NCT Number: NCT03705299
Non-Invasive CVP Method to Standard CVP Method
This is a prospective, comparative, internally controlled device study that will enroll a total of 100 adult patients that require the placement of central venous catheter and CVP monitoring to assess volume status and cardiac preload.
Looking for future studies?
Notify MeKey information
Conditions
Age range
18 year and older
Sex eligibility
All sexes
Study type
Interventional
Phase
Not applicable
About this study
Standard CVP Method
The CVP line is inserted into the internal jugular vein central venous catheter or subclavian vein catheter and threaded to the point where the superior vena cava meets the inferior vena cava prior to entry into the right atrium of the heart. The catheter is hooked to a pressure bag which is pressurized to 200 cm water. This pressure allows the infusion of about 3ml of fluid each hour but keeps the line patent. The distal tip of the CVP line is connected to a transducer and taped to the arm at a level which is in line with the heart.
Research Procedure (non-invasive CVP Method)
A physiologic data acquisition system (Zoe Medical, Topsfield MA)) will be used to monitor CVP, upper arm impedance changes, and upper arm cuff pressures.
The subject's CVP pressure transducer will be zeroed at the level of the right atrium (midaxillary line) with the use of a carpenter's level.
A tetra-polar impedance configuration will be used to measure upper arm impedance. A pair of current-injecting electrodes will be placed in positions 1 & 4 and two sensing electrodes will be placed in positions 2 & 3 (where a traditional blood pressure cuff would be positioned, overlying the brachial-axillary vein system). These leads extend from the Zoe Medical device. (Figure 1 taken from Ward et al., in press).
Electrobioimpedance will be measured in the upper portion of an upper extremity using an electrobioimpedance amplifier contained within the Zoe Medical device. A constant current source (1mA, 100kHz) will be sent through the current electrode and the voltage drop between the two sensing electrodes will be amplified.
A blood pressure cuff attached to the Zoe Medical Device will be positioned over the two sensing electrodes.
The cuff pressure will be quickly inflated to a value higher than CVP but lower than the diastolic arterial pressure (40 mm Hg) and kept at that pressure for 45-60 seconds.
At the end of the inflation hold period, release the cuff pressure valve and open to atmosphere to allow rapid self-deflation
Repeat measures 3 times in each subject for a period of 10 minutes
If the subject requires volume resuscitation, diuresis, or are administered inotropic drugs, we will reevaluate CVP with the non-invasive method during these circumstances. CVP will be measured every 1 minute to capture rapid changes in CVP. The length of time that the investigators monitor CVP will depend on the length of time required to detect changes in CVP.
Who can participate
Healthy volunteers accepted: Yes
Only the study team can determine whether someone qualifies for participation.
Inclusion criteria
- Patients aged ≥ 18 years
- Require CVP monitoring in the supine position as part of routine care
- Central venous access obtained by either internal or external jugular, or subclavian vein
Exclusion criteria
- Patients aged < 18 years
- Patients who are pregnant
- Patients that will be in the prone position
- Central venous access obtained by the femoral vein
Treatment and study plan
Primary outcomes
-
Assessing if the NeuMeDx NICVP device is as effective as the standard invasive method at measuring CVP in patients that require central catheter placement and monitoring.
Time frame: 6 months
The primary objective of the study is to measure the CVP using the NeuMeDx NICVP (Non-Invasive Central Venous Pressure) device. The critical endpoint is the average of the average of three CVP, by calibrated forced transducer, and NICVP measurements in mmHg over a 10-minute period in a patient in the supine position.
Sponsors and collaborators
Lead sponsor
The Cooper Health System
Other
Collaborators
- NeuroDx Development
Registry information
Official study title
A Prospective, Comparative Study to Evaluate the Accuracy of a Non-Invasive Central Venous Pressure Method to the Standard CVP Method
Important dates
- Study start
- 2012
- Primary completion
- 2018
- Study completion
- 2018
- First posted
- Oct 15, 2018
- Registry last updated
- Oct 15, 2018
OpenTrials presents study information sourced from ClinicalTrials.gov. The official registry record should be consulted for the latest information.
View the official ClinicalTrials.gov record (opens in a new tab)This listing is for discovery and informational purposes only. It is not medical advice, does not guarantee that a study is recruiting, and does not determine eligibility. Contact the study team and a qualified healthcare professional when considering participation.
Related clinical trials
Published trials that share one or more normalized conditions with this study.
Pulse Pressure Variation vs. Central Venous Pressure for Fluid Management in Intracranial Tumor Surgery
NCT06776666
Brain Diseases, Brain Neoplasms
Denizli, Pamukkale, Turkey (Türkiye)
View Trial DetailsInternal Jugular Vein Cross Sectional Area
NCT01099254
Central Venous Pressure
Iowa City, Iowa, United States
View Trial DetailsPeripheral Venous Pressure as an Indicator of Central Venous Pressure in Post Operative Surgical Patients: A Randomised Controlled Trial
NCT05476666
Central Venous Pressure, Fluid/Volume Status Monitoring of Patient
Karachi, Sindh, Pakistan
View Trial DetailsCentral Venous Pressure Monitoring and Prognosis of High-risk Operating Patients
NCT04596332
Acute Kidney Injury, Central Venous Pressure
View Trial Details