Skip to main content
OpenTrials
Completed

NCT Number: NCT01709786

Non-Invasive Hemoglobin Monitoring in Patients With Hemorrhage

This study will evaluate the accuracy of two rapid methods of measuring hemoglobin in patients with suspected hemorrhage. These methods will be compared with standard laboratory measurements.

Completed

Looking for future studies?

Notify Me

Key information

Age range

18 year–100 year

Sex eligibility

All sexes

Study type

Observational

Primary location

University Hospital

Cincinnati, Ohio, 45219, United States

About this study

Anemia and bleeding are major causes of morbidity and mortality in both surgical and nonsurgical patients. The current standard of care for monitoring patients at risk for bleeding is serial measurement of hemoglobin levels. At present, the photometric cyanmethemoglobin method is the most widely used technique for monitoring hemoglobin in the lab, and is currently the gold standard. However, this method has potential for delay before final results are obtained.

Immediate hemoglobin measurements are available with portable point-of-care devices such as the iSTAT, which can produce a measurement of hemoglobin concentration in less than 1 minute. Unfortunately, the accuracy of this device has been reported to vary with hemoglobin level, and as such may not be as accurate in detecting blood loss when compared with the gold standard of laboratory analysis.

Recently, a noninvasive, spectrophotometry-based monitoring technology has been developed. This novel technology measures the differential optical density of wavelengths of light passed through the finger in a method similar to conventional pulse oximetry. While some studies have reported that this device appears to be accurate in patients undergoing elective surgical procedures, more recent work suggests that this accuracy degrades with increased blood loss, lower oximeter signal quality and lower absolute Hgb values.

We will evaluate the accuracy of point-of-care and non-invasive SpHb measurements and utility of continuous hemoglobin monitoring in an intensive care unit setting. If these methods of rapid hemoglobin measurement can be validated in patients at risk for ongoing hemorrhage, use of this technology may result in earlier detection of ongoing hemorrhage, expedite appropriate treatment, and improve patient outcomes.

Who can participate

Healthy volunteers accepted: No

Only the study team can determine whether someone qualifies for participation.

Inclusion criteria

  • admitted to surgical intensive care unit (SICU), and
  • at risk of ongoing bleeding, and
  • requires serial CBC measurements

Exclusion criteria

  • < 18 years of age, or
  • a prisoner, or
  • unable to have pulse oximetry readings (due to injuries, burns, amputations, or related problems)

Treatment and study plan

Primary outcomes

  1. CBC Hemoglobin Measurement Compared to Non-invasive Radical-7 Measurement

    Time frame: n ≥ 1 measurements were taken each day. All measurements (n ≥ 7) from ICU Days 1-7 were used in Bland-Altman analysis, equally weighted.

    Whenever blood was drawn for laboratory measurement of serum hemoglobin, we used one drop of blood to make point-of-care measurements using the CDC and Radical-7 methods.

    For purposes of reporting outcomes measures, we took an equally weighted average of measurements on each device (i.e., all measurement occasions on all patients). These are the means reported in the Outcome Measures Data Table.

  2. CBC Hemoglobin Measurement Compared to Non-invasive iSTAT Measurement

    Time frame: n ≥ 1 measurements were taken each day. All measurements (n ≥ 7) from ICU Days 1-7 were used in Bland-Altman analysis, equally weighted.

    When blood was drawn for laboratory measurement of serum hemoglobin, one drop of blood was used to make point of care measurements using the CBC and iSTAT methods.

    For purposes of reporting outcomes measures, we took an equally weighted average of measurements on each device (i.e., all measurement occasions on all patients). These are the means reported in the Outcome Measures Data Table.

Sponsors and collaborators

Lead sponsor

University of Cincinnati

Other

Collaborators

  • United States Air Force

Registry information

Official study title

Non-Invasive Hemoglobin Monitoring in the Patient With Suspected Hemorrhage

Important dates

Study start
2012
Primary completion
2013
Study completion
2013
First posted
Oct 18, 2012
Registry last updated
Aug 9, 2016

OpenTrials presents study information sourced from ClinicalTrials.gov. The official registry record should be consulted for the latest information.

View the official ClinicalTrials.gov record (opens in a new tab)

This listing is for discovery and informational purposes only. It is not medical advice, does not guarantee that a study is recruiting, and does not determine eligibility. Contact the study team and a qualified healthcare professional when considering participation.