Cochlear™ Osia® System (Osia® 2)
DeviceConsists of Osia 2 Sound Processor, Osia 2 Sound Processor Magnets, and Osia Fitting Software 2.
NCT Number: NCT07156461
The goal of this interventional study is to evaluate the clinical performance and sound quality of the Osia 3 Sound Processor and to demonstrate its superiority compared with the Osia 2 Sound Processor in adults with mixed hearing loss, conductive hearing loss, or single-sided deafness.
The main questions this study aims to answer are:
* Do the findings confirm the clinical performance of the Osia 3 Sound Processor? * Is the Osia 3 Sound Processor superior compared to the Osia 2 Sound Processor?
Participants will:
* Undergo speech performance testing in both quiet and noisy environments * Provide ratings for a self-reported questionnaire
Looking for future studies?
Notify Me18 year and older
All sexes
Interventional
Not applicable
Cochlear Macquarie, Macquarie University, Sydney, New South Wales, Australia
Healthy volunteers accepted: No
Only the study team can determine whether someone qualifies for participation.
Inclusion criteria
Exclusion criteria
Consists of Osia 2 Sound Processor, Osia 2 Sound Processor Magnets, and Osia Fitting Software 2.
Consists of Osia 3 Sound Processor, Osia 3 Sound Processor Magnets, Osia 3 Charger, and Osia Fitting Software 3.
Time frame: At Baseline (Day 0), Visit 1 (7-14 days after Baseline), and Visit 2 (7-14 days after Visit 1)
Speech perception in noise will be assessed using the Australian Speech Test in Noise (AuSTIN) at the Australian sites and the Bamford-Kowal-Bench Speech-In-Noise (BKB-SIN) test at the US site. The AuSTIN uses BKB-SIN like target sentences presented in noise and comprises of 80 lists of 20 sentences each. The BKB-SIN test contains 18 list pairs of sentences, each containing 3-4 key words, presented at 65 dB SPL with adaptive noise levels. Both tests aim to determine the signal-to-noise ratio (SNR) at which 50% of the key words are correctly repeated. For the unaided and aided measurements, speech performance in noise will be measured with speech and noise presented from a front speaker, positioned at 1 m from the subject.
Time frame: At Baseline (Day 0), Visit 1 (7-14 days after Baseline), and Visit 2 (7-14 days after Visit 1)
Speech perception in quiet will be measured using the consonant-nucleus-consonant (CNC) monosyllabic words presented at 65 dB SPL from the front speaker. The CNC word test consists of 30 lists each with 50 words per list. The goal of the speech perception assessment in quiet is to obtain the percentage of correctly repeated words in a list, i.e., the word recognition score.
Time frame: At Baseline (Day 0), Visit 1 (7-14 days after Baseline), and Visit 2 (7-14 days after Visit 1)
Thresholds in sound field will be measured using warble tones presented from a front speaker positioned at 1 m from the subject following the modified Hughson-Westlake procedure. Frequencies to be tested are 250, 500, 750, 1000, 1500, 2000, 3000, 4000, 6000, and 8000 Hz.
Time frame: At Visit 1 (7-14 days after Baseline) and Visit 2 (7-14 days after Visit 1)
Speech perception in noise with a fixed SNR will be measured using the AzBio sentence test, which consists of 15 lists of 20 sentences each. AzBio sentences are spoken by different talkers in a conversational style with limited contextual cues that the listener can use to predict or 'fill in' unintelligible words. The sentences will be presented at a fixed level of 65 dB SPL at a fixed SNR of 0 dB SNR using the accompanying noise stimulus. Each word in the sentence counts towards the overall score (percent correct). Speech will be presented from the front, while noise will be presented from the rear.
