Gold tip catheter
DeviceRadiofrquency ablation using gold tip catheter
NCT Number: NCT00326001
The purpose of this study is to demonstrate the advantage of using a gold alloy tip electrode over a platinum/iridium alloy tip electrode in ablation of the cavotricuspid isthmus in patients with atrial flutter.
Looking for future studies?
Notify Me18 year and older
All sexes
Interventional
Phase 4
Institute of Clinical and Experimental Medicine, Prague, Czechia
Transvenous catheter ablation has become the therapy of choice for patients with recurring, isthmus-dependent right atrial flutter. Achieving bidirectional conduction block in the cavotricuspid isthmus is decisive for both acute and long-term therapy success and essentially depends on the selected ablation method and the lesion size. By using an 8 mm tip electrode instead of a conventional 4 mm electrode, deeper lesions can be made, thus significantly reducing the required number of energy applications for achieving a bidirectional conduction block. Experimental studies have proven that using an ablation electrode made of gold alloy allows the creation of deeper lesions than with conventional platinum-iridium electrodes. Due to the greater heat conductivity of the gold alloy as opposed to platinum-iridium, the cooling of the ablation electrode is improved and more electric energy can be transmitted to the tissue at identical temperatures.
The combination of both technologies in the form of an 8 mm-long gold electrode anticipates that the lesion depth required for an isthmus block can be achieved more quickly in comparison to the platinum-iridium electrode.
Healthy volunteers accepted: No
Only the study team can determine whether someone qualifies for participation.
Inclusion criteria
Exclusion criteria
Late Exclusion Criteria:
Radiofrquency ablation using gold tip catheter
Radiofrequency ablation using platinum-iridium tip catheter
Time frame: ablation procedure
Cumulative amount of time current is flowing through the catheter tip. The current (in the radiofrequency range) is applied to ablate the cavotricuspid isthmus in the right atrium.
Time frame: ablation procedure
Delivery of radiofrequency current was repeated until a cavotricuspid isthmus (CTI) conduction block was detected. The final bidirectional CTI block test (well documented in the literature) was performed 20 minutes after the last radiofrequency current delivery to assess ablation success (Y/N).
Positive final bidirectional cavotricuspid isthmus condution block test means "ablation successful".
Negative final bidirectional cavotricuspid isthmus condution block test means "ablation unsuccessful"; ablation should be continued until success or terminated and classified as unsuccess.
Time frame: 6 months after ablation
No recurrence of atrial flutter after ablation
Time frame: ablation procedure
Char or coagulum formation on the catheter tip
Biotronik SE & Co. KG
Industry
AURUM 8 - Ablation of the Cavotricuspid Isthmus in Patients With Atrial Flutter Using an 8 mm Gold Alloy Tip Electrode
OpenTrials presents study information sourced from ClinicalTrials.gov. The official registry record should be consulted for the latest information.
View the official ClinicalTrials.gov record (opens in a new tab)This listing is for discovery and informational purposes only. It is not medical advice, does not guarantee that a study is recruiting, and does not determine eligibility. Contact the study team and a qualified healthcare professional when considering participation.
Published trials that share one or more normalized conditions with this study.
NCT05635864
Arrhythmias, Cardiac, Atrial Fibrillation
Hanover, Lower Saxony, Germany
View Trial DetailsNCT03231826
Arrhythmias, Cardiac, Arrythmia
Tampere, Pirkanmaa, Finland
View Trial DetailsNCT03376438
Arrhythmias, Cardiac, Atrial Flutter
Phoenix, Arizona, United States
View Trial DetailsNCT04210622
Arrhythmias, Cardiac, Atrial Fibrillation
Vilnius, Lithuania
View Trial Details