Instantaneous wave-free ratio (iFR™) is a novel non-hyperaemic index of the functional significance of a coronary stenosis. Previous studies have shown variable levels of correlation with the established hyperaemic index FFR. In addition it has been proposed that iFR™ has superior diagnostic accuracy when compared to mean whole cardiac cycle Pd/Pa which can also be used to predict FFR.
We plan to undertake a prospective clinical study in consecutive patients already undergoing FFR assessment in the cardiac catheterisation laboratory to compare the ability of iFR™ and Pd/Pa (both measured using the proprietary Volcano system) to predict FFR. We will explore the level of misclassification of flow limiting disease that results from use of iFR™ and resting Pd/Pa employed using either binary cut-off algorithms or in a hybrid decision making protocol. We plan to analyse 260 vessels over a 18 month period. Hyperaemia will be induced by intravenous adenosine (140 ug/kg/min) administered wherever possible via an antecubital vein. Intra-coronary nitrates will also be given in line with the standard care procedure for FFR measurement. Final clinical decisions following coronary physiology will be based on steady state FFR.
Looking for future studies?
Notify Me18 year and older
All sexes
Observational
Glasgow, Dunbartonshire, G81 4DY, United Kingdom
Title:
A comparative study of resting Pd/Pa, instantaneous wave-free ratio and fractional flow reserve in an unselected population referred for invasive angiography.
Instantaneous wave-free ratio (iFR™) is a novel non-hyperaemic index for assessing the functional significance of a coronary stenosis without coronary vasodilatation. In previous studies it has been compared to the hyperaemic index FFR with variable results. As a guide to determining the need for revascularisation it has been employed using a dichotomous cut-off without FFR or within a hybrid strategy in which lesions with intermediate iFR™ values are further interrogated using FFR.
The comparative diagnostic utility of iFR™ vs resting pressure (Pd/Pa) in reference to FFR is uncertain. We plan to undertake a prospective clinical study in consecutive patients undergoing clinically-indicated FFR assessment in the cardiac catheterisation laboratory with 30-80% diameter stenosis on quantitative coronary angiography (QCA). We will will use a proprietary (Volcano) pressure wire system and iFR ™ algorithm in order to calculate iFR™ and Pd/Pa in both resting and hyperemic conditions as well as FFR.
The sample size is 260 vessels and the enrolment period is 18 months. Hyperaemia will be induced by intravenous adenosine (140 ug/kg/min) administered wherever possible via an antecubital vein. Intra-coronary nitrates will also be given in line with the standard care procedure for FFR measurement.
Design:
In this prospective single centre cohort study all consecutive patients undergoing FFR are eligible for inclusion.
Active Hypothesis: (1) In comparison to an FFR for all strategy, revascularisation decisions made using binary cut-off values of iFR™ or resting Pd/Pa will result in similar levels of disagreement.
Active Hypothesis: (2) In comparison to an FFR for all strategy, revascularisation decisions using hybrid strategies incorporating iFR™ or resting Pd/Pa and FFR will result in similar levels of disagreement.
Active Hypothesis (3): Compared to iFR™ measured under resting conditions, hyperaemic iFR™ has a stronger correlation with FFR. Should this be the case, then the diagnostic efficiency of iFR™ can be interpreted as being improved with pharmacological vasodilatation. The null hypothesis is that there is no difference in diagnostic efficiency between iFR™ and hyperaemic iFR compared to FFR.
The clinical decisions in the catheter laboratory will align with routine care and be informed by all available clinical data and the FFR results.
This study is being conducted independently in the National Health Service without industry support or involvement.
Sample size: 260 vessels.
Statistical Analysis: Independent analysis of the completed dataset will be performed by Dr John McClure a biostatistician and lecturer in the Institute of Cardiovascular and Medical Sciences in Glasgow.
Methods: We will measure resting indices Pd/Pa and iFR™. Following this we will then measure hyperaemic readings including FFR and hyperaemic iFR™ (HiFR) sequentially using peripherally administered adenosine. Upon completion of enrollment we will produce summary statistics describing demographics and procedural data for the study cases. We will then calculate the discriminatory power of iFR™ using both the pre-specified binary cut-off values of 0.90 for iFR™ and 0.92 for resting Pd/Pa and the adenosine zones for iFR of 0.86-0.93 and resting Pd/Pa of 0.87-0.94. We will also analyse the correlation of HiFR with FFR
Healthy volunteers accepted: No
Only the study team can determine whether someone qualifies for participation.
Inclusion criteria
All patients ≥18 years of age, already undergoing pressure wire assessment and able to consent
Exclusion criteria
Inability to receive adenosine
Extremely tortuous vessels
Highly calcified lesions unsuitable for pressure wire assessment
coronary artery occlusion
acute MI within 3 days
Time frame: 30 days
Resting Pd/Pa will be compared to iFR™ in reference to FFR.
Time frame: 30 days
Adenosine zone for iFR is 0.86-0.93, Adenosine zone for resting Pd/Pa is 0.87-0.94.
Time frame: 30 days
FFR recorded by the operator as the minimum value will be compared with the steady state value and the impact on vessel classification will be assessed
Time frame: 30 days
iFR™ will be reassessed after administration of adenosine to evaluate whether this increases agreement with FFR.
Time frame: 30 days
An analysis of the total number of vessels and patients that would potentially undergo revascularisation will be used to compare iFR™ versus FFR guided managment.
Time frame: 30 days
Patients will be monitored during coronary physiology studies to assess the safety of intravenous adenosine infusion with assessment of side effects leading to early termination of the infusion.
Golden Jubilee National Hospital
Other Gov
A Comparative Study of Resting Coronary Pressure Gradient, Instantaneous Wave-free Ratio and Fractional Flow Reserve in an Unselected Population Referred for Invasive Angiography: The VERIFY 2 Study
Acronym: VERIFY2
OpenTrials presents study information sourced from ClinicalTrials.gov. The official registry record should be consulted for the latest information.
View the official ClinicalTrials.gov record (opens in a new tab)This listing is for discovery and informational purposes only. It is not medical advice, does not guarantee that a study is recruiting, and does not determine eligibility. Contact the study team and a qualified healthcare professional when considering participation.
Published trials that share one or more normalized conditions with this study.
NCT02033694
Arterial Occlusive Diseases, Arteriosclerosis
Los Angeles, California, United States
View Trial DetailsNCT06508437
Arterial Occlusive Diseases, Arteriosclerosis
Tours, France
View Trial DetailsNCT04382443
Acute Coronary Syndrome, Arterial Occlusive Diseases
Ciudad de Buenos Aires, Buenos Aires, Argentina
View Trial DetailsNCT04460482
Arterial Occlusive Diseases, Arteriosclerosis
Odense, Denmark
View Trial Details