NCT Number: NCT04030104

Optoacoustic Images Versus Imagio® Ultrasound

The Imagio® Pivotal Study is intended to evaluate if the results observed in the previous Feasibility Study can be confirmed for pre-specified effectiveness endpoints. Using Intention-to-Diagnose (ITD) masses from the PIONEER Pivotal study, 480 to 840 masses are to be used for the Pivotal Study, the remaining masses will be reserved for training the SenoGram® model.

Completed

Looking for future studies?

Notify Me

Key information

Age range

18 year and older

Sex eligibility

Female

Study type

Interventional

Phase

Not applicable

Primary location

American College of Radiology Center for Research and Innovation

Philadelphia, Pennsylvania, 19103, United States

About this study

This will be a controlled, blinded, multi-reader, multi-case (MRMC) study using a sequential design. The study will include 15 readers with an additional 5 as back-up readers depending on qualifications and availability. Readers that participated in the Seno PIONEER Study (NCT01943916) or the Reader-01 Feasibility Study (NCT03708393) are not eligible to participate as readers in this Pivotal Study.

Imagio® [Ultrasound (IUS) + Optoacoustic (OA)] (IUS)(IUS+OA) training will be completed prior to any reads taking place. Read 1 will be immediately followed by Read 2 within the same read session.

Who can participate

Healthy volunteers accepted: No

Only the study team can determine whether someone qualifies for participation.

Inclusion criteria

  • One analyzable mass per patient: BI-RADS 3 and 4a, 4b, 4c and 5 masses as declared by clinical site investigator via PIONEER study inclusion criteria and categorized as BIRADS 3, 4a, 4b 4c and 5 by conventional diagnostic ultrasound (CDU)
  • Masses declared to be in the PIONEER Intention to Diagnose (ITD)/analysis population, including high risk cases per original PIONEER protocol
  • Patient age, indication for study entry and available medical history
  • Evaluable mammograms and OA and IUS video loops and stills for each mass

Exclusion criteria

  • Critical missing IUS or OA still image and/or video loop views or incorrect IUS or OA stills and video loops that would preclude a case from being evaluated by readers
  • Reader-02 Proficiency Test and training cases

Treatment and study plan

Reader Study - Imagio Ultrasound

Device

Imagio ultrasound images to be reviewed as part of Reader study

Reader Study Imagio Ultrasound + Optoacoustic Imaging

Device

Imagio Ultrasound + Optoacoustic images to be reviewed as part of Reader study

Mammography

Device

Mammography as available per standard of care

Other names: Mammo

Primary outcomes

  1. Gain in Specificity at Fixed 98% Sensitivity (fSp)

    Time frame: Baseline to 12 months +/- 30 days follow-up

    Primary effectiveness endpoint was the difference (gain) in specificity (fSp) at fixed 98% sensitivity for the Imagio IUS+OA relative to IUS alone, across all 15 independent readers; both imaging modalities used in each subject (subject as own control); results for each imaging modality compared to biopsy diagnosis or 12-month follow-up ruling of benign as determined by truth panel (ground truth). fSp derived from empirical receiver operating characteristic (ROC) using endpoint interpolation.

Secondary outcomes

  1. Negative Likelihood Ratio (NLR)

    Time frame: Baseline to 12 months +/- 30 days follow-up

    NLR for the Imagio IUS+OA relative to IUS alone, across all 15 independent readers; both imaging modalities used in each subject (subject as own control); results for each imaging modality compared to biopsy diagnosis or 12-month follow-up ruling of benign as determined by truth panel (ground truth). NLR = [(1-sensitivity) / specificity].

  2. Positive Likelihood Ratio (PLR)

    Time frame: Baseline to 12 months +/- 30 days follow-up

    PLR for the Imagio IUS+OA relative to IUS alone, across all 15 independent readers; both imaging modalities used in each subject (subject as own control); results for each imaging modality compared to biopsy diagnosis or 12-month follow-up ruling of benign as determined by truth panel (ground truth). PLR = [sensitivity / (1-specificity)].

  3. Partial Area Under the Curve (pAUC)

    Time frame: Baseline to 12 months +/- 30 days follow-up

    pAUC considers only the region of the area under the ROC curve that corresponds to clinically relevant values of sensitivity, defined as 95% to 100%, for Imagio IUS+OA vs. IUS alone, averaged across 15 readers. Both imaging modalities used in each subject (subject as own control); results for each imaging modality compared to biopsy diagnosis or 12-month follow-up ruling of benign as determined by truth panel (ground truth). A higher pAUC indicates better diagnostic accuracy.

Sponsors and collaborators

Lead sponsor

Seno Medical Instruments Inc.

Industry

Registry information

Official study title

Imagio® Pivotal Multi-Reader, Multi-Case Study of Optoacoustic Images Versus Imagio® Ultrasound to Guide Decision to Biopsy

Acronym: Reader-02

Important dates

Study start
2019
Primary completion
2019
Study completion
2019
First posted
Jul 23, 2019
Registry last updated
Jul 12, 2021