MUNICH TRANSCATHETER MITRAL VALVE REPLACEMENT SYSTEM
DeviceThe replacement valve consists of a self-expanding, tri-leaflet, dry bovine -pericardial valve. The dry tissue allows the valve to be conveniently pre-loaded. The key characteristics of Munich valve are:
- Nitinol stent frame
- Bovine pericardium
- Polyester skirt
- Anchoring system allows anchoring the frame to the annulus
- Hooks are used to reduce the mobility of the native leaflets and decrease risk of embolization.
The Munich TMVR system design proposed for this clinical investigation represents a significant evolution from previous TMVR designs:
- Transfemoral / transseptal access
- Self-expanding
- Dry pericardium (can be pre-loaded)
- No anchors
- 27 Fr delivery catheter (
- Sizes: 40/48/55mm
- Height 30 mm The valve is available in 3 sizes with a 30mm profile and has been designed to reduce the complexity of implantation in comparison to other TMVR systems.