NCT Number: NCT00950495

Oral Appliance Therapy in Obstructive Sleep Apnea

The aim of the study was to compare the effects of a mandibular advancement device (MAD) with those of nasal continuous positive airway pressure(nCPAP) and of an intra-oral placebo device in obstructive sleep apnea (OSA) patients. The hypothesis for this study was that both MAD therapy and nCPAP therapy have similar, better treatment effects than placebo therapy in OSA. The study was performed according to the CONSORT (consolidated standards of reporting trials) statement (Altman et al., 2001), employing a parallel-group, randomized, placebo-controlled trial design.

Completed

Looking for future studies?

Notify Me

Key information

Age range

18 year and older

Sex eligibility

All sexes

Study type

Interventional

Phase

Not applicable

Primary location

Academic Centre for Dentistry Amsterdam (ACTA), Research Institute Move, University of Amsterdam and VU University Amsterdam

Amsterdam, 1066 EA, Netherlands

About this study

Setting and participants:

Eligible OSA patients, living in the greater Amsterdam area, were referred to the Slotervaart Medical Center by their family physician. All patients underwent a thorough medical examination, including a full polysomnographic (PSG) recording, at the departments of Neurology, Pulmonary Medicine, and ENT, as well as a thorough dental examination at the Department of Oral Kinesiology of ACTA. OSA patients were invited for participation in this study when they fulfilled the inclusion criteria and exclusion criteria.

Randomisation and allocation:

After written informed consent was obtained, the patients were randomly allocated to one of three therapy groups (MAD, nCPAP, or placebo; see below). To ensure that the groups were of approx. the same size, block randomisation was used. Block sizes were 6, 12, and 18; sizes were randomly varied. The allocation sequence was automatically generated and subsequently concealed by an independent co-worker, who kept a paper copy in a lockable drawer. Sealed opaque envelopes were used to conceal the allocation from the principal investigator.

Interventions and blinding:

Three forms of therapy interventions were used in this parallel-group study. First, an individually fabricated MAD with an adjustable protrusive mandibular position at a constant vertical dimension was used. Second, nCPAP of the REMstar Pro system was used (Respironics, Herrsching, Germany). Third, a thin (< 1 mm), hard acrylic-resin palatal splint with only a partial palatal coverage was used as a placebo.

Patients were blinded to the nature of the assigned therapy (placebo or active). After evaluating the therapy, all patients were asked if they were of the opinion that they had received an active or placebo treatment. Blinding of the analyst was ascertained by assigning codes to data sets and by analyzing these sets in random blocks.

Procedure:

From all patients, two full polysomnographic (PSG) recordings were obtained in the sleep laboratory of the Slotervaart Medical Center, using Siesta hardware and Pro-Fusion software (Compumedics, Abbotsford, Australia): one before therapy assignment (baseline PSG) and one after 6 ± 2 months (mean ± SD) of treatment (therapy evaluation PSG). For the MAD and nCPAP groups, the third and fourth PSG recordings for therapy evaluation were performed 6 months and one year after the first therapy evaluation (long-term follow-up). The primary and secondary outcome measures were obtained at baseline and at therapy evaluations.

The MAD and nCPAP were titrated before the start of the treatment. The titration of the nCPAP was performed during a third sleep laboratory examination. The pressure was increased in incremental steps of 1 cm H2O/h, until respiratory disturbances and respiration-related arousals were reduced to ≤ 5/h, and snoring was minimized.

For the titration of the MAD, four ambulatory PSG recordings were obtained at regular intervals, using Monet hardware and Rembrandt Software (Medcare Automation B.V., Amsterdam, The Netherlands). The most effective protrusion position of the MAD (i.e., the mandibular position that yielded the lowest AHI value) was chosen from among four randomly offered positions (viz., 0%, 25%, 50%, and 75% of the maximum protrusion).

For the placebo group, the study procedure was made equally intense as that for the MAD group by making four ambulatory PSG recordings at regular intervals as well.

Who can participate

Healthy volunteers accepted: No

Only the study team can determine whether someone qualifies for participation.

Inclusion criteria

  • age > 18 years
  • apnea-hypopnea index (AHI) between 5 and 45 events per hour

Exclusion criteria

  • evidence of respiratory/ sleep disorders other than obstructive sleep apnea
  • a body mass index (BMI) > 40
  • medication usage that could influence respiration or sleep
  • periodic limb movement disorder
  • previous treatment with CPAP or MAD
  • reversible morphological upper airway abnormalities (e.g., enlarged tonsils)
  • temporomandibular disorders (based on a functional examination of the masticatory system)
  • untreated periodontal problems
  • dental pain
  • lack of retention possibilities for a mandibular advancement device or a placebo appliance

Treatment and study plan

Mandibular advancement device (MAD)

Device

This oral appliance keeps the mandible and the tongue in a protruded position. This results in an upper respiratory tract widening and/or reduced collapsibility of the upper airway, thereby preventing the upper airway collapse during sleep.

Other names: mandibular repositioning appliance, mandibular advancement appliance, mandibular repositioning device, mandibular repositioning splint, mandibular advancement splint, oral appliance

nasal CPAP

Device

The CPAP machine delivers a stream of compressed air via a hose to a nose mask, splinting the airway (keeping it open under air pressure) so that unobstructed breathing becomes possible, reducing and/or preventing apneas and hypopneas

PLACEBO

Device

as a placebo, a thin (< 1 mm), hard acrylic-resin palatal splint with only a partial palatal coverage was used

Primary outcomes

  1. apnea-hypopnea index (AHI)

    Time frame: 6 months

Secondary outcomes

  1. other respiratory variables (apnea-hypopnea index during different sleep stages and positions);sleep variables (sleep stages and arousals);excessive daytime sleepiness;health perception;compliance;snoring reports; side-effects

    Time frame: 6 months

Sponsors and collaborators

Lead sponsor

VU University of Amsterdam

Other

Registry information

Official study title

Effects of an Oral Appliance on Obstructive Sleep Apnea: A Randomized, Placebo-Controlled Trial

Important dates

Study start
2003
Primary completion
2009
Study completion
2009
First posted
Jul 31, 2009
Registry last updated
Dec 25, 2014

OpenTrials presents study information sourced from ClinicalTrials.gov. The official registry record should be consulted for the latest information.

View the official ClinicalTrials.gov record (opens in a new tab)

This listing is for discovery and informational purposes only. It is not medical advice, does not guarantee that a study is recruiting, and does not determine eligibility. Contact the study team and a qualified healthcare professional when considering participation.

Published trials that share one or more normalized conditions with this study.