NCT Number: NCT02004873

Micra Transcatheter Pacing Study

The purpose of this clinical study is to evaluate the safety and efficacy of the Micra Transcatheter Pacing System and to assess long term performance.

Completed

Looking for future studies?

Notify Me

Key information

Age range

18 year and older

Sex eligibility

All sexes

Study type

Interventional

Phase

Not applicable

Primary location

Princess Alexandria Hospital, Woolloongabba, Queensland, Australia

Loading trial locations.

About this study

The study is being conducted in 56 sites located around the world, including 30 sites in the United States (U.S.).

Up to 780 subjects were planned to be enrolled (i.e. consented) to allow up to 720 subjects to be implanted enabling at least 600 subjects to be followed for at least 12-months post-implant.

There are two primary objectives in this study (one safety and one efficacy). The primary safety objective of the study is to evaluate major complications related to the Micra system or procedure. The primary safety endpoint was pre-specified to be evaluated at 6-months (183 days) post-implant. The primary efficacy objective, Micra pacing capture thresholds, was also pre-specified to be evaluated six months post-implant. This study also has a separate long term safety objective that will provide additional long-term safety data following potential regulatory submissions.

All subjects will be followed until official study closure (official study closure is defined as when Medtronic and/or FDA requirements have been satisfied per the Clinical Investigational Plan and/or by a decision by Medtronic or regulatory authority).

Additionally, the Micra Accelerometer Sensor Sub-Study (MASS) is being conducted at 4 centers already active in the Micra study in Austria, Spain, Hungary and France. Forty (40) subjects were enrolled in the Sub-Study, and enrollment in the Sub-Study was complete as of March 31, 2016. The purpose of the Sub-Study is to test feasibility for future enhancements in the Micra device functionalities.

An algorithm was designed to sense the atrial contraction using the Micra accelerometer.

Who can participate

Healthy volunteers accepted: No

Only the study team can determine whether someone qualifies for participation.

Inclusion criteria

  • Subjects who have a Class I or II indication for implantation of a single chamber ventricular pacemaker according to ACC/AHA/HRS 2008 guidelines and any national guidelines
  • Subjects who are able and willing to undergo the study requirements and are expected to be geographically stable for the duration of the follow-up.
  • Subjects who are at least 18 years of age (or older, if required by local law).

Exclusion criteria

  • Subjects who are entirely pacemaker dependent (escape rhythm <30 bpm). (Please note: Subjects who are entirely pacemaker dependent (escape rhythm <30 bpm) can now be included in the study. Medtronic notified all sites on July 23, 2014 that the restriction against pacemaker dependent subjects was lifted, following review of the Early Performance Assessment.)
  • Subject has an existing or prior pacemaker, ICD or CRT device implant.
  • Subject has unstable angina pectoris or has had an acute myocardial infarction (AMI) in the 30 days prior to eligibility assessment.
  • Subjects with current implantation of neurostimulator or any other chronically implanted device which uses current in the body. Note that a temporary pacing wire is allowed.
  • Subjects with a mechanical tricuspid valve, implanted vena cava filter, or left ventricular assist device (LVAD).
  • Subjects who are morbidly obese and physician believes telemetry communication of ≤5 inches (12.7 cm) could not be obtained with programmer head.
  • Subjects whose femoral venous anatomy is unable to accommodate a 23 French introducer sheath or implant on the right side of the heart (for example, due to obstructions or severe tortuosity) in the opinion of the implanter.
  • Subjects who are considered as unable to tolerate an urgent sternotomy
  • Subjects with a known intolerance to Nickel-Titanium (Nitinol) Alloy.
  • Subjects for whom a single dose of 1.0mg dexamethasone acetate may be contraindicated.
  • Subjects with a life expectancy of less than 12- months.
  • Subjects who are currently enrolled or planning to participate in a potentially confounding drug or device trial during the course of this study. Coenrollment in concurrent trials is only allowed when document pre-approval is obtained from the Medtronic study manager.
  • Pregnant women, or women of child bearing potential and who are not on a reliable form of birth control.
  • Subjects with exclusion criteria required by local law (e.g. age, breast feeding, etc.).

Treatment and study plan

Micra Pacemaker Implant

Device

Primary outcomes

  1. Major Complications

    Time frame: Implant to 6 Months Post Implant

    Micra system and/or procedure related major complication free rate at 6-months post-implant.

  2. Pacing Capture Threshold

    Time frame: 6 Months Post Implant

    Subjects that have an adequate pacing capture threshold (PCT) at the 6-month post-implant visit, which is defined as PCT <=2 volts at 0.24 ms pulse width and the increase in PCT from implant to 6 months <=1.5 volts. The pacing capture threshold is the minimal electrical stimulus required to produce consistent cardiac depolarization. It is the minimum amount of energy that is required for a pacemaker to pace the heart.

Secondary outcomes

  1. Ventricular Capture Management Threshold

    Time frame: 6 Months Post Implant

    Subjects that have a ventricular capture management threshold (VCMT) that is within 0.5 Volts of the manual (auto decrement) PCT (at 0.24 ms pulse width) at the 6-month post-implant visit. The VCMT is an automatically measured pacing capture threshold that is measured by the Micra device's pacing algorithm. In contrast, the manual (auto decrement) pacing capture threshold is measured by the clinician during a study visit.

  2. Rate Response Operation of Micra

    Time frame: 3 Months and 6 Months Post Implant (combined analysis)

    Assessment of whether the Micra sensor-indicated rate derived from the input of the accelerometer during the Minnesota Pacemaker Response Exercise Protocol (M-PREP) treadmill test conducted at the 3-month and 6-month follow-up visits was proportional to the workload. The sensor-indicated rate (in min^-1) and workload (in METS) were normalized for each subject relative to their minimum and maximum possible values so the normalized values have a minimum possible value of zero and a maximum possible value of 1. These normalized values were used in a random effect linear regression model to assess the relationship between the sensor-indicated rate and workload via estimation of the Kay-Wilkoff slope parameter. The tests at 3-month and 6-month visits were combined in one analysis.

Sponsors and collaborators

Lead sponsor

Medtronic Cardiac Rhythm and Heart Failure

Industry

Registry information

Important dates

Study start
2013
Primary completion
2015
Study completion
2017
First posted
Dec 9, 2013
Registry last updated
Jan 17, 2018