The RANE Center for Venous and Lymphatic Diseases
Jackson, Mississippi, 39216, United States
NCT Number: NCT06944041
We used a new and easy-to-measure iPad 3D scanner to measure swelling in lower limbs because current technology is not easily available or affordable. We want to see how reliable the measurements are and what might affect how consistent they are.
Looking for future studies?
Notify MeAll sexes
Observational
Jackson, Mississippi, 39216, United States
Twenty limbs from ten healthy volunteers were scanned using an iPad-based Structure Sensor and software. Initial scans followed standard manufacturer instructions, but high variance rendered data unsuitable for clinical use. To improve accuracy, a standardized scanning protocol was developed, incorporating anatomical calibration, scanning distance standardization, and scanning time control. A 254-mm calf segment was defined using a 3D marker placed on the medial malleolus to ensure consistent volume measurement. The scanning distance was fixed between 50-59 cm to reduce zoom parallax errors, and scans were conducted after 3 PM to minimize diurnal volume fluctuations. Multiple technicians performed repeat scans on the same limb to assess intra-observer and inter-observer scan reliability.
Healthy volunteers accepted: Yes
Only the study team can determine whether someone qualifies for participation.
Inclusion criteria
Exclusion criteria
The advent of Apple iPad based 3D scanner opens a new avenue for measuring limb volume. The latest ambient light scanner model was utilized for the current study. 3D scanning technology has gained broad industrial use, including the manufacturing of high-precision tools. A commercially available software package was used for limb volumetry.
Time frame: During Procedure
The scanning protocol and the required modifications of "out of the box" instructions for optimal clinical application are described; The 10 inch (254mm) long limb segment scanned must be anatomically constant. We chose the medial malleolus as the lower border of the scan. The scannable limb volume is 254 mm long, measured up from the malleolus.
The RANE Center for Venous and Lymphatic Diseases at St. Dominic's Hospital
Other
Limb Volumetry Using iPad Based 3D Scanner for Assessment of Edema
OpenTrials presents study information sourced from ClinicalTrials.gov. The official registry record should be consulted for the latest information.
View the official ClinicalTrials.gov record (opens in a new tab)This listing is for discovery and informational purposes only. It is not medical advice, does not guarantee that a study is recruiting, and does not determine eligibility. Contact the study team and a qualified healthcare professional when considering participation.
Published trials that share one or more normalized conditions with this study.
NCT07414615
Hemic and Lymphatic Diseases, Lymphatic Diseases
Istanbul, Turkey (Türkiye)
View Trial DetailsNCT06905639
Behavior, Breast Diseases
Ponferrada, León, Spain
View Trial DetailsNCT02506530
Breast Diseases, Breast Neoplasms
Toulouse, Midi Pyrenees, France
View Trial DetailsNCT07612865
Cardiovascular Diseases, Cervical Cancer
Ürümqi, Xinjiang Uygur Autonomous Region, China
View Trial Details