Time frame: At Visit 1 (7-14 days after Baseline) and Visit 2 (7-14 days after Visit 1)
Speech perception in noise with a fixed SNR will be measured using the AzBio sentence test, which consists of 15 lists of 20 sentences each. AzBio sentences are spoken by different talkers in a conversational style with limited contextual cues that the listener can use to predict or 'fill in' unintelligible words. The sentences will be presented at a fixed level of 65 dB SPL at a fixed SNR of +5 dB SNR using the accompanying noise stimulus. Each word in the sentence counts towards the overall score (percent correct). Speech will be presented from the front, while noise will be presented from the rear.
Time frame: At Visit 1 (7-14 days after Baseline) and Visit 2 (7-14 days after Visit 1)
The SSQ-12 involves a self-reported hearing function and the impact of hearing treatment in daily life across a variety of listening situations. It includes 3 subcategories: Speech Hearing, Spatial Hearing, and Qualities of Hearing, with each question rated on a Likert scale from 0-10 (minimum - maximum hearing abilities).
Time frame: At Visit 1 (7-14 days after Baseline) and Visit 2 (7-14 days after Visit 1)
Speech perception in quiet will be measured using the consonant-nucleus-consonant (CNC) monosyllabic words presented at 50 dB SPL from the front speaker. The CNC word test consists of 30 lists each with 50 words per list. The goal of the speech perception assessment in quiet is to obtain the percentage of correctly repeated words in a list, i.e., the word recognition score.
Time frame: At Visit 1 (7-14 days after Baseline) and Visit 2 (7-14 days after Visit 1)
Speech perception in quiet will be measured using the consonant-nucleus-consonant (CNC) monosyllabic words presented at 65 dB SPL from the front speaker. The CNC word test consists of 30 lists each with 50 words per list. The goal of the speech perception assessment in quiet is to obtain the percentage of correctly repeated words in a list, i.e., the word recognition score.
Time frame: At Visit 1 (7-14 days after Baseline) and Visit 2 (7-14 days after Visit 1)
Speech perception in noise will be assessed using the Australian Speech Test in Noise (AuSTIN) at the Australian sites and the Bamford-Kowal-Bench-Speech in Noise (BKB-SIN) test at the US site. The AuSTIN uses BKB-SIN like target sentences presented in noise and comprises of 80 lists of 20 sentences each. The BKB-SIN test contains 18 list pairs of sentences, each containing 3-4 key words, presented at 65 dB SPL with adaptive noise levels. Both tests aim to determine the signal-to-noise ratio (SNR) at which 50% of the key words are correctly repeated. For the unaided and aided measurements, speech performance in noise will be measured with speech and noise presented from a front speaker, positioned at 1 m from the subject.
Time frame: At Visit 1 (7-14 days after Baseline) and Visit 2 (7-14 days after Visit 1)
Thresholds in sound field will be measured using warble tones presented from a front speaker positioned at 1 m from the subject following the modified Hughson-Westlake procedure. Frequencies to be tested are 250, 500, 750, 1000, 1500, 2000, 3000, 4000, 6000 and 8000 Hz.
Cochlear
Industry
A Pivotal, Interventional, Prospective, Multi-centre Clinical Investigation of Hearing Outcomes With the Osia 3 Sound Processor Compared to the Osia 2 Sound Processor in Adult Osia Implant Recipients With Mixed Hearing Loss, Conductive Hearing Loss or Single-Sided Deafness
Acronym: SONUS
OpenTrials presents study information sourced from ClinicalTrials.gov. The official registry record should be consulted for the latest information.
View the official ClinicalTrials.gov record (opens in a new tab)This listing is for discovery and informational purposes only. It is not medical advice, does not guarantee that a study is recruiting, and does not determine eligibility. Contact the study team and a qualified healthcare professional when considering participation.
Published trials that share one or more normalized conditions with this study.
NCT07130136
Behavior, Bone Conduction
London, United Kingdom
View Trial DetailsNCT06634784
Children Under 5 Years, Ear Diseases
Hanover, Lower Saxony, Germany
View Trial DetailsNCT04761926
Ear Diseases, Hearing Disorders
Aalborg, Denmark
View Trial DetailsNCT04427033
Ear Diseases, Hearing Disorders
Innsbruck, Austria
View Trial Details