Incidence & Predictors of Mechanical Abnormalities of QuickSite® and QuickFlex® Left Ventricular Pacing Leads

On April 2, 2012 the manufacturer of the Quicksite® and Quickflex® leads, St. Jude Medical, released a voluntary medical device advisory regarding these leads' potential for problems with the silicone insulation covering one end of the lead, where it meets another section of the lead which is coated with a different insulating material (polyurethane). There has been a voluntary suspension of sales of these leads.

There are approximately 179,000 of the above leads implanted worldwide with an estimated 101,000 leads currently in service (59,000 in US). There have been 39 confirmed cases of the conductor portion of the lead becoming exposed from the silicone insulation (externalization), and the rate of occurrence of this type of problem is conservatively estimated at less than one percent. However, a review done by St. Jude Medical of returned Quicksite® and Quickflex® leads and available X-ray and fluoroscopic (x-ray movie) images of patients with these leads (from 1,219 patients) revealed a higher rate of 3-4%. It is presumed that there are patients whose leads have not yet been affected by this problem, but whose leads may show signs of structural changes on x-ray or fluoroscopy. This study seeks to determine what structural changes may be occurring in these leads, what signs might predict these changes, and what impact these changes may have on the lead and its performance.

Completed

Looking for future studies?

Notify Me

Key information

Age range

18 year and older

Sex eligibility

All sexes

Study type

Observational

Primary location

University of Kansas Medical Center

Kansas City, Kansas, 66160, United States

Who can participate

Healthy volunteers accepted: No

Only the study team can determine whether someone qualifies for participation.

Inclusion criteria

  • Patient that received Quicksite or Quickflex lead at the University of Kansas Hospital

Exclusion criteria

  • Not willing to consent to data collection

Treatment and study plan

Quicksite

Device

St. Jude Medical produced lead called Quicksite

Quickflex

Device

St. Jude Medical produced lead called Quickflex

Primary outcomes

  1. Evaluate the mechanical/structural abnormalities in implanted St Jude Medical QuickSite® and QuickFlex® left ventricular (LV) pacing leads

Sponsors and collaborators

Lead sponsor

Dhanunjaya Lakkireddy, MD, FACC

Other

Registry information

Official study title

Incidence & Predictors of Mechanical Abnormalities of QuickSite® and QuickFlex® Left Ventricular Pacing Leads: Impact on Electrophysiological Profile

Important dates

Study start
2012
First posted
Jul 10, 2012
Registry last updated
Aug 9, 2013

OpenTrials presents study information sourced from ClinicalTrials.gov. The official registry record should be consulted for the latest information.

View the official ClinicalTrials.gov record (opens in a new tab)

This listing is for discovery and informational purposes only. It is not medical advice, does not guarantee that a study is recruiting, and does not determine eligibility. Contact the study team and a qualified healthcare professional when considering participation.