IBDoc calprotectin test
Diagnostic TestPatients are asked to perform a single IBDoc calprotectin home test.
NCT Number: NCT03408249
The objective of the IBDoc® user performance evaluation is to demonstrate the ease-of-use of the IBDoc® calprotectin home test to allow patients with inflammatory bowel disease (IBD) to independently and correctly determine calprotectin concentration in their own stool sample. The study is based on Chapter 8 of the International Organization for Standardization (ISO) Standard, ISO 15197:2013 "In vitro diagnostic test systems - Requirements for blood-glucose monitoring systems for self-testing in managing diabetes mellitus". The study is a prospective, multicenter study, not blinded for patients, and includes a total of 61 patients.
Looking for future studies?
Notify Me12 year and older
All sexes
Interventional
Not applicable
Division of Gastroenterology, Department of Medicine, Faculty of Medicine University of Alberta,, Edmonton, Alberta, Canada
The patient will determine the calprotectin concentration in their own stool sample using the IBDoc® assay. For this, the patient will collect a stool sample in a standard collection tube at home. The patients will return to the clinical site with the specimen, where they will receive a standardized IBDoc® training. They will also be given opportunity to perform an IBDoc® practice test using the collected sample. The patients will then perform the full IBDoc® assay using the collected stool sample under the observation of the healthcare provider, who will assess their performance. The final result - the interpretation of the IBDoc® lateral flow assay by the CalApp® smartphone application, to yield a final μg/g calprotectin concentration in stool, will be sent to the health care provider (HCP) via the IBDoc® Web Portal.
The usability of the IBDoc® test will be evaluated according to the following factors:
To exclude bias or inaccuracy that may arise from the analytical performance characteristics of the IBDoc® assay itself, a comparative IBDoc® assay will be performed according to the instructions for use, with the patients' samples in a laboratory. For the comparative IBDoc® measurement the HCP at the clinical site will obtain one CALEX® Valve extract from the patient's stool sample, which will be stored frozen and shipped to the laboratory site.
-IBDoc® Calprotectin test usability, clearness of obtained result, and comprehensiveness of the Instruction for Use will be assessed in the following ways:
The ability of patients to correctly collect and extract stool samples using the CALEX® Valve device will be analysed as an additional module within the IBDoc® User Performance Evaluation. The CALEX® Valve extracts obtained by the patients as part of the IBDoc® procedure will be stored frozen and shipped to the laboratory site. Calprotectin levels in the patients' CALEX® Valve extracts will be determined with the BÜHLMANN fCAL® ELISA and compared to calprotectin reference values obtained from the same stool samples by the HCP's.
Healthy volunteers accepted: No
Only the study team can determine whether someone qualifies for participation.
Inclusion criteria
Exclusion criteria
Patients are asked to perform a single IBDoc calprotectin home test.
Patients evaluate the ease-of-use of the test, clearness of test results and comprehensiveness of the instructions for use through a specific questionnaire implementing Likert scales.
Time frame: 24 hours
Patients' IBDoc results are compared to reference calprotectin values obtained by laboratory methods
Time frame: 24 hours
Ease-of-use of the test, clearness of test results and comprehensiveness of the instructions for use is assessed in a questionnaire, with 12 statements, scored on a 1-to-5 Likert scale, with one (1) indicating "I strongly disagree" and five (5) "I strongly agree". The mean and standard deviation of scores obtained from all participating patients will be evaluated for each statement.
Time frame: 24 hours
Quantifiable feedback from the healthcare providers about the ability of the patients to correctly perform the IBDoc® assay. Assessment will be made through a questionnaire with five (5) statements, scored on a 1-to-5 Likert scale, with one (1) indicating "I strongly disagree" and five (5) "I strongly agree". The mean and standard deviation of scores obtained from all participating healthcare providers will be evaluated for each statement.
Bühlmann Laboratories AG
Industry
OpenTrials presents study information sourced from ClinicalTrials.gov. The official registry record should be consulted for the latest information.
View the official ClinicalTrials.gov record (opens in a new tab)This listing is for discovery and informational purposes only. It is not medical advice, does not guarantee that a study is recruiting, and does not determine eligibility. Contact the study team and a qualified healthcare professional when considering participation.
Published trials that share one or more normalized conditions with this study.
NCT06094608
Colitis, Colitis, Ulcerative
Ann Arbor, Michigan, United States
View Trial DetailsNCT04923100
Colitis, Colitis, Ulcerative
Guangzhou, Guangdong, China
View Trial DetailsNCT04647578
Colitis, Colitis, Ulcerative
Lille, France
View Trial DetailsNCT04191655
Colitis, Colitis, Ulcerative
Boston, Massachusetts, United States
View Trial Details