NCT Number: NCT01514305

Observational Study to Evaluate the Effectiveness and Safety Study of the Dexcom G4™ Continuous Glucose Monitoring System

The purpose of this study is to evaluate the effectiveness and safety of the Dexcom G4 System when used as an adjuvant to blood glucose testing.

Device performance will be primarily evaluated in terms of the proportion of glucose values within a pre-specified range compared to reference values.

Safety data of the G4 System will also be collected vis-a-vis adverse event reporting characterized by the incidence and severity of Serious Adverse Device Events, and Adverse Device Events experienced by study participants.

Completed

Looking for future studies?

Notify Me

Key information

Age range

18 year and older

Sex eligibility

All sexes

Study type

Observational

Primary location

Rocky Mountain Diabetes and Osteoporosis Center

Idaho Falls, Idaho, 83404, United States

Who can participate

Healthy volunteers accepted: No

Only the study team can determine whether someone qualifies for participation.

Inclusion criteria

  • Age 18 years or older;
  • Have been diagnosed with insulin-requiring diabetes
  • Abstain from injecting insulin or wear an insulin pump insertion set within 3 inches from the sensor site during sensor wear;
  • Insert sensors on their own and wear 2 systems simultaneously;
  • Use only the blood glucose meter provided for all blood glucose measurements performed during sensor wear and not allow others to use this meter during the study;
  • Participate in one in-clinic session comprising of fingersticks per hour and have blood draws for the entire in-clinic session;
  • Have an intravenous catheter inserted for 4 blood draws per hour
  • Willing to perform SMBG during home use with the meter provided;
  • Refrain from the use of acetaminophen during sensor wear period and the day prior to sensor insertion;
  • Speak, read, and write English;
  • Willing and able to be compliant with provisions laid out in this protocol.

Exclusion criteria

  • Have extensive skin changes/diseases that preclude wearing devices on normal skin (e.g., extensive psoriasis, recent burns or severe sunburn, extensive eczema, extensive scarring, extensive tattoos, dermatitis herpetiformis) at the proposed wear sites;
  • Allergy to medical-grade adhesives;
  • Pregnant as demonstrated by a positive pregnancy test within 72 hours of sensor insertion,
  • Dialysis treatment;
  • Hematocrit that is outside the range of 30-55% at screening visit;

Treatment and study plan

Primary outcomes

  1. Absolute Relative Difference (%) to Reference Standard

    Time frame: one time measure (day 1)

    The primary objective is to characterize the System performance with respect to laboratory reference venous sample measurements. The device performance will be primarily evaluated in terms of point and rate accuracy of the G4 System in reference to YSI. The point accuracy is measured as the proportion of G4 System values that are within ±20% of YSI reference value for glucose levels >80 mg/dL and ±20 mg/dL at YSI glucose levels <80 mg/dL.

Sponsors and collaborators

Lead sponsor

DexCom, Inc.

Industry

Registry information

Important dates

Study start
2011
Primary completion
2011
Study completion
2011
First posted
Jan 23, 2012
Registry last updated
Jan 23, 2012

OpenTrials presents study information sourced from ClinicalTrials.gov. The official registry record should be consulted for the latest information.

View the official ClinicalTrials.gov record (opens in a new tab)

This listing is for discovery and informational purposes only. It is not medical advice, does not guarantee that a study is recruiting, and does not determine eligibility. Contact the study team and a qualified healthcare professional when considering participation.

Published trials that share one or more normalized conditions with this study.