Ziekenhuis Oost-Limburg
Genk, 3600, Belgium
NCT Number: NCT01759563
Assessment of safety and feasibility of using a nasopharyngeal cooling technique to provide targeted brain cooling during percutaneous implantation of aortic valves. Was it possible to obtain brain temperatures below 34° within a reasonable time interval (max of 1hr) without delaying the procedure of valve implantation.
Looking for future studies?
Notify MeAll sexes
Interventional
Phase 3
Genk, 3600, Belgium
OCCTAVI-I Protocol
6 patients selected for Transcutaneous Aortic Valve Implantation (TAVI)-procedure will be subjected to intra-operative transnasal cooling (RhinoChill) ; cooling will be started after induction of anesthesia until tympanic temperature achieves 34° (cooling will be continued until 33°C of tympanic temperature, below 33°C active cooling will be stopped)
No intra-aortic catheter will be inserted before tympanic temperature reaches 34°
The aim of this study is to test the safety (as referring to routinely used neuromonitoring : ForeSight cerebral tissue oxygen saturation (SctO2) monitoring; and as referring to cardiovascular stability : blood pressure & heartrate) and feasibility (as referring to temperature characteristics) of brain cooling during TAVI (cerebral (or tympanic) cooling rate, differential tympanic-rectal cooling rate, rectal cooling rate, time needed to obtain tympanic temperature below 34°)
Patients will be monitored with FORE-SIGHT cerebral oximeter (regional cerebral oxygen saturation, sensors placed on patient's forehead) and BIS VISTA (bilateral continuousEEG).
Tympanic temperature will be measured every 5 minutes while oesophageal and blood temperature will be continuously measured.
Healthy volunteers accepted: No
Only the study team can determine whether someone qualifies for participation.
Inclusion criteria
Exclusion criteria
Other names: RhinoChill Device, BeneChill company
Time frame: Day1: start of cooling until brain temperature above 35.5°C
By the use of nasopharyngeal cooling, brain temperature will be decreased. Time to brain temperature below 34° will be recorded. Local side effects could occur due to the use of this specific nasopharyngeal cooling technique and systemic side effects (especially cardiovascular) could occur due to the initiation of hypothermia
Time frame: Day 1: start of cooling until brain temperature above 35.5°C
NIRS (or Near-infrared Spectroscopy) enables non-invasive monitoring of cerebral oxygen saturation during TAVI procedure and during the installation of targeted brain cooling. NIRS will be used as cerebral monitoring technique during this study protocol
Time frame: Day 1: start of cooling until brain temperature above 35.5°C
Induction of nasopharyngeal cooling and targeted brain cooling will result in systemic hypothermia. Systemic temperature (rectal temperature) will be recorded to assess the influence of nasopharyngeal cooling on systemic body temperature
Hasselt University
Other
Safety and Feasibility Study of Nasopharyngeal Cooling (RhinoChill Device) During Transcutaneous Aortic Valve Implantation
Acronym: OCCTAVI-I
OpenTrials presents study information sourced from ClinicalTrials.gov. The official registry record should be consulted for the latest information.
View the official ClinicalTrials.gov record (opens in a new tab)This listing is for discovery and informational purposes only. It is not medical advice, does not guarantee that a study is recruiting, and does not determine eligibility. Contact the study team and a qualified healthcare professional when considering participation.
Published trials that share one or more normalized conditions with this study.
NCT03424941
Aortic Stenosis, Aortic Valve Disease
Graz, Austria
View Trial DetailsNCT06542380
Aortic Valve Disease, Aortic Valve Stenosis
Mitte, State of Berlin, Germany
View Trial DetailsNCT03966417
Aortic Valve Disease, Aortic Valve Stenosis
Ljubljana, Slovenia
View Trial DetailsNCT04286893
Aortic Valve Disease, Aortic Valve Stenosis
Ljubljana, Slovenia
View Trial Details