Skip to main content
OpenTrials
Completed

NCT Number: NCT03287856

NVT ALLEGRA TAVI System TF in Failing Surgical Aortic Bioprosthesis

The study investigates the technical feasibility of the replacement of failed surgical bioprosthetic aortic valves by the ALLEGRA Transcatheter Heart Valve (THV) and describes the safety and performance profile

Completed

Looking for future studies?

Notify Me

Key information

Age range

18 year and older

Sex eligibility

All sexes

Study type

Interventional

Phase

Not applicable

Primary location

Segeberger Kliniken, Herzzentrum, Bad Segeberg, Germany

Loading trial locations.

Who can participate

Healthy volunteers accepted: No

Only the study team can determine whether someone qualifies for participation.

Inclusion criteria

  • ≥18 years
  • Symptomatic degeneration of aortic bioprosthesis showing echocardiographically mean aortic gradient >40mmHg or peak jet velocity >4.0m/s and AVA<1.0cm2 OR symptomatic patients with severe bioprosthetic valve insufficiency.
  • High risk for redo surgery defined by STS ≥10% / EuroScore II ≥7% OR as assessed by the heart team
  • Has signed the Patient Informed Consent Form
  • Willing and able to comply with requirements of the study, including all follow-up visits
  • Female patients of childbearing potential have a negative pregnancy test ≤7 days before the procedure and are willing to use a reliable method of birth control for the duration of study participation

Exclusion criteria

  • Low position of the coronary ostia, especially in combination with shallow sinuses
  • Echocardiographic evidence of intracardiac mass, thrombus or vegetation
  • Significant aortic disease such as severe obstructive calcification or marked tortuosity or kinking which will preclude a safe advancement of the ALLEGRA TAVI System TF
  • Iliofemoral vessel conditions such as severe obstructive calcification, severe tortuosity or kinking that would preclude safe placement of an 18 Fr introducer sheath or make endovascular access to the aortic valve impossible
  • Severe mitral insufficiency
  • Internal diameter of the bioprosthesis is ≤16 mm or >28 mm
  • Patient prosthesis mismatch (EOAi ≤0.65 cm2/m2) as the underlying cause of the poor valve function and need for re-intervention
  • Non-valvular stenosis as the underlying cause of the poor valve function and need for re-intervention
  • Failing pre-existing prosthetic heart valve or prosthetic ring in any other position than aortic
  • Partially detached leaflets that in the aortic position may obstruct a coronary ostium.
  • Existence of aortic conduit, aortic arch replacement, stentless bioprosthesis and autologous valve replacement
  • Paravalvular leak of the failing surgical bioprosthesis (between failing surgical bioprosthesis and native annulus)
  • LVEF <20%
  • Evidence of active endocarditis or other acute infections
  • End stage renal disease requiring chronic dialysis or creatinine clearance <20 ml/min or serum creatinine >3.0 mg/dl (264 µmol/l)
  • Known hypersensitivity to contrast media, which cannot be adequately pre-medicated or contraindication to anticoagulant or anti-platelet medication or to Nitinol alloy or to bovine tissue
  • Evidence of an acute myocardial infarction within the past 30 days
  • Cerebral vascular accident (TIA, Stroke) within past 6 months (≤180 days)
  • Evidence of active peptic ulcer or upper gastrointestinal bleeding within past 90 days
  • Untreated clinically significant coronary artery disease requiring revascularization
  • Hemodynamic instability (e.g. cardiogenic shock) requiring inotropic support or mechanical heart assistance (e.g. VAD, IABP)
  • Uncontrolled (therapy resistant) atrial fibrillation
  • Need for emergency surgery for any reason
  • Life expectancy ≤ 12 months due to other medical illness
  • Currently participating in another investigational drug or device study

Treatment and study plan

Transcatheter Aortic Valve Implantation (TAVI)

Device

Implantation of the ALLEGRA Transcatheter Heart Valve in failing surgical bioprosthesis

Primary outcomes

  1. Mean pressure gradient

    Time frame: immediately after index-procedure

  2. 30-days survival

    Time frame: 30 days

    Proportion of participants with overall survival at 30 days

Secondary outcomes

  1. Cardiovascular mortality

    Time frame: 30 days

  2. Hemodynamic parameters

    Time frame: up to12 months

    pressure gradient, effective orifice area, velocity

  3. Technical success of implantation

    Time frame: immediately after index-procedure

    defined as absence of procedural mortality AND correct positioning of a single prosthetic heart valve into the proper anatomical location AND no severe (<0.65 cm2/m2) prosthesis - patient mismatch AND mean aortic valve gradient <20 mmHg AND no moderate or severe prosthetic valve regurgitation

  4. Early safety

    Time frame: 30 days

    all-cause mortality, all stroke (disabling and non-disabling), life-threatening bleeding, acute kidney injury stage 2 or 3 (including renal replacement therapy), coronary artery obstruction requiring Intervention, major vascular complication, valve-related dysfunction requiring repeat procedure

  5. Time related valve safety

    Time frame: up to 12 months

    structural valve deterioration as defined by

    • requiring repeat procedure (transcatheter or surgical heart valve replacement)
    • valve-related dysfunction defined by
    • mean aortic valve gradient ≥20 mmHg AND
    • no moderate or severe prosthetic valve regurgitation
  6. New Pacemaker implantation

    Time frame: up to 12 months

    Proportion of patients with pacemaker implantations after index-procedure

  7. NYHA classification

    Time frame: up to 12 months

Sponsors and collaborators

Lead sponsor

NVT GmbH

Industry

Registry information

Official study title

Investigation of the Safety and Performance of the NVT ALLEGRA TAVI System TF in Patients With Failed Surgical Aortic Bioprosthesis and Elevated Surgical Risk

Acronym: VIVALL

Important dates

Study start
2017
Primary completion
2018
Study completion
2019
First posted
Sep 19, 2017
Registry last updated
Jan 10, 2020

OpenTrials presents study information sourced from ClinicalTrials.gov. The official registry record should be consulted for the latest information.

View the official ClinicalTrials.gov record (opens in a new tab)

This listing is for discovery and informational purposes only. It is not medical advice, does not guarantee that a study is recruiting, and does not determine eligibility. Contact the study team and a qualified healthcare professional when considering participation.