Each participant will require (1) visit, which will last approximately 4-5 hours.
At start of visit:
- Vitals are taken (temperature, heart rate, SpO2, respiration rate, blood pressure, Borg dyspnea score, Mmrc score) while participant is seated and using their normal oxygen prescription.
- Patient will remain seated during the following procedure with SpO2 and heart rate continuously being monitored.
Pulsed Oxygen Control at Rest:
- Participant's oxygen will be turned off for 10 minutes as a washout period.
- Using the oxygen tank with the pulse regulator, the participant will be titrated to a pulse rate that maintains their SpO2 at an average of 95%.
a. Increasing volumes of oxygen in 1 L(liter)/min increments every thirty seconds until an average of 95% SpO2 is maintained for 30 seconds.
- Once the participant is titrated to the correct amount of oxygen, they will remain on the oxygen for an additional 3 minutes
- Final oxygen flow rate, SpO2, Heart Rate, and Borg Dyspnea Score will be recorded in the last minute of the test.
Mixed Continuous Air/Oxygen Efficacy at Rest:
- Participant's oxygen will be turned off for 10 minutes as a washout period.
- The participant will be delivered pulsed oxygen and high flow ambient air using the oxygen tank with the pulse regulator and Vapotherm setup, the participant will be titrated to a pulse rate that maintains their SpO2 at an average of 95%.
- High flow ambient air will be set to 15 L/min
- Increasing volumes of oxygen from in 1 L/min increments every thirty seconds until an average of 95% SpO2 is maintained for 30 seconds.
- Once the participant is titrated the correct amount of oxygen, they will remain on the oxygen for an additional 3 minutes.
- Final oxygen flow rate, SpO2, Heart Rate, and Borg Dyspnea Score will be recorded in the last minute of the test.
- Steps 5-8 are then repeated two more times for pulsed oxygen plus continuous flow ambient air at flow rates of 20 and 25 L/min.
Mixed Pulsed Air/Oxygen Efficacy at Rest:
- Participant's oxygen will be turned off for 10 minutes as a washout period
- The participant will be delivered the out-of-phase pulsed oxygen and high flow ambient air using the oxygen tank with the pulse regulator and Vapotherm setup, the participant will be titrated to a pulse rate that maintains their SpO2 at an average of 95%
- High flow ambient air will be set to 15 L/min
- Increasing volumes of oxygen from in 1L/min increments every thirty seconds until an average of 95% SpO2 is maintained for 30 seconds.
- Once the participant is titrated the correct amount of oxygen, they will remain on the oxygen for an additional 3 minutes
- Final oxygen flow rate, SpO2, Heart Rate, and Borg Dyspnea Score will be recorded in the last minute of the test.
- Steps 10-13 are then repeated two more times for out-of phase pulsed oxygen plus pulsed high flow ambient air at flow rates of 20 and 25 L/min.