Hôpital La Pitié Salpêtrière
Paris, 75013, France
NCT Number: NCT01297283
The leadless electrocardiogram (LECG) is a new technology incorporated into the Consulta CRT-P to obtain an ECG signal from the Consulta CRT-P similar to a surface ECG obtained from the device programmer (PECG) or an external ECG machine.
The purpose of the study is to obtain more data on leadless ECG to determine whether LECG during standard CRT follow-up can indeed adequately replace surface ECG and to evaluate if it can be used during remote follow-up evaluations.
Looking for future studies?
Notify Me18 year and older
All sexes
Interventional
Not applicable
Paris, 75013, France
The leadless electrocardiogram (LECG) is a new technology incorporated into the Consulta CRT-P to obtain an ECG signal from the Consulta CRT-P similar to a surface ECG obtained from the device programmer (PECG) or an external ECG machine. The LECG signals are measured from three electrodes mounted on the outside of the pacemaker housing and provides an electrical far field signal of the electrical activity of the heart. The LECG provides three ECG channels as different projections of the electrical activity of the heart, similar to the surface ECG. Clinical interest of LECG is threefold. First, ECG recordings are routinely used to perform pacemaker and cardiac resynchronization systems in-office follow-up mainly to determine pacing thresholds. Connection of ECG electrodes to the patient as well as the time needed to acquire an acceptable ECG signal during routine follow-up could be saved using LECG which would make follow-up easier and less time consuming. Secondly, connecting ECG electrodes requires the patient to be present at the clinic for the follow-up. Use of LECG in conjunction with a transmitting system will allow remote patient follow-up. In that case, correct ventricular capture confirmation by the LECG is of key importance. Finally, LECG stored in device memory at the time of an arrhythmia episode occurrence, can help better classify it.
The following factors might influence the quality of the LECG and/or the axis of the LECG:
Healthy volunteers accepted: No
Only the study team can determine whether someone qualifies for participation.
Inclusion criteria
Exclusion criteria
Measurement of the pacing threshold with the support of a leadless ECG
Pacing threshold measurements are done with the support of the programmer ECG
Time frame: 30 to 120 days
During CRT-P standard follow-up, ECG is used to determine atrial, left and right ventricular pacing thresholds. As primary endpoint, we will consider the proportion of patients for which for all leads LECG provides pacing threshold values that are clinically equivalent to those obtained with PECG taken as reference. The analysis will be performed on data collected at the 1-Month Follow-Up visit when the device pocket healing process is completed. Clinical equivalence will be defined as the LECG threshold values being no more than 0.5 volts different from the PECG threshold values.
Time frame: 30 to 120 days
The investigator will simulate loss of capture (LOC) at 1-Month Follow-Up visit by printing strips of LOC in both ventricular leads, LOC in LV lead only, LOC in RV lead only, no LOC. One strip randomly selected among them by the study manager will be submitted to an independent reviewer. With help of the PHD template LECG strips (intrinsic, RV paced, LV paced, BiV paced) of this patient, he/she will determine which lead is capturing.
The endpoint is the proportion of correct classifications of ventricular capture done by then independent reviewer of LECG.
Time frame: 30 to 120 days
Two parameters will be used to evaluate LECG changes between PHD and 1-Month on the same LECG vector: intrinsic R wave amplitude and P waves visibility (selection the LECG vector at PHD with best combination of the highest R wave and most visible P wave).
R wave amplitude measurements as well as P wave visibility assessment will be done by the independent reviewer.
The endpoints evaluated are:
Time frame: 30 to 120 days
The two following parameters will be used to compare the quality of LECG in new implants versus device replacements at PHD and 1-Month on the same LECG vector:
R wave amplitude measurements as well as P wave visibility assessment will be done by the independent reviewer.
The endpoints will be:
Time frame: 30 to 120 days
The intrinsic R wave amplitude and P waves visibility will be taken to evaluate the effect of posture changes and artifact-inducing maneuvers on the quality of LECG at the 1-Month Follow-Up visit (LECG vector with best combination of the highest R wave and most visible P wave).
The endpoints will be R wave amplitude and P wave visibility in different positions (meaning visible or not visible in both positions).
R wave changes and proportion of patients with stable P waves visibility at lying position versus other positions will be calculated by an independent reviewer.
Time frame: 30 to 120 days
The following factors will be evaluated to investigate whether they influence the LECG quality:
The endpoints will be:
Medtronic Cardiac Rhythm and Heart Failure
Industry
Leadless ECG Evaluation Study - Prospectively, Randomized, Cross-over, Multi-center, Interventional, Post-market Release Study to Evaluate the LECG System.
Acronym: LECG
OpenTrials presents study information sourced from ClinicalTrials.gov. The official registry record should be consulted for the latest information.
View the official ClinicalTrials.gov record (opens in a new tab)This listing is for discovery and informational purposes only. It is not medical advice, does not guarantee that a study is recruiting, and does not determine eligibility. Contact the study team and a qualified healthcare professional when considering participation.
Published trials that share one or more normalized conditions with this study.
NCT03446651
Cardiovascular Diseases, Heart Diseases
New Haven, Connecticut, United States
View Trial DetailsNCT05806606
Behavior, Cardiovascular Diseases
Hong Kong
View Trial DetailsNCT06847438
Cardiovascular Diseases, Heart Diseases
Atlanta, Georgia, United States
View Trial DetailsNCT07721909
Cardiovascular Diseases, Heart Diseases
Muscat, Oman
View Trial Details