Investigational Blood Glucose Monitoring System
DeviceSubjects with diabetes used a new blood glucose monitoring system with capillary blood. Certain results were compared to a laboratory glucose method.
NCT Number: NCT00803777
The purpose of this study is to evaluate the performance and acceptability of a new blood glucose meter.
Looking for future studies?
Notify Me4 year–24 year
All sexes
Interventional
Not applicable
AMCR Institute, Escondido, California, United States
Healthy volunteers accepted: No
Only the study team can determine whether someone qualifies for participation.
Inclusion criteria
Exclusion criteria
Subjects with diabetes used a new blood glucose monitoring system with capillary blood. Certain results were compared to a laboratory glucose method.
Time frame: 1-2 hours
Subjects with diabetes (or parents/guardians) and healthcare professionals (HCPs) used a new blood glucose monitoring system (BGMS) with subject capillary blood. BGM results were compared to a lab glucose method - Yellow Springs Instrument (YSI) Analyzer. BG results were obtained in duplicate from subjects(158 BG results possible). The number of capillary results within +/- 15mg/dL (for reference blood glucose values <75mg/dL) or +/- 20% (for reference blood glucose values >/= 75mg/dL)of the reference results were calculated.
Time frame: 1-2 hours
Duplicate subject Blood Glucose Monitoring System (BGMS) results were compared to healthcare professional (HCP) BGMS results (possible number of results = 292). The number of Subject BGM results within +/- 15mg/dL (for reference BG values <75mg/dL) and within +/- 20% (for reference BG values >= 75mg/dL)of the HCP results was calculated.
Time frame: 1-2 hours
The Parkes Error Grid, developed from a survey of 100 clinicians, plots combinations of BG meter measurements against reference method results. Each (x,y) point on the grid is within a risk category, assigned by the clinicians surveyed. Risk categories (increasing severity): ZoneA: No effect on clinical action; ZoneB: Altered clinical action or little/no effect on clinical outcome; ZoneC: Altered clinical action likely to effect clinical outcome; ZoneD: Altered clinical action could have significant medical risk; ZoneE: Altered clinical action could have dangerous consequences
Time frame: 1-2 hours
Subjects or parents/guardians, as applicable, performed meter tasks after reading the User Guide and Quick Reference Guide. The study staff rated the participants on their success at performing the tasks as follows:
Time frame: 1-2 hours
The average Coefficient of Variation (CV) was computed by calculating the square of the CV for each pair of Blood Glucose (BG) self-test results, averaging this square value over the number of subjects included in the analysis, and then taking the square root.
Ascensia Diabetes Care
Industry
In-Clinic and Home-Use Study of the AVATAR Blood Glucose Monitoring System in Children and Young Adults
OpenTrials presents study information sourced from ClinicalTrials.gov. The official registry record should be consulted for the latest information.
View the official ClinicalTrials.gov record (opens in a new tab)This listing is for discovery and informational purposes only. It is not medical advice, does not guarantee that a study is recruiting, and does not determine eligibility. Contact the study team and a qualified healthcare professional when considering participation.
Published trials that share one or more normalized conditions with this study.
NCT00837122
Cardiovascular Diseases, Diabetes
Accra, Ghana
View Trial DetailsNCT03899883
Diabetes, Diabetes Complications
Aurora, Colorado, United States
View Trial DetailsNCT00065676
Body Weight, Diabetes
Bethesda, Maryland, United States
View Trial DetailsNCT00862433
Body Weight, Diabetes
Bethesda, Maryland, United States
View Trial Details