of the experiment
One patient provided blood and artificial samples are generated from venous heparinized blood. These blood samples were adjusted to five different BG levels (2.8 mmol/L, 5.6 mmol/L, 8.3 mmol/L, 13.9 mmol/L, 19.4 mmol/L) at five different Hct levels (35%, 40%, 45%, 50%, 55%) to fulfill the range required by the TNO 2001 and tested by all devices.
IKFE Institute for Clinical Research and Development
Other
Performance Evaluation of BGStar and iBGStar in Terms of Accuracy (Polaris), Intra-assay Precision, and Hematocrit Interference (Helios Substudy) Following ISO15197 and TNO Guidelines
Acronym: Polaris
OpenTrials presents study information sourced from ClinicalTrials.gov. The official registry record should be consulted for the latest information.
View the official ClinicalTrials.gov record (opens in a new tab)This listing is for discovery and informational purposes only. It is not medical advice, does not guarantee that a study is recruiting, and does not determine eligibility. Contact the study team and a qualified healthcare professional when considering participation.
Published trials that share one or more normalized conditions with this study.
NCT00038727
Body Weight, CVD
Rockville, Maryland, United States
View Trial DetailsNCT06029517
Arterial Occlusive Diseases, Arteriosclerosis
Buffalo, New York, United States
View Trial DetailsNCT01934660
Cardiovascular Disease, Cardiovascular Diseases
Bethesda, Maryland, United States
View Trial DetailsNCT07728162
Diabetes Mellitus, Disease Attributes
Bolu, Turkey (Türkiye)
View Trial Details