INVSENSOR00013 Respiratory Rate Clinical Performance Study

The objective of this prospective study is to compare the noninvasive respiration from pleth (RRp) performance of the Masimo INVSENSOR00013 device against the respiratory rate measured by capnography (RRref), Masimo's RAM technology and the manual annotation.

Completed

Looking for future studies?

Notify Me

Key information

Conditions

Age range

18 year–70 year

Sex eligibility

All sexes

Study type

Interventional

Phase

Not applicable

Primary location

Masimo Clinical Lab

Irvine, California, 92618, United States

Who can participate

Healthy volunteers accepted: Yes

Only the study team can determine whether someone qualifies for participation.

Inclusion criteria

  • 18 to 70 years old
  • Physical status of ASA I or II
  • Must be able to read and communicate in English
  • Has signed all necessary related documents, e.g. written informed consent, confidentiality agreement
  • Passed health assessment screening
  • Negative pregnancy test for female subjects of child bearing potential

Exclusion criteria

  • Physical status of ASA III, IV, or V
  • Subject has any medical condition which in the judgement of the investigator, renders them inappropriate for participation in the study
  • Inability to tolerate sitting still or minimal movement for at least 30 minutes
  • Positive pregnancy test for female subjects
  • Refusal to take pregnancy test for women of child bearing potential
  • Nursing female subjects
  • Refusal of male subjects to agree to shave hair off areas where the sensor will be applied (neck) when deemed necessary
  • Subjects wearing acrylic nails or subjects refusing to remove nail polish
  • Subjects who have a nail deformity on the measurement finger
  • Subjects who do not have adequate skin integrity on the measurement finger
  • Excluded at the Principal Investigator's discretion

Treatment and study plan

INVSENSOR00013

Device

Investigational pulse oximeter device that will be placed on the subject's finger. This will be used as a spotcheck device

Primary outcomes

  1. RRp Arms of Sensor Accuracy

    Time frame: 1-5 hours

    Accuracy will be determined by comparing the noninvasive respiratory rate by pleth (RRp) measurement of the INVSENSOR00013 to the respiratory rate reference (RRef) and calculating the arithmetic root mean square (Arms) value. In order to obtain the Arms value, the RR measurement from the reference is subtracted from the INVSENSOR00013 RRp measurement for a number of samples. The average of this difference is computed as the bias. The standard deviation of the differences is computed as the precision. The square root of the sum of the squares of bias and precision is computed as the Arms value.

Sponsors and collaborators

Lead sponsor

Masimo Corporation

Industry

Registry information

Important dates

Study start
2018
Primary completion
2018
Study completion
2018
First posted
Feb 22, 2018
Registry last updated
Apr 23, 2019

OpenTrials presents study information sourced from ClinicalTrials.gov. The official registry record should be consulted for the latest information.

View the official ClinicalTrials.gov record (opens in a new tab)

This listing is for discovery and informational purposes only. It is not medical advice, does not guarantee that a study is recruiting, and does not determine eligibility. Contact the study team and a qualified healthcare professional when considering participation.

Published trials that share one or more normalized conditions with this study.