NCT Number: NCT06535152
Lifelight is a novel, calibration-free, software-only medical device that contactlessly measures pulse rate and blood pressure by measuring remote photoplethysmography (rPPG) signals of a patient's face using a standard smartphone or tablet camera.
By conducting this validation study we aimed to demonstrate the accuracy of Lifelight for classifying the full range of clinically relevant blood pressures as hypertensive or non-hypertensive and demonstrate its accuracy for measuring pulse rate and the blood pressure of people with blood pressures relevant to Stage 1 hypertension.
Looking for future studies?
Notify Me18 year–85 year
All sexes
Interventional
Not applicable
Louisville, Colorado, 80027, United States
Lifelight was investigated in a study that followed the data collection and data analysis methodology outlined in ISO 81060-2:2018/AMD 1:2020 "Non-invasive Sphygmomanometers - Part 2: Clinical investigation of automated measurement type". This validation study was conducted by independent laboratory Element Materials Technology Boulder (formerly Clinimark), a global leader for clinical testing of vital sign data for medical devices and consumer wearable products. The study generated data from 85 people aged 18-85 with the wide-ranging distribution of blood pressures specified in ISO 81060-2:2018/AMD 1:2020.The accuracy of Lifelight's blood pressure measurements was assessed by comparing Lifelight's measurements with measurements made by dual-observer manual auscultation using the same-arm sequential method specified in ISO 81060-2:2018/AMD 1:2020. The accuracy of Lifelight's pulse rate measurements was assessed by comparing Lifelight's measurements with concurrent ECG-derived heart rate values. Data was collected in an Independent Review Board-approved study conducted by independent laboratory Clinimark (now Element) in the USA and repurposed for the testing of the new improved model in the latest version of the Lifelight® EA device. The data collected consists of 85 participants, with three readings each. The participants were resting before their measurements were taken to ensure stable blood pressure. For each Lifelight measurement, two sets of blood pressure measurements were taken using dual-observer manual auscultation. The first measurement was taken before the Lifelight measurement and the second just after. If there was too much variation between the before and after measurement, or between the two independent observers, the measurement was discarded and repeated. For this validation procedure, the measurements taken during the original study were re-processed in the new updated algorithm.
Healthy volunteers accepted: Yes
Only the study team can determine whether someone qualifies for participation.
Inclusion criteria
Exclusion criteria
Participants could choose to withdraw themselves from the study without prejudice or they could be withdrawn by study investigators for predetermined reasons. Data excluded from the analysis was documented with justifications.
Lifelight is a artificial-intelligence software application that enables completely contactless spot measurements (i.e., one-off measurements without prior calibration) of PR and BP. Based on the science of remote photoplethysmography (rPPG), Lifelight works by detecting tiny changes in the colour of facial skin that occur every time the heart beats. All that is required to generate PR and BP measurements is that the software application is downloaded onto a computer device (e.g., smartphone or tablet) with a standard camera, and that the patient's face stays in the line of sight of the camera for up to 60 seconds.
Blood Pressure Measurement
Time frame: Up to 3 Months
Demonstrate the accuracy of Lifelight for classifying the full range of clinically relevant blood pressures as hypertensive or non-hypertensive.
Time frame: Up to 3 Months
Demonstrate its accuracy for measuring pulse rate (bpm)
Time frame: Up to 3 Months
the blood pressure (mmHg) of people with blood pressures relevant to Stage 1 hypertension
Xim Limited
Industry
CLINIMARK Clinical Investigation Plan Lifelight Non-Invasive Blood Pressure Validation Study PR 2021-443
OpenTrials presents study information sourced from ClinicalTrials.gov. The official registry record should be consulted for the latest information.
View the official ClinicalTrials.gov record (opens in a new tab)This listing is for discovery and informational purposes only. It is not medical advice, does not guarantee that a study is recruiting, and does not determine eligibility. Contact the study team and a qualified healthcare professional when considering participation.
Published trials that share one or more normalized conditions with this study.
NCT05610059
Aging, Behavior
Tucson, Arizona, United States
View Trial DetailsNCT02439775
Cardiovascular Diseases, Hypertension
Huntsville, Alabama, United States
View Trial DetailsNCT07735468
Behavior, Cardiovascular Diseases
Osuna, Sevilla, Spain
View Trial DetailsNCT00837122
Cardiovascular Diseases, Diabetes
Accra, Ghana
View Trial Details