McGill University Health Centre
Montreal, Quebec, H4A 3J1, Canada
NCT Number: NCT05913323
The purpose the study is to determine the agreement between the information obtained about lung mechanics (lung function) from the Rapid Expiratory Occlusion Monitor (REOM) handheld portable device with that obtained by conventional oscillometry as well as by pulmonary function testing (PFT), in patients with both 'mild' as well as 'very severe' forms of chronic obstructive pulmonary disease (COPD). This study also intends to study the user experience with this handheld portable device.
Looking for future studies?
Notify Me40 year and older
All sexes
Observational
Montreal, Quebec, H4A 3J1, Canada
The primary objective of this cross-sectional study is to determine the concordance and agreement between the parameters obtained from the REOM, Reo1 and Reo2, and the conventional resistance parameters obtained by the tremoflo oscillometer device (R19 and R5, respectively) in adults with COPD confirmed by diagnostic pulmonary function testing (PFT).
The secondary objectives are:
Hypothesis:
We hypothesize in the adult COPD patient population that the REOM parameter Reo1 and tremoflo device parameter R19 will correlate closely, and that the REOM parameter Reo2 and tremoflo device parameter R5 will correlate closely. We also hypothesize that the REOM, in particular, Reo2, will be able to distinguish between 'mild' and 'very severe' COPD, and that participants will report a positive user experience with the REOM.
Healthy volunteers accepted: No
Only the study team can determine whether someone qualifies for participation.
Inclusion criteria
Exclusion criteria
Device for which Health Canada approval (ITA Authorization) has been obtained for use in the present study.
tremoflo device-100, Thorasys Thoracic Medical Systems Inc., Montreal, Canada
Standard-of-care respiratory test.
Time frame: Single-visit study: all tests performed at one single visit.
Correlation will be tested using the Spearman correlation coefficient (r) between REOM-obtained (Reo1 and Reo2) and tremoflo-obtained (R19 and R5) resistance measurements.
Time frame: Single-visit study: all tests performed at one single visit.
Agreement will be tested using the Bland-Altman test between REOM-obtained (Reo1 and Reo2) and tremoflo-obtained (R19 and R5) resistance measurements.
Time frame: Single-visit study: all tests performed at one single visit.
The t-test (between 2 independent means) will be used to determine the discriminative capacity of the REOM to distinguish between 'mild' and 'very severe' COPD.
Time frame: Single-visit study: all tests performed at one single visit.
Each participant will complete a 'participant satisfaction questionnaire' as well as the validated System Usability Scale (SUS) for each test (REOM, standard oscillometry, pulmonary function test). Data will be presented descriptively.
McGill University Health Centre/Research Institute of the McGill University Health Centre
Other
Measuring Lung Mechanics in Patients with Chronic Obstructive Pulmonary Disease (COPD) Using the Rapid Expiratory Occlusion Method (REOM) Handheld Portable Oscillometer: a Cross-Sectional Study.
OpenTrials presents study information sourced from ClinicalTrials.gov. The official registry record should be consulted for the latest information.
View the official ClinicalTrials.gov record (opens in a new tab)This listing is for discovery and informational purposes only. It is not medical advice, does not guarantee that a study is recruiting, and does not determine eligibility. Contact the study team and a qualified healthcare professional when considering participation.
Published trials that share one or more normalized conditions with this study.
NCT03927820
Asthma, Bronchial Diseases
Nashville, Tennessee, United States
View Trial DetailsNCT06614959
Bronchial Diseases, Bronchitis
Montreal, Quebec, Canada
View Trial DetailsNCT04754490
COPD, Chronic Disease
Abercynon, Wales, United Kingdom
View Trial DetailsNCT06336252
Arthritis, Basal Ganglia Diseases
Copenhagen, Denmark
View Trial Details