Skip to main content
OpenTrials
Completed

NCT Number: NCT05873335

In Vivo Pulse Oximeter Validation Study

The goal of this clinical trial is to compare values of SpO2 provided by the DUT and the SaO2 used as a reference in healthy adults. The main question it aims to answer is: what is the accuracy of the subject device compared to the 510(k)-cleared reference device

A radial arterial cannula was placed in the wrist of each subject for arterial blood sampling and a subject device was placed on each arm of the patients. Then, hypoxia was induced to different stable levels of oxyhemoglobin saturation between 70-100% by having subjects breathe mixtures of nitrogen, air, and carbon dioxide controlled by the study physician. Values of SpO2 and SaO2 were then compared to provide a root mean square error and conclude on the subject device accuracy.

Completed

Looking for future studies?

Notify Me

Key information

Age range

21 year–50 year

Sex eligibility

All sexes

Study type

Interventional

Phase

Not applicable

Primary location

Hypoxia Research Laboratory

San Francisco, California, 94143, United States

Who can participate

Healthy volunteers accepted: Yes

Only the study team can determine whether someone qualifies for participation.

Inclusion criteria

  • Healthy adult
  • Normal hemoglobin levels (Hemoglobin > 10 g/dL)

Exclusion criteria

  • Systemic illness
  • Obesity

Treatment and study plan

Mixtures of nitrogen, air, and carbon dioxide

Combination Product

Hypoxia was induced to different and stable levels of oxyhemoglobin saturation (between 70-100%) by having subjects breathe mixtures of nitrogen, air, and carbon dioxide controlled by the study physician by adjusting gas flows with valves according to breath-by-breath estimates of oxygen saturation calculated from end-tidal PO2 and PCO2 displayed on a screen using LabVIEW 2015

Other names: Radial arterial cannula

Primary outcomes

  1. Measure of Accuracy (ARMS [%])

    Time frame: 1 hour

    Arms, given in percentage or rms error, is the calculation of root mean square between SpO2 values from the subject device and the SaO2 from the Hemoximeter. It is a way of averaging the absolute values of errors over the full measurement range and is a similar measurmeent to a "rms error". It indicates how far (accurate) the provided value is from the reference used.

Sponsors and collaborators

Lead sponsor

Gabi SmartCare

Industry

Registry information

Official study title

In Vivo Testing for SpO2 Accuracy on Healthy Adults Under Laboratory Conditions

Important dates

Study start
2023
Primary completion
2023
Study completion
2023
First posted
May 24, 2023
Registry last updated
Dec 12, 2024

OpenTrials presents study information sourced from ClinicalTrials.gov. The official registry record should be consulted for the latest information.

View the official ClinicalTrials.gov record (opens in a new tab)

This listing is for discovery and informational purposes only. It is not medical advice, does not guarantee that a study is recruiting, and does not determine eligibility. Contact the study team and a qualified healthcare professional when considering participation.