Comparison of SureCALL® EMG Labor Monitor® and Tocodynamometer to Intrauterine Pressure Catheter Measurement (IUPC) Measurement in Pregnant Women
NCT01282710
Pregnancy - Labor Monitoring
Phoenix, Arizona, United States
View Trial DetailsNCT Number: NCT01282723
This multi-reader multi-case study was designed to evaluate the performance of SureCALL® EMG Labor Monitor® (SureCALL®)to the predicate Tocodynamometer device (TOCO) using the performance of an Intrauterine Pressure Catheter device (IUPC) as the "gold standard".
Looking for future studies?
Notify Me15 year–40 year
Female
Observational
St.Joseph's Hospital and Medical Center, Phoenix, Arizona, United States
The SureCALL® EMG Labor Monitor® (SureCALL®)is a transabdominal electromyography monitor intended to measure intrapartum uterine activity. It is intended for use on term pregnant women who are in labor, with singleton pregnancies, using surface electrodes on the maternal abdomen. The device is intended for use by healthcare professionals in a clinical setting.
This multi-reader multi-case study was designed to evaluate the performance of SureCALL® to the predicate Tocodynamometer device (TOCO) using the performance of an Intrauterine Pressure Catheter device (IUPC) as the "gold standard".
This study involved 20 women at term, in labor, at three clinical sites. Each study subject was instrumented with three technologies for measuring uterine activity:
An assessment was obtained and quantified from masked, qualified clinical obstetrician readers for the evaluation of the equivalence of the three independent labor monitor devices by identifying in isolation the individual uterine contraction signals and the peak of each contraction signal produced by a full-term patient in term labor as independently collected by each labor monitor device.
Healthy volunteers accepted: No
Only the study team can determine whether someone qualifies for participation.
Inclusion criteria
Exclusion criteria
Time frame: 9 - 42 Minutes
The presence of contractions measured by the SureCALL®, the presence of contractions measured by the TOCO, and the presence of contractions measured by the IUPC were determined by independent Readers. The Odds Ratio of SureCALL® contractions to IUPC contractions was calculated, and the Odd Ratio of TOCO contractions to IUPC contractions was calculated. Reader Correspondence was determined by a General Linear Mixed Model.
Reproductive Research Technologies, LP
Industry
OpenTrials presents study information sourced from ClinicalTrials.gov. The official registry record should be consulted for the latest information.
View the official ClinicalTrials.gov record (opens in a new tab)This listing is for discovery and informational purposes only. It is not medical advice, does not guarantee that a study is recruiting, and does not determine eligibility. Contact the study team and a qualified healthcare professional when considering participation.
Published trials that share one or more normalized conditions with this study.
NCT01282710
Pregnancy - Labor Monitoring
Phoenix, Arizona, United States