Hemodynamics Measurement in Radiofrequency Catheter Ablation

This study will occur in patients undergoing routine cardiac ablation of his/her arrhythmia with planned use of an irrigated ablation catheter. The main objective of this study is to take intracardiac pressure measurements and pressure waveforms with both a "gold standard" balloon tipped pulmonary artery catheter placed for this study and an irrigated ablation catheter placed as standard of care for the procedure.

Completed

Looking for future studies?

Notify Me

Key information

Age range

18 year and older

Sex eligibility

All sexes

Study type

Interventional

Phase

Not applicable

Primary location

Emory Clinic, Atlanta, Georgia, United States

Loading trial locations.

About this study

Measuring internal heart pressures allows for the assessment and diagnosis of a variety of cardiac conditions which are commonly found in patients undergoing catheter ablation, including congestive heart failure, diastolic dysfunction, and valvular heart disease. Currently it is unknown if taking pressure measurements with the ablation catheter is accurate and it is not standard practice to insert the additional 'gold standard' pressure sensing catheter. Catheter ablation of cardiac arrhythmia is performed to offer symptomatic relief from a variety of atrial, ventricular and atrio-ventricular rhythms. Irrigated ablation catheters are commonly used for ablation of arrhythmias arising from the left heart and coronary sinus. These catheters contain internal channels allowing for infusion of an external fluid directly to the tip of the catheter, where the fluid is sprayed through numerous end holes allowing for cooling of the catheter ablation tip.

Fluid filled catheters are also the primary tool utilized to assess intravascular and intracardiac pressures and hemodynamics in the cardiac catheterization suite and intensive care unit. However, the fluid filled catheters commonly used for these assessments are single lumen with a single end hole and generally have a larger minimal diameters compared to the ablation catheter. "Dampening" of pressure waveforms are observed in smaller diameter catheters and may lead to underestimation of peak pressure and overestimation of the nadir pressure gradient. It is unknown if transduction of pressures from the tip of an irrigated ablation catheter will yield equivalent results compared to use of a single 5 French (5F) sized lumen catheter.

This study will seek to enroll patients already scheduled to have a cardiac ablation for a clinical arrhythmia at Emory University Hospital. After completion of the standard ablation, the ablation catheter will be positioned in the heart and pressures will be measured. Next, a standard pressure sensing catheter will be positioned in the same area in the heart and pressures will be measured. The pressure values for each heart chamber assessed will be compared between the catheters. The purpose of this study is to test the hypothesis that transduction of intracardiac pressures using an irrigated ablation catheter placed as standard of care for the ablation procedure is equivalent to transduction of pressures using a standard 5F balloon tipped PA catheter placed only for this research study.

Who can participate

Healthy volunteers accepted: No

Only the study team can determine whether someone qualifies for participation.

Inclusion criteria

  • patient undergoing routine cardiac ablation of his/her arrhythmia with planned use of an irrigated ablation catheter

Exclusion criteria

  • inability to provide informed consent

Treatment and study plan

Irrigated ablation catheter

Device

After completion of ablation, the ablation catheter will be serially positioned in the cardiac chambers (either the left or right side of the heart). The stop cock will be turned to the pressure transducer and waveforms will be stored corresponding to each chamber.

Balloon tipped PA catheter

Device

After removal of the irrigated ablation catheter, a 5F balloon tipped PA catheter which will be positioned under fluoroscopic guidance in each previously evaluated chamber and the chambers will be measured again.

Primary outcomes

  1. Systolic Pressure Value

    Time frame: Day 1 (during procedure)

    The systolic pressure values for atrium and ventricle heart chambers are presented for the irrigated ablation catheter derived measurements and the balloon tipped PA catheter measurements.

  2. Diastolic Pressure Value

    Time frame: Day 1 (during procedure)

    The diastolic pressure values for ventricle heart chambers are presented for the irrigated ablation catheter derived measurements and the balloon tipped PA catheter measurements.

Sponsors and collaborators

Lead sponsor

Emory University

Other

Registry information

Important dates

Study start
2020
Primary completion
2020
Study completion
2020
First posted
Jan 13, 2020
Registry last updated
Jun 14, 2021

OpenTrials presents study information sourced from ClinicalTrials.gov. The official registry record should be consulted for the latest information.

View the official ClinicalTrials.gov record (opens in a new tab)

This listing is for discovery and informational purposes only. It is not medical advice, does not guarantee that a study is recruiting, and does not determine eligibility. Contact the study team and a qualified healthcare professional when considering participation.

Published trials that share one or more normalized conditions with this study.