Electrophysiological Study
ProcedureSubjects will receive pacing from one right ventricular lead and one left ventricular catheter/lead with multiple LV pacing spots during electrophysiological study procedure.
NCT Number: NCT02914457
The SYNSEQ study intends to assess the positive left ventricular dP/dt max achieved by MultiSpot LV pacing (either simultaneously or sequentially) in comparison to the response achieved by the current (standard) BiV pacing configuration in patients indicated/recommended for cardiac resynchronization therapy.
Looking for future studies?
Notify Me18 year and older
All sexes
Interventional
Not applicable
Klinika Choroby Wieńcowej, Warsaw, Poland
The following pacing configurations will be evaluated.
Biventricular pacing (BiV):
Pacing will be performed on one LV electrode pair, (at 3 different longitudinal locations), and on the tip of the RV-lead. In total, three different pacing BiV settings will be evaluated. Configuration 1: RV + LV lateral Apex, Configuration 2: RV + LV lateral Mid, Configuration 3: RV + LV lateral Base (Reference: Standard CRT)
MultiSpot simultaneous LV-ventricular pacing (MultiSpot-SYN):
Pacing will be performed on 3 electrodes on the LV wall, placed at different longitudinal locations, and on the tip of the RV-lead simultaneously. Configuration 4: RV + LV lateral Apex + LV lateral Mid + LV lateral Base
MultiSpot sequential LV-ventricular pacing (MultiSpot-SEQ):
3 electrodes on the LV wall will be paced sequentially. The RV electrode will be paced simultaneously with last paced LV electrode.The timing-sequence and the amount of spots will depend on the electrical delays measured during the experiments. Configuration 5: LV lateral Apex => LV lateral Mid => LV lateral Base + RV
Healthy volunteers accepted: No
Only the study team can determine whether someone qualifies for participation.
Inclusion criteria
Exclusion criteria
Subjects will receive pacing from one right ventricular lead and one left ventricular catheter/lead with multiple LV pacing spots during electrophysiological study procedure.
Time frame: Participants were followed for the time of the EP procedure, which had a median duration of 48 min
LV dP/dt max is a measurement of the initial velocity of myocardial contraction and is derivative of the LV-pressure. The percentage changes correspond to a percentage change between a pacing configuration (pacing on, e.g., Multispot pacing) and baseline (LV pacing off). There are several repetitions of pacing off and on for each pacing configuration. For one repetition, the percentage change is determined as ([median dP/dt max during pacing On] - (median baseline dP/dt max during pacing Off])/[median dP/dt max during pacing Off]. From all percentage changes for a given pacing configuration and subject, a regression analysis is performed to determine the regression predicted highest percentage change. The presented percentage change is the average over all subjects.
Time frame: Participants were followed for the time of the EP procedure, which had a median duration of 48 min
Measured by invasive arterial blood line connected to a sensitive membrane displacement sensor. Systolic blood pressure (SBP) and diastolic blood pressure (DBP) were collected.The percentage changes correspond to a percentage change between a pacing configuration (pacing on, e.g., Multispot pacing) and baseline (LV pacing off). There are several repetitions of pacing off and on for each pacing configuration. For one repetition, the percentage change is determined as ([median pressure during pacing On] - (median pressure during pacing Off])/[median pressure during pacing Off]. Correlation will be summarized over all pacing configurations and time points since the interest is in the overall correlation between LV dP/dt max and other measurements, not in the correlation per pacing configuration or per time point. The general linear model as described in Blank & Altman, Biometrical Journal 310 (1995), p 446, was used to determine the correlation.
Time frame: Participants were followed for the time of the EP procedure, which had a median duration of 48 min
Acquired through finger volume clamp
Time frame: Participants were followed for the time of the EP procedure, which had a median duration of 48 min
Derived from intra-cardiac leads (invasive) and surface electrodes (non-invasive) respectively. The Q-LV interval is defined as the time from the onset of the QRS width of the surface ECG to the first large positive or negative peak of the LV electrogram (EGM) during a cardiac cycle. The Q-LV ratio will be calculated as Q-LV/QRS duration. Correlation will be summarized over all pacing configurations since the interest is in the overall correlation between LV dP/dt max and other measurements, not in the correlation per pacing configuration. The general linear model as described in Blank & Altman, Biometrical Journal 310 (1995), p 446, was used to determine the correlation between % change LV dP/dt max and Q-LV ratio.
Time frame: Participants were followed for the time of the EP procedure, which had a median duration of 48 min
Derived from surface electrodes (non-invasive). The percentage change is determined as ([QRS width during pacing On] - (QRS width during pacing Off])/[QRS width during pacing Off]. The correlation between % change LV dP/dt max and % change QRS width will be summarized over all pacing configurations since the interest is in the overall correlation between LV dP/dt max and other measurements, not in the correlation per pacing configuration or per time point. The general linear model as described in Blank & Altman, Biometrical Journal 310 (1995), p 446, was used to determine the correlation.
Medtronic Cardiac Rhythm and Heart Failure
Industry
Left Ventricular Synchronous Versus Sequential MultiSpot Pacing for CRT
Acronym: SYNSEQ
OpenTrials presents study information sourced from ClinicalTrials.gov. The official registry record should be consulted for the latest information.
View the official ClinicalTrials.gov record (opens in a new tab)This listing is for discovery and informational purposes only. It is not medical advice, does not guarantee that a study is recruiting, and does not determine eligibility. Contact the study team and a qualified healthcare professional when considering participation.
Published trials that share one or more normalized conditions with this study.
NCT03446651
Cardiovascular Diseases, Heart Diseases
New Haven, Connecticut, United States
View Trial DetailsNCT05806606
Behavior, Cardiovascular Diseases
Hong Kong
View Trial DetailsNCT06847438
Cardiovascular Diseases, Heart Diseases
Atlanta, Georgia, United States
View Trial DetailsNCT07721909
Cardiovascular Diseases, Heart Diseases
Muscat, Oman
View Trial Details