Completed

NCT Number: NCT02363907

Observational Study to Evaluate the Effectiveness and Safety of the Dexcom G4 Continuous Glucose Monitoring System in Pediatrics

The purpose of this study is to evaluate the effectiveness and safety of the Dexcom G4 System when used as an adjuvant to blood glucose testing over a 7-day period in people 2 to 17 years-old with diabetes mellitus.

Completed

Looking for future studies?

Notify Me

Key information

Age range

2 year–17 year

Sex eligibility

All sexes

Study type

Observational

Primary location

Rocky Mountain Diabetes and Osteoporosis Center

Idaho Falls, Idaho, 83404, United States

About this study

Performance of the system will be primarily evaluated by comparing the blood glucose measurements from the reference device, a blood glucose meter, to the CGM System values collected during the study. Specifically, performance will be primarily evaluated in terms of the proportion of G4 System values that are within ±20% of glucose meter reference value for glucose levels >80 mg/dL and ±20 mg/dL of glucose meter reference values <80 mg/dL. These G4-meter matched pairs will be analyzed across the 7 days of wear.

Safety data of the CGM System will also be collected and safety will be characterized by the incidence of anticipated device-related adverse events (ADEs) of subjects who participate in the study.

Who can participate

Healthy volunteers accepted: No

Only the study team can determine whether someone qualifies for participation.

Inclusion criteria

  • Ages 2 to 17 years;
  • Diagnosis of diabetes mellitus;
  • Treatment of diabetes with insulin and/or oral diabetes medication Note: Insulin-using subjects only - willing to avoid insulin injections or wear an insulin pump insertion set within 3 inches from the sensor site;
  • Willing to refrain from the use of medications containing acetaminophen during the sensor insertion period and for one day prior to sensor insertion;
  • Able to perform up to 7 fingersticks per day during home use (approximately 2 calibration fingersticks or as often as prompted by the device and approximately 5 for diabetes management and/or detection of hyper and hypoglycemia); and
  • Subject or guardian is able to speak, read, and write English.

Exclusion criteria

  • Extensive skin changes/diseases that preclude wearing the required number of devices on normal skin (e.g. extensive psoriasis, recent burns or severe sunburn, extensive eczema, extensive scarring, extensive tattoos, dermatitis herpetiformis) at the proposed wear sites;
  • Known allergy to medical-grade adhesives;
  • For subjects of childbearing potential: are pregnant as demonstrated by a positive pregnancy test within 72 hours prior to device insertion;
  • Hematocrit value outside the range of the blood glucose monitoring system used in the study;
  • Current participation in another investigational study protocol (if a subject has recently completed participation in another drug study, the subject must have completed that study at least 30 days prior to being enrolled in this study);
  • Planned MRI scan, CT scan, or diathermy during the week of the study; or
  • Any condition that, in the opinion of the Investigator, would interfere with their participation in the trial (e.g., marked visual impairment) or pose an excessive risk to study staff handling venous blood samples (e.g., known history of hepatitis B or C, or HIV).

Treatment and study plan

Primary outcomes

  1. Point Accuracy of CGM ISF Readings to Blood Glucose Measured by a Reference Device

    Time frame: Measured during clinic session during 7 day sensor wear period

    "Point Accuracy was evaluated in terms of the percentage of CGM values that were within ±20% of glucose meter reference value for glucose levels >80 mg/dL and ±20 mg/dL of glucose meter reference values for glucose levels <80 mg/dL

Sponsors and collaborators

Lead sponsor

DexCom, Inc.

Industry

Registry information

Official study title

Observational Study to Evaluate the Effectiveness and Safety of the Dexcom G4

Important dates

Study start
2012
Primary completion
2012
Study completion
2012
First posted
Feb 16, 2015
Registry last updated
Jul 27, 2017

OpenTrials presents study information sourced from ClinicalTrials.gov. The official registry record should be consulted for the latest information.

View the official ClinicalTrials.gov record (opens in a new tab)

This listing is for discovery and informational purposes only. It is not medical advice, does not guarantee that a study is recruiting, and does not determine eligibility. Contact the study team and a qualified healthcare professional when considering participation.

Published trials that share one or more normalized conditions with this study.