Molecule Removal and Ease of Use: A Comparison of Two Different Dialyzers

The Polyflux HD-C4 (Gambro Renal Products) is a new dialyzer based on the existing Polyflux H membrane and dialyzer technology, but which achieves a performance comparable to Polyflux 210H in a significantly smaller device.

The objectives of this study are:

1. To compare and contrast small and large molecule removal by the Polyflux HD-C4 and Polyflux 210H dialyzers under conditions of routine hemodialysis; and, 2. To compare and contrast the ease of use of the Polyflux HD-C4 dialyzer with that of the Polyflux 210H under conditions of routine clinical use for hemodialysis.

Completed

Looking for future studies?

Notify Me

Key information

Age range

18 year and older

Sex eligibility

All sexes

Study type

Interventional

Phase

Not applicable

Primary location

DaVita Clinical Research Facility

Minneapolis, Minnesota, 55404, United States

Who can participate

Healthy volunteers accepted: No

Only the study team can determine whether someone qualifies for participation.

Inclusion criteria

  • Adults 18 years of age or older
  • Stable hemodialysis prescription using Polyflux 210H or Polyflux 21R for at least 2 months prior to study enrollment
  • Dialyzing through a native fistula or Gore-Tex graft.
  • Blood access must be able to provide a blood flow rate of 400 ml/min.

Exclusion criteria

  • Non-compliance with dialysis
  • Hematocrit less than 28%
  • Active Infection

Treatment and study plan

Polyflux 210H dialyzer

Device

Three consecutive treatments with the Polyflux 210H dialyzer. All study treatments will be performed in the standard hemodialysis mode.

Other names: Gambro Polyflux 210H dialyzer

Polyflux HD-C4

Device

Three consecutive treatments with the Polyflux HD-C4 dialyzer. All study treatments will be performed in the standard hemodialysis mode.

Other names: Gambro Revaclear

Primary outcomes

  1. Comparison of Urea Removal Under Conditions of Routine Hemodialysis.

    Time frame: 2 weeks = duration required for each subject to complete 6 consecutive dialysis treatments

    Urea removal is correlated with successful clinical outcomes. Kt/V is a way of measuring the delivered dose of dialysis where K=clearance of urea, t=treatment time and V=volume of body water. Single-pool (sp) Kt/V assumes that urea is removed from a single compartment in the human body during dialysis. However, because there are multiple compartments in the human body, rebound occurs following hemodialysis which lowers the Kt/V. Equilibrated (e) Kt/V is an equation that has been devised to predict the amount of rebound based on the ratio of K/V. Outcomes are posted for both spKt/V and eKt/V.

  2. A Comparison of Pre- to Post-dialysis Reduction of Small and Large Molecules Under Conditions of Routine Hemodialysis.

    Time frame: 2 weeks = duration required for each subject to complete 6 consecutive dialysis treatments

    Overall removal of urea, phosphorus and β2-microglobulin was determined from the pre- to post-dialysis change in plasma concentration and from the amount of solute recovered in the dialysate. This outcome measure is reported as the percentage of pre- to post-dialysis reduction of urea, phosphorus and β2-microglobulin.

Secondary outcomes

  1. Comparison of Dialyzer Ease of Use Between the Polyflux HD-C4 Dialyzer and the Polyflux 210H

    Time frame: 2 weeks = duration required for each subject to complete 6 consecutive dialysis treatments

    Assessment of blood side priming: 1=Very Easy 2=Acceptable 3=Difficult 4=Very Difficult / Assessment of dialysate side priming: 1=Perfect 2=Acceptable 3= Not Acceptable / Appearance of dialyzer fibers: 1=Very Good 2=Good 3=Poor 4=Very Poor / Appearance of dialyzer arterial header: 1=Very Good 2=Good 3=Poor 4=Very Poor / Appearance of venous header: 1=Very Good 2=Good 3=Poor 4=Very Poor /

Sponsors and collaborators

Lead sponsor

Vantive Health LLC

Industry

Collaborators

  • Baxter Healthcare Corporation
  • Gambro Renal Products, Inc.

Registry information

Official study title

Urea, Phosphate and b2-Microglobulin Removal and Ease of Use: Comparing the Polyflux HD-C4 With the Polyflux 210H

Important dates

Study start
2008
Primary completion
2008
Study completion
2009
First posted
Mar 14, 2008
Registry last updated
Jul 14, 2025

OpenTrials presents study information sourced from ClinicalTrials.gov. The official registry record should be consulted for the latest information.

View the official ClinicalTrials.gov record (opens in a new tab)

This listing is for discovery and informational purposes only. It is not medical advice, does not guarantee that a study is recruiting, and does not determine eligibility. Contact the study team and a qualified healthcare professional when considering participation.

Published trials that share one or more normalized conditions with this study.