Common Anesthetic Management
OtherStandard intraoperative monitoring including BIS, pulse oximetry, temperature, and anesthesia workstation-derived respiratory variables.
- Arterial blood gas sampling after intubation, before pneumoperitoneum, after pneumoperitoneum/positioning, hourly during pneumoperitoneum, at the end of pneumoperitoneum in supine position, and before extubation.
- Routine safety limits on the anesthesia machine: end-tidal carbon dioxide upper alarm 45 mmHg, inspired oxygen lower alarm 35%, inspired carbon dioxide upper alarm 3 mmHg.
- Routine device self-test each morning and between patients.
- Minimal dead space strategy with avoidance of unnecessary circuit extension.
- Close monitoring of soda lime; replacement if inspired carbon dioxide reaches 3 mmHg even without obvious color change.
- If clinically necessary because of blood gas deterioration, BIS changes, or any safety concern, fresh gas flow may be increased and the participant may be withdrawn from the protocol intervention.