Wills Eye Hospital
Philadelphia, Pennsylvania, 19107, United States
NCT Number: NCT02622178
Evaluate sensitivity and specificity of NOVA-DN visually evoked potentials (VEP) protocol and new software method (Corda) for glaucoma detection using optical coherence tomography (OCT) images to differentiate between normal subjects and glaucoma suspects.
Looking for future studies?
Notify Me18 year–90 year
All sexes
Interventional
Not applicable
Philadelphia, Pennsylvania, 19107, United States
The aim of this study was to evaluate the ability of NOVA-DN VEP protocol and Corda parameters to discriminate between healthy eyes and eyes with early to moderate glaucomatous visual field loss. We will evaluate measurements of the NOVA-DN VEP protocol, Corda software and Cirrus spectral domain (SD) OCT software in order to compare and correlate.
Hypotheses is that NOVA-DN VEP protocol and Corda analysis results in a high sensitivity, specificity and area under the Receiver Operating Characteristic (ROC) curves (ROC area) for glaucoma detection. Second hypothesis is NOVA-DN VEP protocol and Corda parameters can differentiate between normal and glaucoma suspects.
Healthy volunteers accepted: Yes
Only the study team can determine whether someone qualifies for participation.
Inclusion criteria
Group 1: Healthy Subjects
Group 2: Glaucoma Suspects
Group3: Glaucoma Patients
Exclusion criteria
Optical coherence tomography (OCT) is a noninvasive imaging modality that provides micrometer-scale resolution.It has been revolutionized in recent years by exploitation of Fourier domain (FD) techniques, which have a significant sensitivity advantage over traditional time domain (TD) OCT. In spectral-domain (SD-OCT) the reference mirror is stationary, and OCT signal is acquired using a spectrometer as detector or by varying the wavelength of the light source.
Other names: OCT
Visual evoked potential is a means of objectively testing visual field by viewing a computer monitor at 1 meter with a square black/white checkerboard pattern reversal stimulus. Electrodes are placed on the head and face to monitor Electroencephalogram (EEG) activity during testing. Output parameters from VEP system include amplitude and latency measures for each stimuli.
Other names: VEP
Time frame: 1 examination, one hour
Diopsys Visual evoked potential (VEP) (Diopsys, Inc. Pine Brook, NJ) is used to objectively measure the functional responses of the entire visual pathway from the anterior segment of the eye to the visual cortex. This is the strength of the signal recorded from the back of the head near where vision is processed in the brain while a visual stimulus is presented to patient.
Time frame: 1 examination, one hour
Optical coherence tomography (OCT) images provide measurement of the retinal nerve fiber layer thickness in microns.
Wills Eye
Other
Sensitivity and Specificity of the NOVA-DN VEP Protocol and a Novel Analysis of Optical Coherence Tomography Images for Glaucoma Diagnosis
OpenTrials presents study information sourced from ClinicalTrials.gov. The official registry record should be consulted for the latest information.
View the official ClinicalTrials.gov record (opens in a new tab)This listing is for discovery and informational purposes only. It is not medical advice, does not guarantee that a study is recruiting, and does not determine eligibility. Contact the study team and a qualified healthcare professional when considering participation.