Pump suspension
DeviceThe study laptop will communicate to the pump causing a suspension based on output from the algorithm which predicts hypoglycemia based on the continuous glucose sensor trend.
NCT Number: NCT01591681
The purpose of this study is to see whether low blood sugar at night can be reduced by using a system that turns off the insulin pump automatically. The study system includes a combination continuous glucose monitor (CGM)/ insulin pump made by Medtronic MiniMed, Inc and a regular laptop computer that runs a computer program that predicts low blood sugar. It works by (1) measuring the glucose levels under the skin with a continuous glucose monitor, (2) using a computer program on a laptop to predict what will happen to the glucose level over the next 35-55 minutes, and (3) turning off the insulin pump when the computer program predicts that low blood sugar will occur.
This study has several phases and will take about 3 months for a patient to complete. Patients will use the Medtronic CGM with the Enlite sensor at home for 10-15 days to be sure that they are able and willing to use this system and to determine if they meet the investigators study criteria to proceed with the next phase of the study. Patients will be provided teaching on how to use CGM data in real time. If a patient is not using a Medtronic CGM already, the patient will first use one at home for 10-15 days to be sure that he/she is able and willing to use it. If a patient is already using a Medtronic CGM, then his/her most recent 10-15 days of data will be used to find out if he/she is eligible. Those who need to complete the CGM run-in phase will have an extra office visit for training.
If eligible to continue in the study, patients will need to use the study system for 5 nights at home so that the investigators can make sure they are able to use it correctly. After that, patients will be asked to use the study system each night for an additional 6 to 8 weeks. If the system is active and predicts that a patient's blood sugar will become low, the insulin pump will shut off for up to 2 hours.
The study will include about 45 individuals at 3 clinical centers in the United States and Canada.
Looking for future studies?
Notify Me15 year–45 year
All sexes
Interventional
Phase 2
St. Joseph's Health Care, London, Ontario, Canada
Patients who are eligible for the clinical trial initially will use a Medtronic Continuous Glucose Monitor (CGM) at home to verify that the subject is able to use the CGM and insert sensors. Subjects already using a Medtronic CGM will skip this phase, provided that the eligibility criteria for CGM use and the study are met.
Patients who meet criteria for sensor use and the study will then use the closed-loop system at home for 5 days to demonstrate their ability to use the system and submit study data to the Coordinating Center.
Patients who successfully demonstrate their ability to use the system at home as described above will be eligible for the randomized trial phase. This phase consists of use of the full system as an outpatient for approximately 42 nights:
Upon completion of the study, patients as well as study clinicians will be asked to complete a questionnaire regarding use of the study system.
There will be one follow-up visit after 21 days in the clinical trial and a second follow-up visit after the completion of 42 nights of successful study system use. A successful night of study system use is defined as use of the system for at least four hours. Phone contacts with the patients will be made once a week.
Healthy volunteers accepted: No
Only the study team can determine whether someone qualifies for participation.
Inclusion criteria
Exclusion criteria
The study laptop will communicate to the pump causing a suspension based on output from the algorithm which predicts hypoglycemia based on the continuous glucose sensor trend.
Time frame: Overnight from system activation to deactivation in the morning upon awakening for 42 nights of system use
Each night is categorized as to whether hypoglycemia occurred. Hypoglycemia is defined as the occurrence of one or more CGM glucose values ≤60 mg/dL. The percentage of hypoglycemic nights will be tabulated separately with versus without the closed-loop control system in use. A repeated measures logistic regression model will be used to compare intervention versus control nights accounting for correlated data from the same subject and adjusting for the baseline (bedtime) sensor glucose.
Time frame: Overnight from system activation to deactivation in the morning upon awakening for 42 nights of system use
The median percentages of the number of glucose values with values of 71-180 mg/dL overall.
Time frame: Overnight from system activation to deactivation in the morning upon awakening for 42 nights of system use
Time frame: Overnight from system activation to deactivation in the morning upon awakening for 42 nights of system use
Time frame: 42 mornings following night of system use
Measured with a study home blood glucose meter.
Time frame: 42 mornings following night of system use
Measured with a study home blood glucose meter.
Time frame: 42 mornings following night of system use
Blood ketones measured with a study blood ketone meter.
Time frame: 42 mornings following night of system use
Urine ketones measured each morning with Ketostix.
Time frame: Overnight from system activation to deactivation in the morning upon awakening for 42 nights of system use
Calculated as the median of the overall mean.
Time frame: Overnight from system activation to deactivation in the morning upon awakening for 42 nights of system use
The measure is reporting area under the curve for glucose concentrations below 250 mg/dL and above 60 mg/dL. Overall time below and above a threshold and area under a curve was divided by total time and multiplied by 8 hours.
Time frame: Overnight from system activation to deactivation in the morning upon awakening for 42 nights of system use
Jaeb Center for Health Research
Other
Reduction of Nocturnal Hypoglycemia by Using Predictive Algorithms and Pump Suspension: An Outpatient Pilot Feasibility and Efficacy Study
OpenTrials presents study information sourced from ClinicalTrials.gov. The official registry record should be consulted for the latest information.
View the official ClinicalTrials.gov record (opens in a new tab)This listing is for discovery and informational purposes only. It is not medical advice, does not guarantee that a study is recruiting, and does not determine eligibility. Contact the study team and a qualified healthcare professional when considering participation.
Published trials that share one or more normalized conditions with this study.
NCT01897688
Autoimmune Diseases, Diabetes Mellitus
Chicago, Illinois, United States
View Trial DetailsNCT05671679
Autoimmune Diseases, Diabetes Mellitus
Bern, Canton of Bern, Switzerland
View Trial DetailsNCT07373236
Autoimmune Diseases, Diabetes Mellitus
Hellerup, Denmark
View Trial DetailsNCT01354288
Acid-Base Imbalance, Acidosis
Le Kremlin-Bicêtre, France
View Trial Details