Novel Portable Diagnostic Device for Automatic Detection of Relative Afferent Pupillary Defect

This study is to evaluate the ability of a newly designed device, Optic Nerve Glass (O-Glass) to detect relative afferent pupillary defect (RAPD). In this prospective study, 44 patients (diagnosed RAPD- positive) enrolled the study. They were examined for an RAPD by O-Glass and also manual swinging flashlight method (SFM) . This newly designed instrument captures and records eye pictures. The images will be processed and analyzed using computerized software to calculate pupillary measurements.

Completed

Looking for future studies?

Notify Me

Key information

About this study

Pupil response to light stimulation is a basic clinical approach to the assessment of a patient with visual loss. Asymmetric response may indicate a relative afferent pupillary defect (RAPD). The rapid detection device for relative afferent pupillary defect (present device) is a newly designed portable facility with computerized software on a mobile device. This allows for field ophthalmic examination and identification of RAPDs quickly and accurately and also record pupillary movements for further processing and analysis or send the information and images via Wi-Fi. We aimed at comparing and evaluating this device with available methods to develop a test which is practically easy and quick with objective results and no need for specialist interpretation, so that any technician can perform the test automatically. Distinctive software allows simple use of the device by field personnel with minimal training. Each patient was investigated by swinging flashlight method and the newly constructed automated O-glass. The swinging flashlight test procedure, also called as Marcus Gunn Test, is well known in ophthalmology science.The hardware for this device includes 4 different parts: Camera and optics, light control system, power control system, and the microcontroller.The two main components of the software are the ability to communicate with the hardware wirelessly, and the pattern recognition system.

Who can participate

Healthy volunteers accepted: No

Only the study team can determine whether someone qualifies for participation.

Inclusion criteria

  • positive relative afferent pupillary defect

Exclusion criteria

-

Treatment and study plan

O-Glass

Device

The new diagnostic device to distinguish relative afferent pupillary defect(RAPD).

Swinging Flashlight Test

Other

The swinging-flashlight test is used to help a practitioner identify a relative afferent pupillary defect.For an adequate test, vision must not be entirely lost. In dim room light, the examiner notes the size of the pupils. The patient is asked to gaze into the distance, and the examiner swings the beam of a penlight back and forth from one pupil to the other, and observes the size of pupils and reaction in the eye that is lit.

Other names: Other diagnostic method

Primary outcomes

  1. Number of Participants With Detected Relative Afferent Pupillary Defect (RAPD)

    Time frame: up to 6 months

    The instrument illuminated the eyes alternatively and took images and recorded pupillary reflex to this light stimulation.

    All of 44 patients were examined with two methods, SFT and O Glass. SFT method: The well known manual method to diagnose RAPD. O glass method:The device consists of camera and light sources.The red light was on and off for a 5 second interval. Then the white light was on for right eye and 3 seconds later the system captured an image. After 0.5 second the right light was off and the left light turned on, 3 seconds later the image was captured. The images were processed and analyzed using computerized software.

Sponsors and collaborators

Lead sponsor

Mashhad University of Medical Sciences

Other

Registry information

Official study title

New Automatic Portable Ophthalmology Device

Important dates

Study start
2015
Primary completion
2016
Study completion
2016
First posted
May 13, 2016
Registry last updated
Dec 29, 2016

OpenTrials presents study information sourced from ClinicalTrials.gov. The official registry record should be consulted for the latest information.

View the official ClinicalTrials.gov record (opens in a new tab)

This listing is for discovery and informational purposes only. It is not medical advice, does not guarantee that a study is recruiting, and does not determine eligibility. Contact the study team and a qualified healthcare professional when considering participation.

Published trials that share one or more normalized conditions with this study.