Skip to main content
OpenTrials
Completed

NCT Number: NCT02155218

Improving the Concentration of MR Contrast in the Arteries

In Contrast-Enhanced Magnetic Resonance Angiography (CE-MRA) a contrast agent is injected into an arm vein using a programmable medical device called a power injector. Once the contrast has circulated to the blood vessel of interest they can be imaged with MRI and distinctly separated from surrounding non-vascular tissue.

This project is designed to find the best way to administer the contrast agent during CE-MRA to produce the best images. The clinical availability of programmable power injectors facilitates the use of variable contrast rate and volume injection protocols that are based on patient specific parameters.

CE-MRA is a valuable imaging tool for diagnosing abnormalities of the major blood vessels in the chest and abdomen. It is often used in conjunction with cardiac magnetic resonance (CMR) examinations to evaluate the blood vessels of the heart and lung. CE-MRA is attractive because 1) it does not involve radiation, and 2) the contrast agents used are not toxic to the kidneys and can be used in patients who are at greater risk for kidney dysfunction following iodinated contrast agents.

We plan to administer a small (1cc) test amount of contrast, taking pictures as the contrast passes through the blood vessels, and analyze the results. Using this information, we will administer an individually patient-tailored injection given in multiple phases (i.e., a "multi-phasic" injection) to maintain constant contrast concentration (and therefore constant signal intensity) throughout the scan. We believe this will reduce image blurring, and improve the sharpness and likely the diagnostic quality of CE-MRA.

Completed

Looking for future studies?

Notify Me

Key information

Age range

18 year and older

Sex eligibility

All sexes

Study type

Interventional

Phase

Not applicable

Primary location

University of Washington

Seattle, Washington, 98195, United States

About this study

This project will build and validate a tool designed to optimize the use of magnetic resonance (MR) contrast agents for contrast-enhanced magnetic resonance angiography (CE-MRA). As with most angiography methods, contrast agents selectively highlight a patient's arteries as compared to surrounding tissue. In CE-MRA the contrast agent is injected into an arm vein using a programmable injector and transported by the body's circulation to the arteries of interest, at which point a MR acquisition depicts the arteries distinctly from surrounding soft tissue.

At present, the clinically accepted method of injecting the contrast agent is essentially the same for every patient, and has not been systematically evaluated or optimized. For most chest and abdominal CE-MRA studies contrast is injected over 5-15 seconds, and the resultant concentration of contrast agent in the arteries over the duration of the MR acquisition is not well known. What this concentration is and how it varies over time, however, has a large impact on the quality and clarity of the resulting images. This project is designed to optimize the concentration of contrast agent in the arteries during the MR scan by using the predictive results of a small "test bolus" injection to tailor the actual contrast injection for each individual patient such that concentration remains as high and as constant as possible during the time of MR acquisition. This will be accomplished by developing algorithms and software tools to tailor the optimal, patient-specific injection parameters. The method will be validated in a preliminary clinical study.

We propose that we can predict what the contrast bolus will look like (i.e., its "concentration vs. time") by first administering a small (1cc) test bolus, taking pictures as the contrast passes through the blood vessels, and analyzing the results. Using this information, we believe that an individually patient-tailored injection given in multiple phases (i.e., a "multi-phasic" injection) can create constant contrast concentration (and therefore constant signal intensity) throughout the scan, eliminating this source of image blurring, and improving the sharpness and likely the diagnostic quality of first-pass CE-MRA. Moreover, by tailoring the contrast to the duration of the scan, the contrast agent is used most efficiently and not wasted outside the acquisition duration, potentially recouping any contrast-to-noise ratio (CNR) loss. It is important to note the results of this study are generalizable to CT Angiography (CTA) as well, where similar efficient use of the contrast will reduce cost and excessive patient exposure to iodinated contrast. We plan to compare an "optimized" multiphasic contrast administration to a standard single phase contrast administration.

Who can participate

Healthy volunteers accepted: No

Only the study team can determine whether someone qualifies for participation.

Inclusion criteria

  • 18 years of age or older;
  • Scheduled at University of Washington Radiology to undergo contrast-enhanced cardiac MR.

Exclusion criteria

  • Unable to understand the nature of the study or to consent;
  • Unwilling to participate;
  • Prisoner;
  • Pregnant;
  • Contraindications for MRI:
  • Metallic or other surgical implants
  • History of metal work; metal pieces in your eyes or other body parts
  • pacemaker
  • defibrillator
  • some aneurysm clips
  • some artificial heart valves
  • some implanted venous access device
  • some cochlear/inner ear implants
  • spinal stimulator
  • some intrauterine devices (IUD)
  • some tattoos or permanent make-up
  • transdermal patch
  • severe anxiety in confined spaces (claustrophobia)
  • Contraindications for contrast:
  • an allergic reaction to gadolinium contrast in the past that required treatment
  • severe kidney or liver disease

Treatment and study plan

Standard Injection Rate

Other

Normally an IV is placed in an arm and the gadolinium contrast (30-40 cc, depending on patient size) is administered through the IV with a power injector as a single bolus. Subjects will have a test bolus and main bolus.

Patient Tailored Injection Rate

Other

Normally an IV is placed in an arm and the gadolinium contrast (30-40 cc, depending on patient size) is administered through the IV with a power injector as a single bolus. Subjects will have a test bolus and main bolus. We will use a mathematical algorithm to rapidly analyze the test bolus and calculate a predicted "best" way to inject the main bolus - likely slower and "multi-phasic", meaning different flow rates as the bolus injection evolves.

gadolinium contrast

Other

Primary outcomes

  1. Arterial Signal Intensity Evaluation

    Time frame: During the one-time contrast-enhanced cardiac MR exam lasting approximately 45 to 60 minutes, signal intensity will be evaluated continuously before, during and after administration of contrast.

    To evaluate the performance of the bolus-tailoring method, measured signal intensity profiles will be evaluated using residual sum-of-squares (RSS) between targeted and measured responses. A smaller RSS indicates agreement with the targeted signal intensity profile. Additionally, full-width-at-half-maximum (FWHM) and start time (10% signal intensity increase over baseline) of the signal intensity profile will be measured and compared between the two groups. T-test will be used to test for significant differences between the two methods.

Sponsors and collaborators

Lead sponsor

University of Washington

Other

Registry information

Official study title

Contrast Dynamics of the Thoracic Vasculature and Heart

Important dates

Study start
2014
Primary completion
2015
Study completion
2015
First posted
Jun 4, 2014
Registry last updated
Oct 26, 2017

OpenTrials presents study information sourced from ClinicalTrials.gov. The official registry record should be consulted for the latest information.

View the official ClinicalTrials.gov record (opens in a new tab)

This listing is for discovery and informational purposes only. It is not medical advice, does not guarantee that a study is recruiting, and does not determine eligibility. Contact the study team and a qualified healthcare professional when considering participation.

Published trials that share one or more normalized conditions with this study.